Repeatability and agreement of a new anterior segment optical coherence tomography with standard devices in normal, keratoconus, and post-refractive eyes

被引:0
|
作者
Kundu, Gairik [1 ]
Shetty, Naren [2 ]
Modak, Durgalaxmi [1 ]
Kaweri, Luci [3 ]
Khamar, Pooja [2 ]
Arora, Vishal [4 ]
Nuijts, Rudy M. M. A. [5 ]
Shetty, Rohit [1 ]
机构
[1] Narayana Nethralaya, Dept Cornea & Refract Surg, Bengaluru, Karnataka, India
[2] Narayana Nethralaya, Dept Cataract & Refract Surg, Bengaluru, Karnataka, India
[3] ASG Eye Hosp, Dept Cataract & Refract Surg, Mysore, Karnataka, India
[4] Aro Hlth, Dept Cataract & Refract Surg, Gurugram, Haryana, India
[5] Maastricht Univ Med Ctr, Dept Ophthalmol, Maastricht, Netherlands
关键词
Agreement; Anterion; Eyestar; 900; MS-39; Pentacam AXL Wave; repeatability; IMAGING-SYSTEMS; OCT; REPRODUCIBILITY; HEALTHY; ABERROMETER; PARAMETERS; BIOMETERS;
D O I
10.4103/IJO.IJO_805_24
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose:To assess repeatability and agreement of Eyestar 900 (Haag-Streit, K & ouml;niz, Switzerland) with Pentacam AXL Wave system (Oculus, Wetzlar, Germany), Anterion (Heidelberg Engineering GmBH, Heidelberg, Germany), and MS-39 (CSO, Florence, Italy) in measuring anterior segment parameters. Methods:Two hundred normal eyes, 50 keratoconus (KC) eyes, and 50 post-refractive surgery eyes underwent three sequential scans on each instrument. The order of scans was randomized. Repeatability was assessed using within-subject standard deviation (Sw) and intraclass correlation coefficient (ICC). Bland-Altman analysis was used to assess the 95% limits of agreement. Results:Eyestar 900 had comparable repeatability to Anterion and MS-39 across groups. ICC was overall >0.8. Pentacam AXL Wave had lower Sw (better repeatability) across all groups. In normal eyes, Eyestar 900 can be used interchangeably with MS-39 and Anterion (P > 0.23). Eyestar 900 and Pentacam AXL Wave did not show agreement on any parameter, except anterior chamber depth (ACD) (P = 0.92). Eyestar 900 in KC eyes can be used interchangeably with MS-39, except for posterior keratometry (pK) and thinnest corneal thickness (TCT) (P < 0.001), and with Anterion, except for TCT and ACD (P < 0.001). In post-refractive eyes, Eyestar 900 can be used interchangeably with MS-39, except for pK (P < 0.001), and with Anterion, except for ACD (P < 0.001). Pentacam AXL Wave and Eyestar 900 cannot be used interchangeably in KC and post-refractive eyes. Conclusions:Newer devices such as Eyestar 900 can be used interchangeably with Anterion and MS-39 in normal eyes and partly in KC and post-refractive eyes. Pentacam AXL Wave and Eyestar 900 cannot be used interchangeably across the various groups. However, these multifunctional hybrid tools of combined topography, AS-OCT, can be helpful in refractive clinics.
引用
收藏
页码:325 / 334
页数:12
相关论文
共 50 条
  • [31] Repeatability of Corneal Epithelial Thickness Measurements Using Fourier-Domain Optical Coherence Tomography in Normal and Post-LASIK Eyes
    Ma, Xingxuan Jack
    Wang, Li
    Koch, Douglas D.
    CORNEA, 2013, 32 (12) : 1544 - 1548
  • [32] Agreement between anterior segment parameters obtained by a new ultrasound biomicroscopy and a swept-source fourier-domain anterior segment optical coherence tomography.
    Li, Xiuyuan
    Chang, Pingjun
    Li, Zhangliang
    Qian, Shuyi
    Zhu, Zehui
    Wang, Qianwei
    Yun-E, Zhao
    EXPERT REVIEW OF MEDICAL DEVICES, 2020, 17 (12) : 1333 - 1340
  • [33] Agreement between a new optical low coherence reflectometry biometer and an anterior segment optical coherence tomographer
    Yiran Wang
    Ting Wan
    Luze Liu
    Yuyuan Xue
    Xinyao Chen
    Giacomo Savini
    Domenico Schiano-Lomoriello
    Xingtao Zhou
    Jinjin Yu
    Jinhai Huang
    Eye and Vision, 10
  • [34] Prospective comparison of ultrasound biomicroscopy and anterior segment optical coherence tomography for evaluation of anterior chamber dimensions in European eyes with primary angle closure
    Mansouri, K.
    Sommerhalder, J.
    Shaarawy, T.
    EYE, 2010, 24 (02) : 233 - 239
  • [35] Repeatability of automatic measurements by a new anterior segment optical coherence tomographer combined with Placido topography and agreement with 2 Scheimpflug cameras
    Savini, Giacomo
    Schiano-Lomoriello, Domenico
    Hoffer, Kenneth J.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2018, 44 (04) : 471 - 478
  • [36] Reproducibility and repeatability of optical coherence tomography imaging of the optic nerve head in normal beagle eyes
    Bemis, Alexandra M.
    Pirie, Christopher G.
    LoPinto, Alexander J.
    Maranda, Louise
    VETERINARY OPHTHALMOLOGY, 2017, 20 (06) : 480 - 487
  • [37] Anterior Segment Optical Coherence Tomography in Pediatric Ocular Pathology: Imaging Study of 115 eyes
    Nadeem, Sana
    EXPERT REVIEW OF MEDICAL DEVICES, 2023, 20 (08) : 691 - 702
  • [38] Evaluation of CorvisST biomechanical parameters and anterior segment optical coherence tomography for diagnosing forme fruste keratoconus
    Shiga, Suguru
    Kojima, Takashi
    Nishida, Tomoya
    Nakamura, Tomoaki
    Ichikawa, Kazuo
    ACTA OPHTHALMOLOGICA, 2021, 99 (06) : 644 - 651
  • [39] The Repeatability of Retinal Layer Thickness Measurements with Spectral-Domain Optical Coherence Tomography in Normal Eyes
    Kang, Min Seok
    Yu, Seung-Young
    Kwak, Hyung Woo
    JOURNAL OF THE KOREAN OPHTHALMOLOGICAL SOCIETY, 2016, 57 (05): : 786 - 793
  • [40] Agreement, repeatability, and reproducibility of quantitative retinal layer assessment using swept-source and spectral-domain optical coherence tomography in eyes with retinal diseases
    Hou, Huiyuan
    Durbin, Mary K.
    El-Nimri, Nevin
    Fischer, Jeffrey L.
    Sadda, Srinivas R.
    FRONTIERS IN MEDICINE, 2023, 10