Monte Carlo modeling and dosimetric validation of GammaMed Plus 192Ir source for HDR brachytherapy application using TOPAS

被引:1
作者
Shende, Ravindra [1 ,2 ]
Dhoble, S. J. [2 ]
Gayake, Umesh [3 ]
机构
[1] Balco Med Ctr, Dept Radiat Oncol, Sect 36, Naya Raipur 493661, Chhattisgarh, India
[2] RTM Nagpur Univ, Dept Phys, Nagpur 440033, Maharashtra, India
[3] Tata Mem Hosp, Adv Ctr Treatment Res & Educ Canc ACTREC, Dept Radiat Oncol, Navi Mumbai 410210, Maharashtra, India
关键词
MC simulation; TOPAS; 192 Ir source; Brachytherapy; DOSE CALCULATION; AAPM; KERMA; RECOMMENDATIONS; PARAMETERS; SPECTRA;
D O I
10.1016/j.apradiso.2024.111584
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Study aimed at Monte Carlo (MC) modeling and dosimetric validation of GammaMed Plus 192Ir source for high dose rate brachytherapy using Geant4 code in TOPAS. Study investigated the several TG-43 guided dosimetric parameters for brachytherapy dose calculation. The excellent agreement was found between simulated and published data. The simulated MC dataset can be served as reference and fed into Treatment planning system (TPS) to verify brachytherapy dose calculation. Present MC modeling of brachytherapy source can be availed in further patient specific studies involving MC simulation.
引用
收藏
页数:7
相关论文
共 22 条
[1]   Technical note:: Monte-Carlo dosimetry of the HDR 12i and plus 192Ir sources [J].
Ballester, F ;
Puchades, V ;
Lluch, JL ;
Serrano-Andrés, MA ;
Limami, Y ;
Pérez-Calatayud, J ;
Casal, E .
MEDICAL PHYSICS, 2001, 28 (12) :2586-2591
[2]   A generic high-dose rate 192Ir brachytherapy source for evaluation of model-based dose calculations beyond the TG-43 formalism [J].
Ballester, Facundo ;
Tedgren, Asa Carlsson ;
Granero, Domingo ;
Haworth, Annette ;
Mourtada, Firas ;
Fonseca, Gabriel Paiva ;
Zourari, Kyveli ;
Papagiannis, Panagiotis ;
Rivard, Mark J. ;
Siebert, Frank-Andre ;
Sloboda, Ron S. ;
Smith, Ryan L. ;
Thomson, Rowan M. ;
Verhaegen, Frank ;
Vijande, Javier ;
Ma, Yunzhi ;
Beaulieu, Luc .
MEDICAL PHYSICS, 2015, 42 (06) :3048-3062
[3]   Beta versus gamma dosimetry close to Ir-192 brachytherapy sources [J].
Baltas, D ;
Karaiskos, P ;
Papagiannis, P ;
Sakelliou, L ;
Loeffler, E ;
Zamboglou, N .
MEDICAL PHYSICS, 2001, 28 (09) :1875-1882
[4]   Report of the Task Group 186 on model-based dose calculation methods in brachytherapy beyond the TG-43 formalism: Current status and recommendations for clinical implementation [J].
Beaulieu, Luc ;
Tedgren, Asa Carlsson ;
Carrier, Jean-Francois ;
Davis, Stephen D. ;
Mourtada, Firas ;
Rivard, Mark J. ;
Thomson, Rowan M. ;
Verhaegen, Frank ;
Wareing, Todd A. ;
Williamson, Jeffrey F. .
MEDICAL PHYSICS, 2012, 39 (10) :6208-6236
[5]   Validation of the TOPAS Monte Carlo toolkit for HDR brachytherapy simulations [J].
Berumen, Francisco ;
Ma, Yunzhi ;
Ramos-Mendez, Jose ;
Perl, Joseph ;
Beaulieu, Luc .
BRACHYTHERAPY, 2021, 20 (04) :911-921
[6]   Spectra and air-kerma strength for encapsulated 192Ir sources [J].
Borg, J ;
Rogers, DWO .
MEDICAL PHYSICS, 1999, 26 (11) :2441-2444
[7]   Monte Carlo-aided dosimetry of a new high dose-rate brachytherapy source [J].
Daskalov, GM ;
Loffler, E ;
Williamson, JF .
MEDICAL PHYSICS, 1998, 25 (11) :2200-2208
[8]  
GLASGOW G P, 1979, Medical Physics (Woodbury), V6, P49, DOI 10.1118/1.594551
[9]  
Hubbell J.H., 2004, Tables of x-ray mass attenuation coefficients and mass energy-absorption coefficients (version 1.4), DOI DOI 10.1118/1.2734726
[10]  
Lopez J.A., 2011, Rev Fis Med, V12, P159