Predicted SAR/temperature changes induced by phase-amplitude steering are minimally affected by uncertainties in tissue properties: a basis for robust on-line adaptive hyperthermia treatment planning

被引:0
作者
Kok, H. P. [1 ,2 ]
Crezee, J. [1 ,2 ]
机构
[1] Amsterdam UMC Locat Univ Amsterdam, Radiat Oncol, Amsterdam, Netherlands
[2] Canc Ctr Amsterdam, Treatment & Qual Life, Canc Biol & Immunol, Amsterdam, Netherlands
关键词
Locoregional hyperthermia; hyperthermia treatment planning; Plan2Heat; Adapt2Heat; RF heating; on-line adaptive planning; REGIONAL HYPERTHERMIA; DEEP HYPERTHERMIA; CERVICAL-CANCER; LOCOREGIONAL HYPERTHERMIA; LOCAL HYPERTHERMIA; BLOOD-FLOW; OPTIMIZATION; SAR; DISTRIBUTIONS; APPLICATOR;
D O I
10.1080/02656736.2025.2483433
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
BackgroundReliability of absolute specific absorption rate (SAR)/temperature levels predicted by treatment planning is strongly affected by tissue parameter uncertainties. Therefore, regular re-optimization to suppress hot spots can accidentally induce new hot spots elsewhere. Adaptive planning methods to avoid this problem re-optimize with respect to the current predicted 3D-distribution. This strategy is robust if reliability of predicted SAR/temperature changes (i.e., increases/decreases) after phase-amplitude adjustments is minimally affected by parameter uncertainties; this work evaluated this robustness.MethodsWe validated the basic concept in an inhomogeneous phantom, followed by a patient model. Uncertainties in electrical conductivity, permittivity and perfusion were mimicked by simulations using 100 random parameter samples from normal distributions. Reliability of predicted SAR/temperature increase/decrease after phase-amplitude adjustments was evaluated. Next, correlations between measured and simulated SAR and SAR changes were determined for phase settings evaluated at the treatment start for a treatment series. Finally, practical use in an adaptive workflow was illustrated.ResultsLocal SAR/temperature increases/decreases after phase-amplitude adjustments can be predicted accurately. For the phantom, the measured 28.5% SAR decrease was predicted accurately(28.5 +/- 0.7%). In the patient model, predicted SAR/temperature changes were typically accurate within a few percent. For the treatment series, correlations between measured and simulated (relative) SAR changes were much better(R2=0.70-0.82) than for absolute SAR levels(R2=0.29). Predictions of steering effects during treatment corresponded qualitatively with measurements/observations.ConclusionPredictions of SAR/temperature increases/decreases induced by phase-amplitude steering are hardly affected by tissue parameter uncertainties. On-line adaptive planning based on predicted changes is thus robust to effectively support clinical steering strategies.
引用
收藏
页数:17
相关论文
共 81 条
[1]   Clinical Evidence for Thermometric Parameters to Guide Hyperthermia Treatment [J].
Ademaj, Adela ;
Veltsista, Danai P. ;
Ghadjar, Pirus ;
Marder, Dietmar ;
Oberacker, Eva ;
Ott, Oliver J. ;
Wust, Peter ;
Puric, Emsad ;
Haelg, Roger A. ;
Rogers, Susanne ;
Bodis, Stephan ;
Fietkau, Rainer ;
Crezee, Hans ;
Riesterer, Oliver .
CANCERS, 2022, 14 (03)
[2]   Hyperthermia treatment planning for cervical cancer patients based on electrical conductivity tissue properties acquired in vivo with EPT at 3 T MRI [J].
Balidemaj, Edmond ;
Kok, Henny Petra ;
Schooneveldt, Gerben ;
van Lier, Astrid L. H. M. W. ;
Remis, Rob F. ;
Stalpers, Lukas J. A. ;
Westerveld, Henrike ;
Nederveen, Aart J. ;
van den Berg, Cornelis A. T. ;
Crezee, Johannes .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2016, 32 (05) :558-568
[3]   SAR OPTIMIZATION IN A PHASED-ARRAY RADIOFREQUENCY HYPERTHERMIA SYSTEM [J].
BARDATI, F ;
BORRANI, A ;
GERARDINO, A ;
LOVISOLO, GA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1995, 42 (12) :1201-1207
[4]  
Bellizzi GG, 2017, IEEE J ELECTROMAG RF, V1, P74, DOI 10.1109/JERM.2017.2766569
[5]   Field and Temperature Shaping for Microwave Hyperthermia: Recent Treatment Planning Tools to Enhance SAR-Based Procedures [J].
Bevacqua, Martina T. ;
Gaffoglio, Rossella ;
Bellizzi, Gennaro G. ;
Righero, Marco ;
Giordanengo, Giorgio ;
Crocco, Lorenzo ;
Vecchi, Giuseppe ;
Isernia, Tommaso .
CANCERS, 2023, 15 (05)
[6]  
Bruggmoser G, 2012, STRAHLENTHER ONKOL, V188, P198, DOI 10.1007/s00066-012-0176-2
[7]   Complaint-adaptive power density optimization as a tool for HTP-guided steering in deep hyperthermia treatment of pelvic tumors [J].
Canters, R. A. M. ;
Franckena, M. ;
van der Zee, J. ;
Van Rhoon, G. C. .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (23) :6799-6820
[8]  
Canters RAM, 2013, STRAHLENTHER ONKOL, V189, P74, DOI 10.1007/s00066-012-0241-x
[9]   Optimizing deep hyperthermia treatments: are locations of patient pain complaints correlated with modelled SAR peak locations? [J].
Canters, R. A. M. ;
Franckena, M. ;
van der Zee, J. ;
van Rhoon, G. C. .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (02) :439-451
[10]   A literature survey on indicators for characterisation and optimisation of SAR distributions in deep hyperthermia, a plea for standardisation [J].
Canters, R. A. M. ;
Wust, P. ;
Bakker, J. F. ;
Van Rhoon, G. C. .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2009, 25 (07) :593-608