Agreement in anterior segment measurements between swept-source optical coherence and dual Scheimpflug tomography devices in keratoconus eyes

被引:0
作者
Lee, Yunjin [1 ,2 ]
Oh, Joo Youn [2 ,3 ,4 ]
Choi, Hyuk Jin [2 ,3 ,4 ,5 ]
Kim, Mee Kum [2 ,3 ,4 ]
Yoon, Chang Ho [2 ,3 ,4 ]
机构
[1] Gachon Univ, Gil Hosp, Dept Ophthalmol, Incheon, South Korea
[2] Seoul Natl Univ, Coll Med, Dept Ophthalmol, Seoul, South Korea
[3] Seoul Natl Univ, Seoul Natl Univ Hosp, Coll Med, Dept Ophthalmol, Seoul, South Korea
[4] Seoul Natl Univ Hosp, Artificial Eye Ctr, Lab Ocular Regenerat Med & Immunol LORMI, Biomed Res Inst, Seoul, South Korea
[5] Seoul Natl Univ Hosp, Healthcare Syst Gangnam Ctr, Dept Ophthalmol, Seoul, South Korea
关键词
Casia; 2; Dual Scheimpflug tomography; Galilei G6; Keratoconus; Swept-source optical coherence tomography; COLLAGEN CROSS-LINKING; ORBSCAN II; REPEATABILITY; RELIABILITY; PENTACAM;
D O I
10.1038/s41598-024-76031-7
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose To evaluate the agreement of ocular biometry measured using a swept-source optical coherence (Casia 2) and a dual Scheimpflug (Galilei G6) tomography in keratoconus. Methods This retrospective study included 102 eyes from 102 keratoconus patient examined using both devices. Parameters compared included flat (Kf) and steep (Ks) keratometry, astigmatism of anterior, posterior, and total keratometry, central (CCT) and thinnest (TCT) corneal thickness. To assess the agreement, intraclass coefficient (ICC) and Bland-Altman analysis with 95% limits of agreement (LoA) were used. Results Anterior and total Ks and Kf showed moderate or good agreement with 95% limits of agreement (LoA) range over 9.75 D. Posterior Ks and Kf were lower in the Galilei G6 (all p < 0.001). Astigmatism showed moderate, poor, and moderate agreement for anterior, posterior, and total keratometry, respectively. CCT and TCT showed excellent agreement; however, the 95% LoA range was over 60 mu m. Conclusion The agreement between the two devices was not excellent for most parameters used to diagnose keratoconus and assess disease progression, and the differences were clinically significant. Therefore, the measurements from these two devices are not interchangeable for patients with keratoconus.
引用
收藏
页数:9
相关论文
共 26 条
[1]   Anterior segment optical coherence tomography [J].
Ang, Marcus ;
Baskaran, Mani ;
Werkmeister, Rene M. ;
Chua, Jacqueline ;
Schmidl, Doreen ;
dos Santos, Valentin Aranha ;
Garhoefer, Gerhard ;
Mehta, Jodhbir S. ;
Schmetterer, Leopold .
PROGRESS IN RETINAL AND EYE RESEARCH, 2018, 66 :132-156
[2]   Agreement of total corneal power between 2 swept-source optical coherence tomography and Scheimpflug tomography in normal and keratoconic patients [J].
Asawaworarit, Rosepon ;
Satitpitakul, Vannarut ;
Taweekitikul, Parichart ;
Pongpirui, Krit .
PLOS ONE, 2022, 17 (05)
[3]   Transepithelial corneal collagen crosslinking for progressive keratoconus: 24-month clinical results [J].
Caporossi, Aldo ;
Mazzotta, Cosimo ;
Paradiso, Anna Lucia ;
Baiocchi, Stefano ;
Marigliani, Davide ;
Caporossi, Tomaso .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2013, 39 (08) :1157-1163
[4]   Comparison and Repeatability of Keratometric and Corneal Power Measurements Obtained by Orbscan II, Pentacam, and Galilei Corneal Tomography Systems [J].
Crawford, Alexandra Z. ;
Patel, Dipika V. ;
McGhee, Charles N. J. .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2013, 156 (01) :53-60
[5]   Assessing progression of keratoconus: novel tomographic determinants [J].
Duncan, Joshua K. ;
Belin, Michael W. ;
Borgstrom, Mark .
EYE AND VISION, 2016, 3
[6]   The Reliability of Successive Scheimpflug Imaging and Anterior Segment Optical Coherence Tomography Measurements Decreases With Increasing Keratoconus Severity [J].
Flockerzi, Elias ;
Elzer, Berit ;
Daas, Loay ;
Xanthopoulou, Kassandra ;
Eppig, Timo ;
Langenbucher, Achim ;
Seitz, Berthold .
CORNEA, 2021, 40 (11) :1433-1439
[7]   Corneal thickness changes after corneal collagen crosslinking for keratoconus and corneal ectasia: One-year results [J].
Greenstein, Steven A. ;
Shah, Vinnie P. ;
Fry, Kristen L. ;
Hersh, Peter S. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (04) :691-700
[8]   Effect of keratoconus grades on repeatability of keratometry readings: Comparison of 5 devices [J].
Hashemi, Hassan ;
Yekta, Abbasali ;
Khabazkhoob, Mehdi .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (05) :1065-1072
[9]   Comparison of corneal tomography using a novel swept-source optical coherence tomographer and rotating Scheimpflug system in normal and keratoconus eyes: repeatability and agreement analysis [J].
Herber, Robert ;
Lenk, Janine ;
Pillunat, Lutz E. ;
Raiskup, Frederik .
EYE AND VISION, 2022, 9 (01)
[10]   Agreement of Total Keratometry and Posterior Keratometry Among IOLMaster 700, CASIA2, and Pentacam [J].
Jin, Aixia ;
Han, Xiaotong ;
Zhang, Jiaqing ;
Qiu, Xiaozhang ;
Zhang, Yifan ;
Qu, Bo ;
Tan, Xuhua ;
Luo, Lixia .
TRANSLATIONAL VISION SCIENCE & TECHNOLOGY, 2023, 12 (03)