Evaluation of non-invasive sensors for monitoring core temperature

被引:0
作者
Thomas, Shavin S. [1 ]
Flickinger, Katharyn L. [2 ]
Elmer, Jonathan [2 ,3 ,4 ]
Callaway, Clifton W. [2 ]
机构
[1] Univ Pittsburgh, Sch Med, Pittsburgh, PA USA
[2] Univ Pittsburgh, Sch Med, Dept Emergency Med, Iroquois Bldg,Suite 400A,3600 Forbes Ave, Pittsburgh, PA 15213 USA
[3] Univ Pittsburgh, Sch Med, Dept Crit Care Med, Pittsburgh, PA 15213 USA
[4] Univ Pittsburgh, Sch Med, Dept Neurol, Pittsburgh, PA 15213 USA
关键词
Monitoring; Temperature; Non-invasive; Cardiac arrest; AGREEMENT;
D O I
10.1007/s10877-025-01289-9
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We evaluated the accuracy and precision of zero-heat flux (ZHF) and dual sensor (DS) non-invasive temperature probes in intensive care unit (ICU) patients undergoing hypothermic temperature control, hypothesizing that both devices would accurately estimate core temperature. In a single-center prospective cohort study, we enrolled 35 ICU patients and applied continuous, non-invasive ZHF and/or DS probes to the lateral forehead or anterior chest to collect 358 observations. Conditions potentially influencing temperature estimation were recorded. Using Bland-Altman analysis with multiple paired observations per individual, we compared the bias between non-invasive probes and direct core temperature measurements. Lin's concordance coefficient (LCC) was computed to quantify precision. The mean bias between the ZHF probe and invasive temperature was + 0.98 degrees C; for the DS probe, it was - 2.19 degrees C. In hypothermic patients, the ZHF probe's accuracy improved (bias + 0.28 degrees C, LCC 0.86), while the DS probe remained inaccurate (bias - 2.52 degrees C, LCC 0.07). Clinical confounders like vasoactive agents or temperature control devices did not consistently affect bias, accuracy, or precision. Neither the ZHF nor DS non-invasive probes provided sufficient accuracy or precision to guide clinical decisions in the ICU. These results contrast with previous studies reporting biases within +/- 0.5 degrees C. However, the ZHF probe showed promising limited deviation, especially in hypothermic patients.
引用
收藏
页数:6
相关论文
共 50 条
  • [21] In Vivo Evaluation of Two Hemorrhagic Shock Resuscitation Controllers with Non-Invasive, Intermittent Sensors
    Rodgers, Tina M.
    Berard, David
    Gonzalez, Jose M.
    Vega, Saul J.
    Gathright, Rachel
    Bedolla, Carlos
    Ross, Evan
    Snider, Eric J.
    BIOENGINEERING-BASEL, 2024, 11 (12):
  • [22] Progressive evaluation in spectroscopic sensors for non-invasive blood haemoglobin analysis-a review
    Kumar, Yogesh
    Dogra, Ayush
    Kaushik, Ajeet
    Kumar, Sanjeev
    PHYSIOLOGICAL MEASUREMENT, 2022, 43 (02)
  • [23] Non-invasive monitoring of muscle blood perfusion by photoplethysmography: evaluation of a new application
    Sandberg, M
    Zhang, Q
    Styf, J
    Gerdle, B
    Lindberg, LG
    ACTA PHYSIOLOGICA SCANDINAVICA, 2005, 183 (04): : 335 - 343
  • [24] Comparison of optical sensors for non-invasive hemoglobin measurement
    Desai, Brinda
    Chaskar, Uttam
    2016 INTERNATIONAL CONFERENCE ON ELECTRICAL, ELECTRONICS, AND OPTIMIZATION TECHNIQUES (ICEEOT), 2016, : 2211 - 2214
  • [25] Non-Invasive Lactate Monitoring System Using Wearable Chipless Microwave Sensors With Enhanced Sensitivity and Zero Power Consumption
    Baghelani, Masoud
    Abbasi, Zahra
    Daneshmand, Mojgan
    Light, Peter E.
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2022, 69 (10) : 3175 - 3182
  • [26] A non-invasive method for gastrointestinal parameter monitoring
    Wen-Xing Wang
    World Journal of Gastroenterology, 2005, (04) : 521 - 524
  • [27] A non-invasive method for gastrointestinal parameter monitoring
    Wang, Wen-Xing
    Yan, Guo-Zheng
    Sun, Fang
    Jiang, Ping-Ping
    Zhang, Wen-Qiang
    Zhang, Gen-Fu
    WORLD JOURNAL OF GASTROENTEROLOGY, 2005, 11 (04) : 521 - 524
  • [28] A Non-Invasive Blood Glucose Monitoring Device
    Kassem, Abdallah
    Hamad, Mustapha
    Harbieh, George G.
    El Moucary, Chady
    2020 IEEE 5TH MIDDLE EAST AND AFRICA CONFERENCE ON BIOMEDICAL ENGINEERING (MECBME), 2020, : 150 - 153
  • [29] Non-invasive monitoring using photoplethysmography technology
    Keisuke Tomita
    Taka-aki Nakada
    Taku Oshima
    Takehiko Oami
    Tuerxun Aizimu
    Shigeto Oda
    Journal of Clinical Monitoring and Computing, 2019, 33 : 637 - 645
  • [30] Trends in chemical sensors for non-invasive breath analysis
    Stewart, Trenton K.
    Carotti, Ines E.
    Qureshi, Yasser M.
    Covington, James A.
    TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2024, 177