Outcomes of proximal interphalangeal joint arthroplasty with the MatOrtho prosthesis after a minimum of 10 years

被引:0
作者
Sigamoney, Kohila Vani [1 ]
Shields, David W. [1 ]
Gillott, Elizabeth [1 ]
Boland, Katy [1 ]
Ricks, Matthew [1 ]
Talwalkar, Sumedh [1 ]
Trail, Ian [1 ]
机构
[1] Wrightington Hosp, Hall Lane, Appley Bridge, Wigan WN6 9EP, Lancs, England
关键词
Arthritis; arthroplasty; finger; interphalangeal; replacement; HAND; RELIABILITY; VALIDITY;
D O I
10.1177/17531934251313996
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
This is a prospective case review on 100 MatOrtho proximal interphalangeal joint arthroplasties in 50 patients with previously published 2-year outcomes. We report the outcomes at a minimum of 10-years' follow-up. Sixty-three implants in 29 patients were available for analysis. The visual analogue scale score for pain revealed a significant improvement of pain at 10 years (mean 0.7, SD 1.3) compared with preoperatively (mean 4.3, SD 0.8) and at 2 years (mean 1.1, SD 1.8). There was no statistically significant improvement at 10 years compared with preoperatively in the mean grip strength (10, SD 6.7), range of motion (27 degrees, SD 20.1,), patient evaluation measure score (43, SD 22), Quick Disabilities of the Arm, Shoulder and Hand (38, SD 23) and Michigan Hand Outcome Questionnaire (65, SD 9,), but all patients remained satisfied with the surgery. Survivorship was 82% at 10 years. We conclude that this implant is suitable to use mainly for pain relief.Level of Evidence: IV
引用
收藏
页数:6
相关论文
共 22 条
[1]   Age-related change in handgrip strength in men and women: is muscle quality a contributing factor? [J].
Abe, Takashi ;
Thiebaud, Robert S. ;
Loenneke, Jeremy P. .
AGE, 2016, 38 (01) :1-7
[2]   FLEXIBLE IMPLANT RESECTION ARTHROPLASTY OF THE PROXIMAL INTERPHALANGEAL JOINT IN PATIENTS WITH SYSTEMIC INFLAMMATORY ARTHRITIS [J].
ADAMSON, GJ ;
GELLMAN, H ;
BRUMFIELD, RH ;
KUSCHNER, SH ;
LAWLER, JW .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 1994, 19A (03) :378-384
[3]  
Ashworth CR, 1997, CLIN ORTHOP RELAT R, P34
[4]   Long-term survivorship of total knee arthroplasty with a single-radius, high-flexion posterior stabilized prosthesis [J].
Baek, Ji-Hoon ;
Lee, Su Chan ;
Choi, Kyungwon ;
Ahn, Hye Sun ;
Nam, Chang Hyun .
KNEE, 2021, 30 :275-282
[5]   Ling-Term Results of Swanson Silicone Arthroplasty for Proximal Interphalangeal Joint Osteoarthritis [J].
Bales, Joshua G. ;
Wall, Lindley B. ;
Stern, Peter J. .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2014, 39 (03) :455-461
[6]  
Beaton D E, 2001, J Hand Ther, V14, P128
[7]   Development of the QuickDASH: Comparison of three item-reduction approaches [J].
Beaton, DE ;
Wright, JG ;
Katz, JN .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (05) :1038-1046
[8]   Reliability and validity testing of the Michigan Hand Outcomes Questionnaire [J].
Chung, KC ;
Pillsbury, MS ;
Walters, MR ;
Hayward, RA .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 1998, 23A (04) :575-587
[9]   Which questionnaire is best? The reliability validity and ease of use of the patient evaluation measure, the disabilities of the arm, shoulder and hand and the Michigan hand outcome measure [J].
Dias, J. J. ;
Rajan, R. A. ;
Thompson, J. R. .
JOURNAL OF HAND SURGERY-EUROPEAN VOLUME, 2008, 33E (01) :9-17
[10]   MatOrtho proximal interphalangeal joint arthroplasty: minimum 2-year follow-up [J].
Flannery, O. ;
Harley, O. ;
Badge, R. ;
Birch, A. ;
Nuttall, D. ;
Trail, I. A. .
JOURNAL OF HAND SURGERY-EUROPEAN VOLUME, 2016, 41 (09) :910-916