Bias Calibration of Optically Pumped Magnetometers Based on Variable Sensitivity

被引:0
|
作者
Chen, Jieya [1 ,2 ]
Ye, Chaofeng [2 ]
Hou, Xingshen [2 ]
Niu, Yaqiong [2 ]
Sun, Limin [1 ]
机构
[1] Chinese Acad Sci, Shanghai Inst Microsyst & Informat Technol, Shanghai 200050, Peoples R China
[2] ShanghaiTech Univ, Sch Informat Sci & Technol, Shanghai 201210, Peoples R China
基金
国家重点研发计划;
关键词
OPM; bias calibration; biomagnetic field measurement; magnetic field compensation; DESIGN; MAGNETOENCEPHALOGRAPHY; COILS;
D O I
10.3390/s25020433
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Optically pumped magnetometers (OPMs) functioning in the spin-exchange relaxation-free (SERF) regime have emerged as attractive options for measuring weak magnetic fields, owing to their portability and remarkable sensitivity. The operation of SERF-OPM critically relies on the ambient magnetic field; thus, a magnetic field compensation device is commonly employed to mitigate the ambient magnetic field to near zero. Nonetheless, the bias of the OPM may influence the compensation impact, a subject that remains unexamined. This paper introduced an innovative bias calibration technique for OPMs. The sensitivity of the OPM was altered by adjusting the cell temperature. The output of the OPM was then documented with varying sensitivity. It is assumed that the signal exhibits a linear correlation with the environmental magnetic field, and the statistical characteristics of the magnetic field are identical for both measurements, upon which the bias of the OPM is assessed. The bias was subsequently considered in the feedback magnetic field compensation mechanism. The results indicate that this technique might successfully reduce environmental magnetic fluctuations and enhance the sensitivity of the OPM. This technique measured the magnetic field produced by the human heart, confirming the viability of the ultra-weak biomagnetic field measurement approach.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Improving the sensitivity of optically pumped magnetometers by hyperfine repumping
    Schultze, V.
    Scholtes, T.
    IJsselsteijn, R.
    Meyer, H. -G.
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA B-OPTICAL PHYSICS, 2015, 32 (05) : 730 - 736
  • [2] Dependence on sensitivity patterns of bias fields for vector measurements using optically pumped vapor magnetometers
    Lou, Janet W.
    Cranch, Geoffrey A.
    AIP ADVANCES, 2020, 10 (10)
  • [3] Calibration and Localization of Optically Pumped Magnetometers Using Electromagnetic Coils
    Iivanainen, Joonas
    Borna, Amir
    Zetter, Rasmus
    Carter, Tony R.
    Stephen, Julia M.
    McKay, Jim
    Parkkonen, Lauri
    Taulu, Samu
    Schwindt, Peter D. D.
    SENSORS, 2022, 22 (08)
  • [4] Sensitivity studies and optimization of arrangements of optically pumped magnetometers in simulated magnetoencephalography
    Eichardt, Roland
    Strohmeier, Daniel
    Hunold, Alexander
    Machts, Rene
    Haueisen, Jens
    Oelsner, Gregor
    Schmidt, Christian B.
    Schultze, Volkmar
    Stolz, Ronny
    Graichen, Uwe
    COMPEL-THE INTERNATIONAL JOURNAL FOR COMPUTATION AND MATHEMATICS IN ELECTRICAL AND ELECTRONIC ENGINEERING, 2019, 38 (03) : 953 - 964
  • [5] Microfabricated Optically-Pumped Magnetometers
    Knappe, Svenja
    2014 CONFERENCE ON LASERS AND ELECTRO-OPTICS (CLEO), 2014,
  • [6] In situ calibration of triaxial coils of a vector optically pumped magnetometers based on a particle swarm optimization algorithm
    Li, Siran
    Ma, Danyue
    Lu, Jixi
    Xing, Bozheng
    Wang, Kun
    Gao, Yanan
    Han, Bangcheng
    MEASUREMENT, 2022, 202
  • [7] Recent progress in optically pumped magnetometers
    Alexandrov, EB
    PHYSICA SCRIPTA, 2003, T105 : 27 - 30
  • [8] Floquet description of optically pumped magnetometers
    Florez, Hans Marin
    Pyragius, Tadas
    PHYSICAL REVIEW A, 2021, 103 (03)
  • [9] VCSEL frequency stabilization for optically pumped magnetometers
    闫业广
    刘刚
    林海笑
    尹凯峰
    王锟
    陆吉玺
    ChineseOpticsLetters, 2021, 19 (12) : 69 - 74
  • [10] VCSEL frequency stabilization for optically pumped magnetometers
    Yan, Yeguang
    Liu, Gang
    Lin, Haixiao
    Yin, Kaifeng
    Wang, Kun
    Lu, Jixi
    CHINESE OPTICS LETTERS, 2021, 19 (12)