Method and system for detecting pulmonary function parameters by integrating four-subdivisions impedance information

被引:0
|
作者
Zhang, Wei [1 ]
Wang, Jingang [1 ]
Li, Tanxiao [2 ]
Zhang, Yapeng [1 ]
He, Wei [1 ]
Yu, Chuanxiang [1 ]
Zhao, Pengcheng [1 ]
机构
[1] Chongqing Univ, Sch Elect Engn, State Key Lab Power Transmiss Equipment Technol, Chongqing 400044, Peoples R China
[2] Chongqing Univ Univ Cincinnati Joint Coop Inst, Chongqing 401331, Peoples R China
关键词
Four subdivisions; Regional impedance integration; Impedance measurement focusing; LUNG; BIOIMPEDANCE; TISSUE; MODEL;
D O I
10.1016/j.measurement.2025.116795
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Pulmonary function parameter testing is the principal medical method for the early diagnosis and postoperative monitoring of diseases with impaired pulmonary ventilation. Bioimpedance measurement, known for its noninvasive nature and absence of respiratory resistance, has garnered widespread attention over traditional flow-rate methods for assessing pulmonary function. However, relying solely on integral thoracic impedance to characterize changes in pulmonary air volume overlooks the potential errors introduced by alterations in nonpulmonary tissues. This article proposes a method for assessing lung ventilation parameters based on regional information that integrates impedance data from four distinct lung subdivisions, constructing a multivariate nonlinear mathematical model of the impedance variation of the four lung branches in relation to lung volume, thereby minimizing the impact of non-pulmonary factors on calculations. The 2-D thoracic equivalent model simulations revealed the ability to achieve regional impedance measurement focusing with a specific drive measurement pattern and four zonal focusing measurement electrode configurations were obtained for the lungs. The precision of the four-zone electrode configurations was confirmed through 3-D lung model simulations, revealing a generalized mathematical relationship between zonal impedance changes and air volume transformations in corresponding lung zones. Optimize and determine the integral calculation equation by selecting the parameters with the highest individual difference correlation through 500 clinical trials. The clinical trial results showed that the maximum relative errors of forced ventilation volume (FVC) and forced expiratory volume in one second (FEV1) were 1.02 % and 1.31 %, respectively, with differences within +/- 1.96 SD.
引用
收藏
页数:18
相关论文
共 50 条
  • [1] IMPEDANCE CARDIOGRAPHY AS A NONINVASIVE METHOD OF MONITORING CARDIAC FUNCTION AND OTHER PARAMETERS OF CARDIOVASCULAR SYSTEM
    KUBICEK, WG
    PATTERSON, RP
    WITSOE, DA
    ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1970, 170 (02) : 724 - +
  • [2] An Integrating Method for Information System Based on CCM
    Zhang, Jian
    Huang, Kun
    Wang, Lu
    ADVANCES IN MANUFACTURING TECHNOLOGY, PTS 1-4, 2012, 220-223 : 2319 - 2322
  • [3] Integrating longitudinal information on pulmonary function and inflammation using asthma phenotypes
    Nagasaki, Tadao
    Matsumoto, Hisako
    Kanemitsu, Yoshihiro
    Izuhara, Kenji
    Tohda, Yuji
    Kita, Hideo
    Horiguchi, Takahiko
    Kuwabara, Kazunobu
    Tomii, Keisuke
    Otsuka, Kojiro
    Fujimura, Masaki
    Ohkura, Noriyuki
    Tomita, Katsuyuki
    Yokoyama, Akihito
    Ohnishi, Hiroshi
    Nakano, Yasutaka
    Oguma, Tetsuya
    Hozawa, Soichiro
    Ito, Isao
    Oguma, Tsuyoshi
    Inoue, Hideki
    Tajiri, Tomoko
    Iwata, Toshiyuki
    Izuhara, Yumi
    Ono, Junya
    Ohta, Shoichiro
    Yokoyama, Tetsuji
    Niimi, Akio
    Mishima, Michiaki
    JOURNAL OF ALLERGY AND CLINICAL IMMUNOLOGY, 2014, 133 (05) : 1474 - U406
  • [4] Regional Pulmonary Ventilation Assessment Method and System Based on Impedance Sensing Information from the Pentapulmonary Lobes
    Zhang, Yapeng
    Song, Chengxin
    He, Wei
    Zhang, Qian
    Zhao, Pengcheng
    Wang, Jingang
    SENSORS, 2024, 24 (10)
  • [5] Errors in Determination of the Parameters of Bioelectric Impedance by the Transient-Function Method
    Akulov, S. A.
    Fedotov, A. A.
    Akulova, A. S.
    MEASUREMENT TECHNIQUES, 2015, 58 (03) : 348 - 351
  • [6] Research on the Method of Cardiac Impedance Acquisition and Cardiac Function Parameters Analysis
    Ke, Li
    Zhan, Ting
    Du, Qiang
    INTERNATIONAL CONFERENCE ON ENERGY, ENVIRONMENT AND MATERIALS ENGINEERING (EEME 2014), 2014, : 981 - 985
  • [7] Errors in Determination of the Parameters of Bioelectric Impedance by the Transient-Function Method
    S. A. Akulov
    A. A. Fedotov
    A. S. Akulova
    Measurement Techniques, 2015, 58 : 348 - 351
  • [8] The Relationship Between Pulmonary System Impedance and Right Ventricular Function in Normal Sheep
    Kuo, Alexander S.
    Sato, Hitoshi
    Reoma, Junewai L.
    Cook, Keith E.
    CARDIOVASCULAR ENGINEERING, 2009, 9 (04) : 153 - 160
  • [9] A new threat assessment method for integrating an IoT infrastructure in an information system
    Dorsemaine, Bruno
    Gaulier, Jean-Philippe
    Wary, Jean-Philippe
    Kheir, Nizar
    Urien, Pascal
    2017 IEEE 37TH INTERNATIONAL CONFERENCE ON DISTRIBUTED COMPUTING SYSTEMS WORKSHOPS (ICDCSW), 2017, : 105 - 112