Hemodynamic analysis of fusiform-shaped inferior vena cava filter with thrombus capture: A computational fluid dynamics study

被引:0
作者
Wang, Jingying [1 ]
Li, Mingrui [1 ]
Chen, Shuai [2 ]
Lee, Chunhian [1 ,3 ]
机构
[1] Shandong Univ, Sch Energy & Power Engn, Jinan 250061, Peoples R China
[2] Jinan Cent Hosp, Dept Stomatol, Jinan 250013, Peoples R China
[3] Beihang Univ, Sch Aeronaut Sci & Engn, Beijing 100191, Peoples R China
关键词
ACUTE PULMONARY-EMBOLISM; BLOOD-FLOW; HEMOSTASIS; SIMULATION; MANAGEMENT;
D O I
10.1063/5.0264241
中图分类号
O3 [力学];
学科分类号
08 ; 0801 ;
摘要
The placement of an inferior vena cava filter (IVCF) has been an effective treatment to prevent fatal pulmonary embolism in clinical practice for decades. Fusiform-shaped filter is one of the main types of IVCFs, but its hemodynamics with thrombus capture has not been clearly understood. In this paper, computational fluid dynamics (CFD) simulations are used to investigate the characteristics of blood flow with thrombi over TrapEase and OptEase filters, respectively, two frequently used fusiform-shaped IVCFs in clinic therapy. The blood flow with thrombi is characterized by the Eulerian two-phase flow model, in which the blood is regarded as a non-Newtonian fluid. CFD results show that there are significant stagnation zones downstream of both filters, with the lengths nearly up to 12 times the diameter of the IVC. The no-slip wall conditions of filter wires induce a viscous block effect and result in blood flow acceleration inside the fusiform-shaped filter. The viscous drag contributes predominantly to the total flow resistance of fusiform-shaped filter, about three times the pressure drag. The presence of thrombi does not greatly affect the hemodynamics, but reduces the nearby blood flow speed when captured by the filter. The thrombi primarily accumulate at the end of the filter, increasing both viscous and pressure drag, leading to a 10%-12% increase in total flow resistance. The present study provides common insights into the hemodynamics with thrombus capture of the fusiform-shaped filter, which would contribute to its future clinical use and design improvement.
引用
收藏
页数:15
相关论文
共 56 条
[1]   A longitudinal comparison of hemodynamics and intraluminal thrombus deposition in abdominal aortic aneurysms [J].
Arzani, Amirhossein ;
Suh, Ga-Young ;
Dalman, Ronald L. ;
Shadden, Shawn C. .
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 2014, 307 (12) :H1786-H1795
[2]   Computational predictions of the embolus-trapping performance of an IVC filter in patient-specific and idealized IVC geometries [J].
Aycock, Kenneth I. ;
Campbell, Robert L. ;
Lynch, Frank C. ;
Manning, Keefe B. ;
Craven, Brent A. .
BIOMECHANICS AND MODELING IN MECHANOBIOLOGY, 2017, 16 (06) :1957-1969
[3]   The Importance of Hemorheology and Patient Anatomy on the Hemodynamics in the Inferior Vena Cava [J].
Aycock, Kenneth I. ;
Campbell, Robert L. ;
Lynch, Frank C. ;
Manning, Keefe B. ;
Craven, Brent A. .
ANNALS OF BIOMEDICAL ENGINEERING, 2016, 44 (12) :3568-3582
[4]   A Computational Method for Predicting Inferior Vena Cava Filter Performance on a Patient-Specific Basis [J].
Aycock, Kenneth I. ;
Campbell, Robert L. ;
Manning, Keefe B. ;
Sastry, Shankar P. ;
Shontz, Suzanne M. ;
Lynch, Frank C. ;
Craven, Brent A. .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2014, 136 (08)
[5]   Inferior Vena Cava Filters and Complications: A Systematic Review [J].
Bajda, Joe ;
Park, Ann N. ;
Raj, Aishwarya ;
Raj, Rhea ;
Gorantla, Vasavi Rakesh .
CUREUS JOURNAL OF MEDICAL SCIENCE, 2023, 15 (06)
[6]   Effects of reverse deployment of cone-shaped vena cava filter on improvements in hemodynamic performance in vena cava [J].
Chen, Ying ;
Xu, Zaipin ;
Deng, Xiaoyan ;
Yang, Shibo ;
Tan, Wenchang ;
Fan, Yubo ;
Han, Yong ;
Xing, Yubin .
BIOMEDICAL ENGINEERING ONLINE, 2021, 20 (01)
[7]   Modeling of coarctation of aorta in human fetuses using 3D/4D fetal echocardiography and computational fluid dynamics [J].
Chen, Zhuo ;
Zhou, Yue ;
Wang, Jingying ;
Liu, Xiaowei ;
Ge, Shuping ;
He, Yihua .
ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2017, 34 (12) :1858-1866
[8]   Inferior vena caval hemodynamics quantified in vivo at rest and during cycling exercise using magnetic resonance imaging [J].
Cheng, CP ;
Herfkens, RJ ;
Taylor, CA .
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 2003, 284 (04) :H1161-H1167
[9]  
CHO YI, 1991, BIORHEOLOGY, V28, P241
[10]   Steady Flow in a Patient-Averaged Inferior Vena Cava-Part II: Computational Fluid Dynamics Verification and Validation [J].
Craven, Brent A. ;
Aycock, Kenneth I. ;
Manning, Keefe B. .
CARDIOVASCULAR ENGINEERING AND TECHNOLOGY, 2018, 9 (04) :654-673