Impact of a Prototype 29:1 Ratio Grid on Image Quality and Radiation Dose in Abdominal Angiography Evaluation in a Pig Model

被引:0
作者
Werncke, Thomas [1 ]
Becker, Lena S. [2 ]
Maschke, Sabine K. [1 ]
Bruesch, Inga [3 ]
Rumpel, Regina [3 ]
Wacker, Frank K. [1 ]
Meyer, Bernhard C. [1 ]
机构
[1] Hannover Med Sch, Dept Diagnost & Intervent Radiol, Carl Neuberg Str 1, D-30625 Hannover, Germany
[2] Yale Sch Med, Dept Radiol & Biomed Imaging, Sect Intervent Radiol, New Haven, CT USA
[3] Hannover Med Sch, Inst Lab Anim Sci & Cent Anim Facil, Hannover, Germany
关键词
interventional radiology; animal study; antiscatter grid; image quality; radiation dose; fluoroscopy;
D O I
10.1097/RLI.0000000000001136
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives The aim of this study was to evaluate the impact of a prototype grid with a 29:1 ratio (r29) and a 15:1 (r15) grid on the image quality (IQ) and radiation dose in abdominal angiography. Materials and Methods Six typical abdominal angiographic image scenarios were created in 4 pigs. Polymethylmethacrylate and aluminum plates were used to add 10 cm of patient equivalent thickness to simulate different body types. Fluoroscopic images were acquired with a source-to-image receptor distance of 120 cm. Tantalum- and iron-specific acquisition protocols at different IQ levels were acquired. IQ of radiation dose equivalent image pairs, created with the r29 and r15 grids, respectively, was quantitatively evaluated using contrast-to-noise ratio (CNR) measurements. Differences in radiation dose were estimated using the dose-weighted CNR. Two blinded readers compared IQ of these images using a Likert scale. In a second step, the readers selected pairs of the r29 and r15 images with subjectively equivalent IQ. Radiation doses were then compared. Results Compared with the r15 grid, the r29 grid images achieved similar CNR at an average of 26% (+/- 12%) lower radiation dose at a mean patient equivalent thickness of 26 cm and 36 cm. Both readers noted a significant increase in IQ (P < 0.001) for dose equivalent images, whereas the interobserver agreement was 0.59. For the selected IQ equivalent images, a radiation dose reduction of 38% (+/- 17%; P < 0.001, interobserver agreement 0.92) was noted when using the r29 grid. Conclusions The use of an r29 grid at a large source-to-image receptor distance can significantly improve the IQ compared with the r15 grid at the same radiation dose in abdominal angiography or can reduce radiation dose while preserving IQ.
引用
收藏
页码:349 / 355
页数:7
相关论文
共 22 条
[1]  
[Anonymous], 1993, Physical principles of medical imaging
[2]  
Bernhardt P., Device and Method for Computer Tomography, Patent No. [US-7260169-B2, 7260169]
[3]  
Cicchetti DV., 1994, PSYCHOL ASSESSMENTS, V6, P284, DOI [DOI 10.1037/1040-3590.6.4.284, 10.1037/1040-3590.6.4.284]
[4]   A Comparison of Reliability Coefficients for Ordinal Rating Scales [J].
de Raadt, Alexandra ;
Warrens, Matthijs J. ;
Bosker, Roel J. ;
Kiers, Henk A. L. .
JOURNAL OF CLASSIFICATION, 2021, 38 (03) :519-543
[5]  
Denham B.E., 2017, Categorical Statistics for Communication Research, P232, DOI [DOI 10.1002/9781119407201.CH11, 10.1002/9781119407201.ch]
[6]   Influence of anatomical structure on antiscatter grid performance in 2D: application to x-ray angiography and a prototype 29:1 ratio grid [J].
Fetterly, Kenneth ;
Bernhardt, Philipp ;
Schueler, Beth .
PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (14)
[7]   Technical evaluation of a prototype ratio 29:1 grid for adult patient cardiovascular angiography imaging conditions [J].
Fetterly, Kenneth A. ;
Schueler, Beth A. ;
Hindal, Mark D. ;
Miller, Daniel L. .
PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (14)
[8]  
Gosch D., 2007, Rofo, V179, p1035 1042
[9]  
IEC, 2007, DIN EN 61223-2-6:2008-12
[10]   Report of AAPM Task Group 272: Comprehensive acceptance testing and evaluation of fluoroscopy imaging systems [J].
Lin, Pei-Jan Paul ;
Goode, Allen R. ;
Corwin, Frank D. ;
Fisher, Ryan F. ;
Balter, Stephen ;
Wunderle, Kevin A. ;
Schueler, Beth A. ;
Kim, Don-Soo ;
Zhang, Jie ;
Zhou, Yifang ;
Jenkins, Peter A. ;
Mahmood, Usman ;
Lin, Teh ;
Zhao, Hui ;
Park, Mi-Ae ;
Trianni, Annalisa ;
Lendle, Markus ;
Kuhls-Gilcrist, Andrew ;
Jans, Jan C. ;
Desponds, Lionel ;
Banasiak, Gene ;
Backes, Steve ;
Snyder, Carl ;
Snyder, Angela ;
Lu, Minghui ;
Gonzalez, Scott .
MEDICAL PHYSICS, 2022, 49 (04) :E1-E49