Device-Less Data-Driven Cardiac and Respiratory Gating Using LAFOV PET Histo Images

被引:1
作者
Overbeck, Nanna [1 ]
Andersen, Thomas Lund [1 ,2 ]
Rodell, Anders Bertil [3 ]
Cabello, Jorge [4 ]
Birge, Noah [4 ]
Schleyer, Paul [4 ]
Conti, Maurizio [4 ]
Korsholm, Kirsten [1 ]
Fischer, Barbara Malene [1 ,2 ]
Andersen, Flemming Littrup [1 ,2 ]
Lindberg, Ulrich [1 ]
机构
[1] Copenhagen Univ Hosp, Dept Clin Physiol & Nucl Med, Rigshosp, DK-2100 Copenhagen, Denmark
[2] Univ Copenhagen, Dept Clin Med, DK-2100 Copenhagen, Denmark
[3] Siemens Healthcare AS, DK-8200 Aarhus, Denmark
[4] Siemens Med Solut USA Inc, Knoxville, TN 37932 USA
关键词
data-driven gating; image-derived frequency; gating; motion correction; respiration; cardiac; histo images; TOF; LIST-MODE; SIGNAL; RECONSTRUCTION; 3D;
D O I
10.3390/diagnostics14182055
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: The outstanding capabilities of modern Positron Emission Tomography (PET) to highlight small tumor lesions and provide pathological function assessment are at peril from image quality degradation caused by respiratory and cardiac motion. However, the advent of the long axial field-of-view (LAFOV) scanners with increased sensitivity, alongside the precise time-of-flight (TOF) of modern PET systems, enables the acquisition of ultrafast time resolution images, which can be used for estimating and correcting the cyclic motion. Methods: 0.25 s so-called [18F]FDG PET histo image series were generated in the scope of for detecting respiratory and cardiac frequency estimates applicable for performing device-less data-driven gated image reconstructions. The frequencies of the cardiac and respiratory motion were estimated for 18 patients using Short Time Fourier Transform (STFT) with 20 s and 30 s window segments, respectively. Results: The Fourier analysis provided time points usable as input to the gated reconstruction based on eight equally spaced time gates. The cardiac investigations showed estimates in accordance with the measured pulse oximeter references (p = 0.97) and a mean absolute difference of 0.4 +/- 0.3 beats per minute (bpm). The respiratory frequencies were within the expected range of 10-20 respirations per minute (rpm) in 16 out of 18 patients. Using this setup, the analysis of three patients with visible lung tumors showed an increase in tumor SUVmax and a decrease in tumor volume compared to the non-gated reconstructed image. Conclusions: The method can provide signals that were applicable for gated reconstruction of both cardiac and respiratory motion, providing a potential increased diagnostic accuracy.
引用
收藏
页数:12
相关论文
共 39 条
  • [1] Assessment of motion correction in dynamic rubidium-82 cardiac PET with and without frame-by-frame adjustment of attenuation maps for calculation of myocardial blood flow
    Armstrong, Ian S.
    Memmott, Matthew J.
    Saint, Kimberley J.
    Saillant, Antoine
    Hayden, Charles
    Arumugam, Parthiban
    [J]. JOURNAL OF NUCLEAR CARDIOLOGY, 2021, 28 (04) : 1334 - 1346
  • [2] Clinical Evaluation of a Data-Driven Respiratory Gating Algorithm for Whole-Body PET with Continuous Bed Motion
    Buether, Florian
    Jones, Judson
    Seifert, Robert
    Stegger, Lars
    Schleyer, Paul
    Schaefers, Michael
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2020, 61 (10) : 1520 - 1527
  • [3] Data-driven gating in PET: Influence of respiratory signal noise on motion resolution
    Buether, Florian
    Ernst, Iris
    Frohwein, Lynn Johann
    Pouw, Joost
    Schaefers, Klaus Peter
    Stegger, Lars
    [J]. MEDICAL PHYSICS, 2018, 45 (07) : 3205 - 3213
  • [4] Detection of respiratory tumour motion using intrinsic list mode-driven gating in positron emission tomography
    Buether, Florian
    Ernst, Iris
    Dawood, Mohammad
    Kraxner, Peter
    Schaefers, Michael
    Schober, Otmar
    Schaefers, Klaus P.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2010, 37 (12) : 2315 - 2327
  • [5] Image Quality and Quantitative PET Parameters of Low-Dose [18F]FDG PET in a Long Axial Field-of-View PET/CT Scanner
    Calderon, Eduardo
    Schmidt, Fabian P.
    Lan, Wenhong
    Castaneda-Vega, Salvador
    Brendlin, Andreas S.
    Trautwein, Nils F.
    Dittmann, Helmut
    la Fougere, Christian
    Kiefer, Lena Sophie
    [J]. DIAGNOSTICS, 2023, 13 (20)
  • [6] Focus on time-of-flight PET: the benefits of improved time resolution
    Conti, Maurizio
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2011, 38 (06) : 1147 - 1157
  • [7] Comparison of List-Mode and DIRECT Approaches for Time-of-Flight PET Reconstruction
    Daube-Witherspoon, Margaret E.
    Matej, Samuel
    Werner, Matthew E.
    Surti, Suleman
    Karp, Joel S.
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2012, 31 (07) : 1461 - 1471
  • [8] Feng T., 2020, P 2020 IEEE NUCL SCI, DOI [10.1109/NSS/MIC42677.2020.9508106, DOI 10.1109/NSS/MIC42677.2020.9508106]
  • [9] Value of 4-Dimensional F-18-FDG PET/CT in the Classification of Pulmonary Lesions
    Garcia Vicente, Ana Maria
    Soriano Castrejon, Angel
    Leon Martin, Antonio Alberto
    Gonzalez Garcia, Beatriz
    Pilkington Woll, John Patrick
    Palomar Munoz, Azahara
    [J]. JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY, 2011, 39 (02) : 91 - 99
  • [10] Respiratory gated PET/CT in a European multicentre retrospective study: added diagnostic value in detection and characterization of lung lesions
    Guerra, Luca
    De Ponti, Elena
    Elisei, Federica
    Bettinardi, Valentino
    Landoni, Claudio
    Picchio, Maria
    Gilardi, Maria Carla
    Versari, Annibale
    Fioroni, Federica
    Dziuk, Miroslaw
    Koza, Magdalena
    Ahond-Vionnet, Renee
    Collin, Bertrand
    Messa, Cristina
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2012, 39 (09) : 1381 - 1390