Enhancing fixation stability in proximal humerus fractures: screw orientation optimization in PHILOS plates through finite element analysis and biomechanical testing

被引:2
作者
Liu, Jichao [1 ]
Zhang, Ziyan [1 ]
Li, Peng [2 ]
Piao, Chengdong [1 ]
机构
[1] Second Hosp Jilin Univ, Dept Orthoped, Changchun, Jilin, Peoples R China
[2] Jilin Univ, Dept Engn Mech, Changchun, Jilin, Peoples R China
关键词
Proximal humerus fractures; Finite element analysis; PHILOS; Biomechanics; Parametric optimization; LOCKING PLATE; BONE-DENSITY; JOINT LOADS; EPIDEMIOLOGY; MODELS;
D O I
10.1038/s41598-024-78702-x
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
The optimal treatment strategy for proximal humerus fractures (PHFs) is debatable owing to the relatively high failure rate of locking plates. Optimizing implants may enhance the fixation stability of PHFs and reduce the rate of mechanical failures. We developed a finite element (FE) model to simulate the treatment of PHFs with Proximal Humerus Internal Locking System (PHILOS) plates. The model evaluated the average bone strain around the screw tips under vertical loading (as an alternative to the risk of cyclic screw cutout failure verified through biomechanical testing) to minimize this strain and maximize predicted fixation stability. After determining the optimal screw configuration, further FE analysis and in vitro biomechanical testing were conducted on both standard and optimized PHILOS screw orientation to assess whether the optimized plates have biomechanical advantages over the standard plates. The FE-based optimized configuration exhibited significantly lower bone strain around the implant than the standard PHILOS screw orientation (- 17.24%, p < 0.001). In both FE analysis and in vitro biomechanical testing, the optimized PHILOS plates achieved significantly lower average bone strain around the screws (p < 0.05), more uniform stress distribution, and greater structural stiffness (p < 0.05) than the standard PHILOS screw orientation. Our results show that biomechanical performance of the PHILOS plates can be improved by altering the orientation of the locking screws. This approach may be useful for future patient-specific design optimization of implants for other fractures.
引用
收藏
页数:14
相关论文
共 40 条
[21]   PHILOS Synthesis for Proximal Humerus Fractures Has High Complications and Reintervention Rates: A Systematic Review and Meta-Analysis [J].
Oldrini, Lorenzo Massimo ;
Feltri, Pietro ;
Albanese, Jacopo ;
Marbach, Francesco ;
Filardo, Giuseppe ;
Candrian, Christian .
LIFE-BASEL, 2022, 12 (02)
[22]   Late screw-related complications in locking plating of proximal humerus fractures: A systematic review [J].
Panagiotopoulou, Vasiliki C. ;
Varga, Peter ;
Richards, R. Geoff ;
Gueorguiev, Boyko ;
Giannoudis, Peter V. .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2019, 50 (12) :2176-2195
[23]  
Patel Akshar H, 2022, JSES Int, V6, P137, DOI 10.1016/j.jseint.2021.08.006
[24]   Epidemiology of proximal humerus fractures managed in a trauma center [J].
Roux, A. ;
Decroocq, L. ;
El Batti, S. ;
Bonnevialle, N. ;
Moineau, G. ;
Trojani, C. ;
Boileau, P. ;
de Peretti, F. .
ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2012, 98 (06) :715-719
[25]   One size may not fit all: patient-specific computational optimization of locking plates for improved proximal humerus fracture fixation [J].
Schader, Jana Felicitas ;
Mischler, Dominic ;
Dauwe, Jan ;
Richards, R. Geoff ;
Gueorguiev, Boyko ;
Varga, Peter .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2022, 31 (01) :192-200
[26]   Role of inferomedial supporting screws for secondary varus deformity in non-osteoporotic proximal humerus fracture: A biomechanical study [J].
Shin, Myung Jin ;
Kim, Hyojune ;
Kim, Dong Min ;
Park, Dongjun ;
Jeon, In-Ho ;
Koh, Kyoung Hwan .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2021, 141 (09) :1517-1523
[27]   Computational analysis of primary implant stability in trabecular bone [J].
Steiner, Juri A. ;
Ferguson, Stephen J. ;
van Lenthe, G. Harry .
JOURNAL OF BIOMECHANICS, 2015, 48 (05) :807-815
[28]   Open Reduction and Internal Fixation of Proximal Humeral Fractures with Use of the Locking Proximal Humerus Plate Results of a Prospective, Multicenter, Observational Study [J].
Suedkamp, N. ;
Bayer, J. ;
Hepp, P. ;
Voigt, C. ;
Oestern, H. ;
Kaeaeb, M. ;
Luo, C. ;
Plecko, M. ;
Wendt, K. ;
Koestler, W. ;
Konrad, G. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2009, 91A (06) :1320-1328
[29]   The influence of bone density and anisotropy in finite element models of distal radius fracture osteosynthesis: Evaluations and comparison to experiments [J].
Synek, A. ;
Chevalier, Y. ;
Baumbach, S. F. ;
Pahr, D. H. .
JOURNAL OF BIOMECHANICS, 2015, 48 (15) :4116-4123
[30]   Biomechanical Testing of Additive Manufactured Proximal Humerus Fracture Fixation Plates [J].
Tilton, Maryam ;
Armstrong, April ;
Sanville, Jennifer ;
Chin, Matthew ;
Hast, Michael W. ;
Lewis, Gregory S. ;
Manogharan, Guha P. .
ANNALS OF BIOMEDICAL ENGINEERING, 2020, 48 (01) :463-476