Sagittal accuracy and functional impact of tibial slope in imageless robotic-assisted Total Knee Arthroplasty

被引:0
|
作者
Eggermont, Erwan [1 ]
Janssens, Raphael [2 ]
Ulrix, Maarten [1 ]
Fils, Jean-Francois [3 ]
Hernigou, Jacques [4 ,5 ]
Everaert, Johnatan [1 ]
Baillon, Bruno [1 ,2 ]
机构
[1] Univ libre Bruxelles, Brussels, Belgium
[2] Iris South Hosp Grp, Dept Orthopaed Surg, Brussels, Belgium
[3] Ars Stat SRL, Nivelles, Belgium
[4] Ath Hosp, Orthopaed Dept, EPICURA Baudour, Nivelles, Hainaut, Belgium
[5] Univ Mons, Mons, Belgium
关键词
Posterior tibial slope; Robotic-assisted; Total knee arthroplasty; Sagittal accuracy; OKS; KOOS; ROBUST ESTIMATION; MECHANICAL AXIS; ALIGNMENT; ANGLE;
D O I
10.1007/s00264-025-06472-w
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
PurposeStudy of the sagittal accuracy of the 'Robotic Surgical Assistant' (ROSA (R)), compared to conventional surgery, regarding the application of the tibial slope (TS). Study of the impact of TS on the range of motion (ROM) and patient-reported outcome measures (PROMS).MethodsInclusion of patients who underwent primary Total Knee Arthroplasty (TKA) between 1/1/2021 and 15/4/2024. Divided into robotic-assisted TKA (RA-TKA) and manual TKA (M-TKA). Measurement of pre- and post-operative TS, using the posterior tibial cortex, on profile knee X-rays. 3 degrees TS applied arbitrarily for both groups. ROM was measured pre-operatively and at three, six and 12 months post-operatively. Patient satisfaction assessed via Knee Injury and Osteoarthritis Outcome Score (KOOS) and Oxford Knee Score (OKS).Results266 patients were included in the study. The M-TKA (110) had a post-operative TS of 3.11 degrees (+/- 2.12 degrees). 81.21% were within 2 degrees of the target and 92.87% within 3 degrees. The RA-TKA (82) had a post-operative TS of -0.11 degrees +/- (1.93 degrees). 36.83% were within 2 degrees of the target and 56.63% within 3 degrees. RA-TKA had a KOOS of 64.43 +/- 12.87 and OKS of 33.05 +/- 6.01. M-TKA had a KOOS of 64.18 +/- 13.11 and OKS of 32.31 +/- 5.97. Maximum flexion at 12 months was 118.74 degrees +/- 8.19 degrees for M-TKA and 121.88 degrees +/- 7.43 degrees for RA-TKA (p = 0.002).ConclusionThe application of TS using ROSA (R) was less precise than the conventional method in achieving post-operative TS values as measured on X-rays. However, there were no clinical differences in ROM or PROMS.
引用
收藏
页码:1387 / 1394
页数:8
相关论文
共 50 条
  • [1] Accuracy and learning curve of imageless robotic-assisted total knee arthroplasty
    Bosco, Francesco
    Rovere, Giuseppe
    Burgio, Carmelo
    Bue, Giorgia Lo
    Cobisi, Claudio Domenico
    Via, Riccardo Giai
    Lucenti, Ludovico
    Camarda, Lawrence
    JOURNAL OF ORTHOPAEDICS, 2025, 66 : 77 - 83
  • [2] The Accuracy and Clinical Success of Robotic-Assisted Total Knee Arthroplasty
    Nickel, Brian T.
    Carroll, Kaitlin M.
    Pearle, Andrew D.
    Kleeblad, Laura J.
    Burger, Joost
    Mayman, David J.
    Westrich, Geoffrey
    Jerabek, Seth A.
    HSS JOURNAL, 2021, 17 (03) : 261 - 266
  • [3] Individualized Functional Knee Alignment in Total Knee Arthroplasty: A Robotic-assisted Technique
    Clark, Gavin W.
    Esposito, Christina, I
    Wood, David
    TECHNIQUES IN ORTHOPAEDICS, 2022, 37 (03) : 185 - 191
  • [4] Learning curve for imageless robotic-assisted total knee arthroplasty in non-fellowship trained joint replacement surgeons
    Stegelmann, Samuel D.
    Butler, Justin
    Eaddy, Samuel G.
    Davis, Trent
    Davis, Kirk
    Miller, Richard
    JOURNAL OF ORTHOPAEDICS, 2023, 45 : 72 - 77
  • [5] Robotic-Assisted Total Knee Arthroplasty in Obese Patients
    Richardson, Mary K.
    Digiovanni, Ryan M.
    McCrae, Brian K.
    Cooperman, Wesley S.
    Ludington, John
    Heckmann, Nathanael D.
    Oakes, Daniel A.
    ARTHROPLASTY TODAY, 2024, 26
  • [6] Robotic-assisted knee arthroplasty
    Banerjee, Samik
    Cherian, Jeffrey J.
    Elmallah, Randa K.
    Jauregui, Julio J.
    Pierce, Todd P.
    Mont, Michael A.
    EXPERT REVIEW OF MEDICAL DEVICES, 2015, 12 (06) : 727 - 735
  • [7] Functional alignment in robotic-assisted total knee arthroplasty: a systematic review
    Filippo Migliorini
    Marco Pilone
    Luise Schäfer
    Francesco Simeone
    Andreas Bell
    Nicola Maffulli
    Archives of Orthopaedic and Trauma Surgery, 2024, 144 : 1741 - 1749
  • [8] Functional alignment in robotic-assisted total knee arthroplasty: a systematic review
    Migliorini, Filippo
    Pilone, Marco
    Schaefer, Luise
    Simeone, Francesco
    Bell, Andreas
    Maffulli, Nicola
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2024, 144 (04) : 1741 - 1749
  • [9] Robotic-assisted total knee arthroplasty is comparable to conventional total knee arthroplasty: a meta-analysis and systematic review
    Onggo, James Randolph
    Onggo, Jason Derry
    De Steiger, Richard
    Hau, Raphael
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2020, 140 (10) : 1533 - 1549
  • [10] Evaluating the cost of robotic-assisted total and unicompartmental knee arthroplasty
    Alexander, Kate
    Karunaratne, Sascha
    Sidhu, Verinder
    Fritsch, Brett
    Gupta, Sanjeev
    Horsley, Mark
    Guzman, Maurice
    Boyle, Richard
    Mcbride, Kate
    Steffens, Daniel
    JOURNAL OF ROBOTIC SURGERY, 2024, 18 (01)