Volume assessment comparing femoral vein and inferior vena cava among chest pain patients presenting to the emergency department

被引:1
作者
Nair, Neeraja A. [1 ]
Sirur, Freston Marc [1 ]
Krishnan, S. Vimal [1 ]
机构
[1] Manipal Acad Higher Educ MAHE, Kasturba Med Coll, Dept Emergency Med, Manipal, Karnataka, India
关键词
POCUS (point of Care Ultrasound); Femoral vein Collapsibility Index; Volume assessment; Inferior Vena Cava; Volume responsiveness; Emergency Department; PREDICT FLUID RESPONSIVENESS; RESPIRATORY VARIATION; DIAMETER; ULTRASOUND;
D O I
10.1186/s12245-024-00762-7
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BackgroundInferior vena cava (IVC) diameter measurement using ultrasound for volume status assessment has shown satisfactory results and is being adopted in Emergency and critical care settings. IVC diameter can vary depending on the cardiac function, respiratory efforts, intraabdominal pressure, and mechanical ventilation. Due to these factors, IVC measurement cannot be considered a stand-alone technique appropriate for every patient. The femoral vein (FV), a more superficial vein than IVC, can be considered an alternative method for assessing fluid responsiveness in patients presenting to the Emergency department. It is easily accessible and can be used in scenarios where IVC cannot be visualized or reliable.MethodsThis was a single-center diagnostic study where 85 patients who presented to the ED with chest pain were enrolled prospectively. IVC and femoral vein collapsibility indices, stroke volume, and cardiac output are measured using an ultrasound machine. The measurements were repeated after a passive leg-raising test. These values were compared with each other to assess an intra-class correlation between IVC and femoral vein collapsibility indices. We have also evaluated the relationship between the collapsibility indices of both veins and cardiac output.Discussion & limitationsOur findings show an insufficient correlation between IVC and FV collapsibility indices. However, both vein diameters significantly increased after passive leg raising (PLR), indicating a response to fluid challenge. Post-PLR reduced IVC, and FV collapsibility index (CI) suggests intravascular volume expansion after a fluid challenge, also reflected in the hemodynamic parameters. Our study was conducted only in a subset of relatively stable patients. The applicability of the study in different subsets of patients presenting to ED is still questionable.ConclusionWe conclude that femoral vein indices may not be an accurate alternative for volume assessment in the chosen cohort of patients. IVC and FV metrics do not correlate and may not be accurate for volume responsiveness. We may need to explore the utility of FV and its indices in a larger population in multiple settings for a better understanding of its role in volume assessment and responsiveness.Trial registration(EC/NEW/INST/2021/1707). Registered 03 January 2023.
引用
收藏
页数:7
相关论文
共 22 条
  • [1] A novel method of calculating stroke volume using point-of-care echocardiography
    Aligholizadeh, Ehson
    Teeter, William
    Patel, Rajan
    Hu, Peter
    Fatima, Syeda
    Yang, Shiming
    Ramani, Gautam
    Safadi, Sami
    Olivieri, Peter
    Scalea, Thomas
    Murthi, Sarah
    [J]. CARDIOVASCULAR ULTRASOUND, 2020, 18 (01)
  • [2] Begum M., 2023, Life Sci, V4, P132, DOI [10.37185/LnS.1.1.272, DOI 10.37185/LNS.1.1.272]
  • [3] Diagnostic accuracy of passive leg raising for prediction of fluid responsiveness in adults: systematic review and meta-analysis of clinical studies
    Cavallaro, Fabio
    Sandroni, Claudio
    Marano, Cristina
    La Torre, Giuseppe
    Mannocci, Alice
    De Waure, Chiara
    Bello, Giuseppe
    Maviglia, Riccardo
    Antonelli, Massimo
    [J]. INTENSIVE CARE MEDICINE, 2010, 36 (09) : 1475 - 1483
  • [4] Basic concepts of fluid responsiveness
    Cherpanath, T. G. V.
    Geerts, B. F.
    Lagrand, W. K.
    Schultz, M. J.
    Groeneveld, A. B. J.
    [J]. NETHERLANDS HEART JOURNAL, 2013, 21 (12) : 530 - 536
  • [5] Predicting Fluid Responsiveness by Passive Leg Raising: A Systematic Review and Meta-Analysis of 23 Clinical Trials
    Cherpanath, Thomas G. V.
    Hirsch, Alexander
    Geerts, Bart F.
    Lagrand, Wim K.
    Leeflang, Mariska M.
    Schultz, Marcus J.
    Groeneveld, A. B. Johan
    [J]. CRITICAL CARE MEDICINE, 2016, 44 (05) : 981 - 991
  • [6] Measurement of Femoral Vein Diameter by Ultrasound to Estimate Central Venous Pressure
    Cho, Roy J.
    Williams, David R.
    Leatherman, James W.
    [J]. ANNALS OF THE AMERICAN THORACIC SOCIETY, 2016, 13 (01) : 81 - 85
  • [7] Monitoring volume and fluid responsiveness: From static to dynamic indicators
    Guerin, Laurent
    Monnet, Xavier
    Teboul, Jean-Louis
    [J]. BEST PRACTICE & RESEARCH-CLINICAL ANAESTHESIOLOGY, 2013, 27 (02) : 177 - 185
  • [8] Accurate Measurement of Left Ventricular Outflow Tract Diameter: Comment on the Updated Recommendations for the Echocardiographic Assessment of Aortic Valve Stenosis
    Hahn, Rebecca T.
    Pibarot, Philippe
    [J]. JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2017, 30 (10) : 1038 - 1041
  • [9] Fluid balance correlates with clinical course of multiple organ dysfunction syndrome and mortality in patients with septic shock
    Huang, Allen Chung-Cheng
    Lee, Tim Yu-Ting
    Ko, Meng-Cheng
    Huang, Chih-Hsien
    Wang, Tsai-Yu
    Lin, Ting-Yu
    Lin, Shu-Min
    [J]. PLOS ONE, 2019, 14 (12):
  • [10] Predicting fluid responsiveness: A review of literature and a guide for the clinician
    Jalil, Bilal A.
    Cavallazzi, Rodrigo
    [J]. AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2018, 36 (11) : 2093 - 2102