Magnesium versus titanium headless compression screws for fixation of a Hoffa-type fracture - a biomechanical in-vitro study

被引:1
作者
Felten, Larissa [1 ]
Migliorini, Filippo [2 ]
Hildebrand, Frank [2 ]
Weber, Christian David [2 ]
机构
[1] Rhein Westfal TH Aachen, D-52074 Aachen, Germany
[2] Univ Hosp Aachen, Dept Orthopaed Trauma & Reconstruct Surg, Pauwelsstr 30, D-52074 Aachen, Germany
关键词
Hoffa fracture; Biomechanics; Magnesium screws; Resorbable implant; Distal femur; Coronal fractures of the femur; TOTAL KNEE ARTHROPLASTY; DISTAL FEMORAL CONDYLE; CORONAL FRACTURES; UNICONDYLAR FRACTURES; EXTENSOR MECHANISM; REMOVAL; ALLOYS; VIVO;
D O I
10.1186/s12891-025-08559-2
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background Intraarticular (e.g., Hoffa-type) fractures are usually stabilized with titanium screws, which may necessitate later implant removal. The aim of this study was to compare the biomechanical strength and stability of magnesium and titanium screws. Methods 18 double-layer bone blocks were fixed with 18 one-layer bone blocks and divided into 2 groups based on the fixation method: magnesium screw fixation (Magnesium group, n = 9) and traditional titanium screw (Titanium group, n = 9). Compressive force was applied to the specimens orthogonally to the screw axis. First, axial stiffness was measured, and a cyclic loading test was performed, after 10, 100, 1000 and 10,000 cycles respectively, and the axial displacements were recorded. Finally, the specimens were loaded to failure. Results There were significant differences between the groups with respect to axial stiffness and cyclic loading. The ultimate failure load was comparable. The average axial stiffness for the magnesium group was 326 +/- 67 N/mm and for the titanium group 266 +/- 72 N/mm (p = 0.031). The axial displacement relative to 100 N preload after 10,000 cycles in the magnesium group was 1.7319 +/- 0.2261 mm and in the titanium group 2.6932 +/- 0.5921 mm (p < 0.001). The average ultimate failure in the magnesium group was 920 +/- 55 N and in the titanium group 944 +/- 40 N (p = 0.293). Conclusions Based on the results magnesium screws show at least a comparable strength and stability as titanium screws in this setting. This study provides support from a biomechanical perspective for the use of magnesium screws in Hoffa fractures.
引用
收藏
页数:8
相关论文
共 59 条
[1]   Coronal plane partial articular fractures of the distal femoral condyle CURRENT CONCEPTS IN MANAGEMENT [J].
Arastu, M. H. ;
Kokke, M. C. ;
Duffy, P. J. ;
Korley, R. E. C. ;
Buckley, R. E. .
BONE & JOINT JOURNAL, 2013, 95B (09) :1165-1171
[2]   Hallux valgus correction utilising a modified short scarf osteotomy with a magnesium biodegradable or titanium compression screws - a comparative study of clinical outcomes [J].
Atkinson, Henry Dushan ;
Khan, Shahnawaz ;
Lashgari, Yasha ;
Ziegler, Andreas .
BMC MUSCULOSKELETAL DISORDERS, 2019, 20 (1) :334
[3]   History of femoral head fracture and coronal fracture of the femoral condyles [J].
Bartonicek, Jan ;
Rammelt, Stefan .
INTERNATIONAL ORTHOPAEDICS, 2015, 39 (06) :1245-1250
[4]   Unicondylar fractures of the distal femur [J].
Bel, J. -C. ;
Court, C. ;
Cogan, A. ;
Chantelot, C. ;
Pietu, G. ;
Vandenbussche, E. .
ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2014, 100 (08) :873-877
[5]   The outcomes of surgical treatment for lateral Hoffa fracture nonunions [J].
Bhowmick, Kaushik ;
Jepegnanam, Thilak Samuel ;
Inja, Dan Barnabas ;
Karuppusami, Reka ;
Nithyananth, Manasseh .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2023, 143 (05) :2509-2517
[6]   Hoffa fracture: fixation using headless compression screws [J].
Borse, V. ;
Hahnel, J. ;
Cohen, A. .
EUROPEAN JOURNAL OF TRAUMA AND EMERGENCY SURGERY, 2010, 36 (05) :477-479
[7]   Hardware removal: Indications and expectations [J].
Busam, ML ;
Esther, RJ ;
Obremskey, WT .
JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2006, 14 (02) :113-120
[8]   Open bicondylar Hoffa fracture associated with extensor mechanism injury [J].
Calmet, J ;
Mellado, JM ;
Forcada, IG ;
Giné, J .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2004, 18 (05) :323-325
[9]   A computed tomography-based classification of Hoffa fracture: Surgical treatment considerations and prognostic outcome with assessment of reproducibility [J].
Chandrabose, Rex ;
Saha, Sayantan ;
Kumar, Harish ;
Tapadiya, Niraj ;
Bharath, H. D. .
JOURNAL OF ORTHOPAEDICS, 2020, 20 :21-27
[10]  
Cheng P. L., 2009, Hong Kong Medical Journal, V15, P385