Robustness of 4D flow MRI derived aortic wall shear stress and pulse wave velocity across different protocols in healthy controls and in patients with bicuspid aortic valve

被引:1
|
作者
Dushfunian, David [1 ]
Maroun, Anthony [1 ]
Berhan, Haben [1 ,2 ]
Baraboo, Justin [1 ]
Johnson, Ethan M. [1 ]
Jarvis, Kelly [1 ]
Allen, Bradley D. [1 ]
Markl, Michael [1 ,2 ]
机构
[1] Northwestern Univ, Feinberg Sch Med, Dept Radiol, Chicago, IL 60208 USA
[2] Northwestern Univ, McCormick Sch Engn, Dept Biomed Engn, Evanston, IL USA
来源
关键词
4D-flow MRI; Wall shear stress; Aortic stiffness; Pulse wave velocity; Net flow; Reproducibility; PHASE-CONTRAST MRI; THORACIC AORTA; REPRODUCIBILITY; VARIABILITY; EXPRESSION; STIFFNESS; DILATION;
D O I
10.1007/s10554-024-03299-1
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Purpose: To evaluate the reproducibility of important biomarkers like wall shear stress (WSS), pulse wave velocity (PWV), and net flow across two 4D flow MRI imaging protocols with different coverages: aorta-targeted 4D flow MRI (AT4D) and whole-heart 4D flow (WH4D) protocols. Methods: Thirty-eight control subjects (43.2 +/- 10.1 years old; 22 males) and ten patients (45.7 +/- 8.9 years old; 7 males) with bicuspid aortic valve (BAV) were included. Each subject underwent AT4D and WH4D scans. Absolute WSS, PWV, and net flow were assessed for each patient across both protocols and compared using Bland-Altman analysis. Areas of elevated WSS were assessed for BAV patients across different WSS thresholds that define WSS to be elevated compared to a normal population average. A sensitivity analysis was conducted to determine the best WSS threshold at which WH4D-derived areas most closely resemble AT4D-derived areas. Inter-rater reproducibility was evaluated in twenty-four subjects. Results: AT4D and WH4D PWV and WSS estimates demonstrated good agreement (PWV: -0.12 +/- 1.84 m/s, p = 0.4; Median WSS: 0.06 +/- 0.13 Pa, p < 0.01; Maximum WSS: 0.04 +/- 0.27 Pa, p = 0.07). Good agreement was also found for AAo net flow (8.14 +/- 24.86 mL/cycle, p < 0.01). PWV correlated with age across protocols (AT4D: r = 0.68, p < 0.01; WH4D: r = 0.72, p < 0.01). Sensitivity analysis identified a WSS threshold where WH4D-derived areas of elevated WSS most closely resembled AT4D-derived areas. Inter-rater assessment of the tested parameters resulted in a small mean difference percentage of < 3%. Data conclusion: PWV, WSS, and net flow demonstrated good agreement across protocols. The WSS threshold should be adjusted for WH4D estimates to optimally match AT4D-derived output. Reproducibility analysis showed good test-retest agreement. This study demonstrates the reproducibility of certain hemodynamic parameters across two 4D flow MRI protocol.
引用
收藏
页码:137 / 149
页数:13
相关论文
共 50 条
  • [1] Expanding the volume on 4D flow CMR Editorial for "Robustness of 4D flow MRI derived aortic wall shear stress and pulse wave velocity across different protocols in healthy controls and in patients with bicuspid aortic valve"
    Fischer, Kady
    Bastiaansen, Jessica A. M.
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2025, 41 (02): : 181 - 183
  • [2] Multiyear Interval Changes in Aortic Wall Shear Stress in Patients with Bicuspid Aortic Valve Assessed by 4D Flow MRI
    Maroun, Anthony
    Scott, Michael B.
    Catania, Roberta
    Berhane, Haben
    Jarvis, Kelly
    Allen, Bradley D.
    Barker, Alex J.
    Markl, Michael
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2024, 60 (06) : 2580 - 2589
  • [3] Editorial for "Multiyear Interval Changes in Aortic Wall Shear Stress in Patients With Bicuspid Aortic Valve Assessed by 4D Flow MRI"
    Guala, Andrea
    Ferrer-Cornet, Marta
    Dux-Santoy, Lydia
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2024, 60 (06) : 2590 - 2591
  • [4] Aortic Pulse Wave Velocity Evaluated by 4D Flow MRI Across the Adult Lifespan
    Jarvis, Kelly
    Scott, Michael B.
    Soulat, Gilles
    Elbaz, Mohammed S. M.
    Barker, Alex J.
    Carr, James C.
    Markl, Michael
    Ragin, Ann
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2022, 56 (02) : 464 - 473
  • [5] CMR 4D-Flow Wall Shear Stress and Aortic Dilation in Bicuspid Aortic Valve
    Salerno, Michael
    Chandrashekhar, Y.
    JACC-CARDIOVASCULAR IMAGING, 2022, 15 (01) : 177 - 179
  • [6] Aortic arch dilation in bicuspid aortic valve patients is related to altered hemodynamics and wall shear stress: a 4D flow CMR study
    Dux-Santoy, L.
    Guala, A.
    Rodriguez-Palomares, J.
    Teixido-Tura, G.
    Ruiz-Munoz, A.
    Granato, C.
    Villalva, N.
    Valente, F.
    Dentamaro, I.
    Sao-Aviles, A.
    Galian, L.
    Gutierrez, L.
    Fernandez, R.
    Gonzalez-Alujas, T.
    Garcia-Dorado, D.
    Evangelista, A.
    EUROPEAN JOURNAL OF CLINICAL INVESTIGATION, 2018, 48 : 61 - 61
  • [7] Editorial for "Aortic Pulse Wave Velocity Evaluated by 4D Flow MRI Across the Adult Lifespan"
    Fyrdahl, Alexander
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2022, 56 (02) : 474 - 475
  • [8] 4D-flow MRI derived wall shear stress for the risk stratification of bicuspid aortic valve aortopathy: A systematic review
    Qin, Jiaxing Jason
    Obeidy, Peyman
    Gok, Mustafa
    Gholipour, Alireza
    Grieve, Stuart M. M.
    FRONTIERS IN CARDIOVASCULAR MEDICINE, 2023, 9
  • [9] Interval changes in aortic peak velocity and wall shear stress in patients with bicuspid aortic valve disease
    Ozair Rahman
    Michael Scott
    Emilie Bollache
    Kenichiro Suwa
    Jeremy Collins
    James Carr
    Paul Fedak
    Patrick McCarthy
    Chris Malaisrie
    Alex J. Barker
    Michael Markl
    The International Journal of Cardiovascular Imaging, 2019, 35 : 1925 - 1934
  • [10] Interval changes in aortic peak velocity and wall shear stress in patients with bicuspid aortic valve disease
    Rahman, Ozair
    Scott, Michael
    Bollache, Emilie
    Suwa, Kenichiro
    Collins, Jeremy
    Carr, James
    Fedak, Paul
    McCarthy, Patrick
    Malaisrie, Chris
    Barker, Alex J.
    Markl, Michael
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2019, 35 (10): : 1925 - 1934