Systematic characterization of new EBT4 radiochromic films in clinical x-ray beams

被引:0
|
作者
Khan, Rao [1 ]
Rahimi, Robabeh [2 ]
Fan, Jiajin [3 ]
Chen, Kuan Ling [3 ]
机构
[1] Howard Univ, Dept Phys & Astron, Med Phys Program, Washington, DC 20059 USA
[2] Univ Maryland, Sch Med, Baltimore, MD USA
[3] Inova Schar Canc Inst, Dept Adv Radiat Oncol & Proton Therapy, Fairfax, VA USA
来源
BIOMEDICAL PHYSICS & ENGINEERING EXPRESS | 2025年 / 11卷 / 01期
关键词
radiochromic film; dosimetry; EBT4; megavoltage therapy; GAFCHROMIC FILMS; DOSE-RESPONSE; DOSIMETRY; DEPENDENCE; MODEL;
D O I
10.1088/2057-1976/ad8c49
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective. We aim to characterize kinetics of radiation-induced optical density in newly released EBT4radiochromicfilms exposed to clinical x-rays. Severalfilm models and batches were evaluated for thefilmsensitivity, optical signal increasing with time, relativefilm noise, and minimum detectable limits(MDL).Approach.Radiochromicfilm pieces from a single batch of EBT3 and three batches of EBT4 were exposedto doses of 77.38 cGy, 386.92 cGy, and 773.84 cGy using a 6 MV x-ray beam. Thefilms were scanned withaflatbed scanner at specific time intervals up to 120 h. The time-series net optical density of red, green andblue colors was corrected for response of the scanner with time and studied to establish the saturationcharacteristics offilm polymerization process. Dose-response from 3.86 cGy to 1935 cGy was alsodetermined for each color. MDL of thefilms was quantitatively defined as the dose that would double thenet optical density of red color above the standard deviation of the residual signal at zero dose. The relativenoise characteristics of EBT3 versus EBT4 were studied as a function of time, dose and scanner resolution.Main Results. For doses >= 100 cGy, analysis revealed a stability of optical density beyond 48 h post-exposure for EBT3 and EBT4films. EBT3films attained 80%-90% of their net optical density at 48 hwithin minutes of irradiation, compared to 72%-88% for EBT4films. The rate of growth was slowest forblue color, fastest for red, while green was in between the two. The MDL for EBT4 averaged 15 cGy forthree batches, whereas EBT3films reliably detected doses as low as 8.5 cGy.Significance. Several batches ofthe new EBT4film showed slightly lower response compared to its predecessor over 3.86 cGy to 1935 Gyrange. For all practical purposes, the post-irradiation growth of polymers ceases between 48 to 60 h forboth EBTfilms. Overall, the EBT4film exhibited noise characteristics similar to EBT3, except for lowerdoses where the noise was observed to be higher than its predecessor.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Spectral characterization and comparison of EBT3, EBT4, and EBT-XD radiochromic films
    Pecic, S.
    Devic, S.
    Belca, I
    Mosic, M.
    Kurij, Lj
    Nidzovic, B.
    Stojadinovic, S.
    PHYSICS IN MEDICINE AND BIOLOGY, 2025, 70 (01):
  • [2] Dose-response curve of EBT, EBT2, and EBT3 radiochromic films to synchrotron-produced monochromatic x-ray beams
    Brown, Thomas A. D.
    Hogstrom, Kenneth R.
    Alvarez, Diane
    Matthews, Kenneth L., II
    Ham, Kyungmin
    Dugas, Joseph P.
    MEDICAL PHYSICS, 2012, 39 (12) : 7412 - 7417
  • [3] Comparative analysis of dosimetric uncertainty using Gafchromic™ EBT4 and EBT3 films in radiochromic film dosimetry
    Akdeniz, Yucel
    RADIATION PHYSICS AND CHEMISTRY, 2024, 220
  • [4] Energy response of the new EBT2 radiochromic film to x-ray radiation
    Butson, Martin J.
    Yu, Peter K. N.
    Cheung, Tsang
    Alnawaf, Hani
    RADIATION MEASUREMENTS, 2010, 45 (07) : 836 - 839
  • [5] Characterization of the Response of EBT-XD Radiochromic Films to Proton Therapy Beams
    Darafsheh, A.
    Zhao, T.
    Khan, R.
    MEDICAL PHYSICS, 2020, 47 (06) : E717 - E717
  • [6] Response characterization of EBT-XD radiochromic films in megavoltage photon and electron beams
    Leon-Marroquin, E. Yazmin
    Mulrow, Daniel
    Darafsheh, Arash
    Khan, Rao
    MEDICAL PHYSICS, 2019, 46 (09) : 4246 - 4256
  • [7] Spectral Analysis of the EBT3 Radiochromic Films for Clinical Photon and Electron Beams
    Leon, Y.
    Mulrow, D.
    Khan, R.
    Darafsheh, A.
    MEDICAL PHYSICS, 2018, 45 (06) : E311 - E311
  • [8] Spectral analysis of the EBT3 radiochromic films for clinical photon and electron beams
    Leon-Marroquin, Elsa Y.
    Mulrow, Daniel J.
    Khan, Rao
    Darafsheh, Arash
    MEDICAL PHYSICS, 2019, 46 (02) : 973 - 982
  • [9] Evaluation of a new radiochromic film dosimeter, Gafchomic EBT4, for VMAT, SABR and HDR treatment delivery verification
    Palmer, Antony L.
    Nash, David
    Polak, Wojciech
    Wilby, Sarah
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (17):
  • [10] Systematic Evaluation of the Spectral Response of EBT3 Radiochromic Films to Proton Therapy Beams
    Darafsheh, A.
    Mulrow, D.
    Leon, Y.
    Zhao, T.
    Mossahebi, S.
    Kassaee, A.
    Khan, R.
    MEDICAL PHYSICS, 2019, 46 (06) : E326 - E326