Repeatability and agreement of a new anterior segment optical coherence tomography with standard devices in normal, keratoconus, and post-refractive eyes

被引:0
|
作者
Kundu, Gairik [1 ]
Shetty, Naren [2 ]
Modak, Durgalaxmi [1 ]
Kaweri, Luci [3 ]
Khamar, Pooja [2 ]
Arora, Vishal [4 ]
Nuijts, Rudy M. M. A. [5 ]
Shetty, Rohit [1 ]
机构
[1] Narayana Nethralaya, Dept Cornea & Refract Surg, Bengaluru, Karnataka, India
[2] Narayana Nethralaya, Dept Cataract & Refract Surg, Bengaluru, Karnataka, India
[3] ASG Eye Hosp, Dept Cataract & Refract Surg, Mysore, Karnataka, India
[4] Aro Hlth, Dept Cataract & Refract Surg, Gurugram, Haryana, India
[5] Maastricht Univ Med Ctr, Dept Ophthalmol, Maastricht, Netherlands
关键词
Agreement; Anterion; Eyestar; 900; MS-39; Pentacam AXL Wave; repeatability; IMAGING-SYSTEMS; OCT; REPRODUCIBILITY; HEALTHY; ABERROMETER; PARAMETERS; BIOMETERS;
D O I
10.4103/IJO.IJO_805_24
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose:To assess repeatability and agreement of Eyestar 900 (Haag-Streit, K & ouml;niz, Switzerland) with Pentacam AXL Wave system (Oculus, Wetzlar, Germany), Anterion (Heidelberg Engineering GmBH, Heidelberg, Germany), and MS-39 (CSO, Florence, Italy) in measuring anterior segment parameters. Methods:Two hundred normal eyes, 50 keratoconus (KC) eyes, and 50 post-refractive surgery eyes underwent three sequential scans on each instrument. The order of scans was randomized. Repeatability was assessed using within-subject standard deviation (Sw) and intraclass correlation coefficient (ICC). Bland-Altman analysis was used to assess the 95% limits of agreement. Results:Eyestar 900 had comparable repeatability to Anterion and MS-39 across groups. ICC was overall >0.8. Pentacam AXL Wave had lower Sw (better repeatability) across all groups. In normal eyes, Eyestar 900 can be used interchangeably with MS-39 and Anterion (P > 0.23). Eyestar 900 and Pentacam AXL Wave did not show agreement on any parameter, except anterior chamber depth (ACD) (P = 0.92). Eyestar 900 in KC eyes can be used interchangeably with MS-39, except for posterior keratometry (pK) and thinnest corneal thickness (TCT) (P < 0.001), and with Anterion, except for TCT and ACD (P < 0.001). In post-refractive eyes, Eyestar 900 can be used interchangeably with MS-39, except for pK (P < 0.001), and with Anterion, except for ACD (P < 0.001). Pentacam AXL Wave and Eyestar 900 cannot be used interchangeably in KC and post-refractive eyes. Conclusions:Newer devices such as Eyestar 900 can be used interchangeably with Anterion and MS-39 in normal eyes and partly in KC and post-refractive eyes. Pentacam AXL Wave and Eyestar 900 cannot be used interchangeably across the various groups. However, these multifunctional hybrid tools of combined topography, AS-OCT, can be helpful in refractive clinics.
引用
收藏
页码:325 / 334
页数:12
相关论文
共 50 条
  • [21] Influence of Keratoconus Severity on Detecting True Progression with Scheimpflug Imaging and Anterior Segment Optical Coherence Tomography
    Ortiz-Toquero, Sara
    Fuente, Carlota
    Auladell, Clara
    Arnalich-Montiel, Francisco
    LIFE-BASEL, 2023, 13 (07):
  • [22] Repeatability in and interchangeability between the macular and the fast macular thickness map protocols: a study on normal eyes with Stratus optical coherence tomography
    Eriksson, Urban
    Alm, Albert
    ACTA OPHTHALMOLOGICA, 2009, 87 (07) : 725 - 730
  • [23] Agreement between a new optical low coherence reflectometry biometer and an anterior segment optical coherence tomographer
    Wang, Yiran
    Wan, Ting
    Liu, Luze
    Xue, Yuyuan
    Chen, Xinyao
    Savini, Giacomo
    Schiano-Lomoriello, Domenico
    Zhou, Xingtao
    Yu, Jinjin
    Huang, Jinhai
    EYE AND VISION, 2023, 10 (01)
  • [24] Keratometry Agreement Between Two Swept-Source OCT Devices in Healthy and Post-refractive Surgery Eyes
    Martinez-Albert, Noelia
    Carlos Nieto-Fernandez, Juan
    Vicente Garcia-Marques, Jose
    JOURNAL OF REFRACTIVE SURGERY, 2023, 39 (05) : 347 - +
  • [25] Comparison of a new anterior segment optical coherence tomography and Oculus Pentacam for measurement of anterior chamber depth and corneal thickness
    Li, Xuepei
    Zhou, Yijing
    Young, Charlotte Aimee
    Chen, Aiming
    Jin, Guangming
    Zheng, Danying
    ANNALS OF TRANSLATIONAL MEDICINE, 2020, 8 (14)
  • [26] Repeatability and reproducibility of a new spectral-domain optical coherence tomography biometer and agreement with swept-source optical coherence tomography based biometer
    Li, Xin
    Lei, Chak Seng
    Ning, Rui
    Liu, Luze
    Chen, Aodong
    Yang, Xinning
    Savini, Giacomo
    Schiano-Lomoriello, Domenico
    Zhou, Xingtao
    Huang, Jinhai
    EYE AND VISION, 2025, 12 (01)
  • [27] Comparison of corneal measurements in keratoconus eyes with two swept-source-optical coherence tomography devices and a Scheimpflug device
    Bograd, Alexandra
    Himmel, Isabelle
    Pfister, Isabel B.
    Seiler, Theo G.
    Frueh, Beatrice E.
    Tappeiner, Christoph
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2023, 262 (3) : 891 - 901
  • [28] Swept-source optical coherence tomography in ocular surface diseases: anterior segment analysis repeatability and its limits
    Recchioni, Alberto
    Venkataraman, Abinaya Priya
    Rauz, Saaeha
    Dominguez-Vicent, Alberto
    FRONTIERS IN MEDICINE, 2024, 11
  • [29] Anterior Segment Optical Coherence Tomography Assessment After Laser Capsulotomy in Pseudophakic Eyes With Pseudoexfoliation
    Eliacik, Mustafa
    Erdur, Sevil Karaman
    Gulkilik, Gokhan
    Ozsutcu, Mustafa
    Aras, Cengiz
    Bayramlar, Huseyin
    Aslan, Cemile Anil
    EYE & CONTACT LENS-SCIENCE AND CLINICAL PRACTICE, 2016, 42 (05): : 318 - 321
  • [30] Corneal Epithelial Mapping Characteristics in Normal Eyes Using Anterior Segment Spectral Domain Optical Coherence Tomography
    Abusamak, Mohammad
    TRANSLATIONAL VISION SCIENCE & TECHNOLOGY, 2022, 11 (03):