Beam hardening of K-edge contrast agents: a phantom study comparing clinical energy-integrating detector and photon-counting detector CT systems

被引:1
|
作者
Amir Pourmorteza [1 ]
Arnaud Richard Choux [2 ]
Thomas Wesley Holmes [3 ]
U. Joseph Schoepf [1 ]
Marly van Assen [1 ]
Carlo De Cecco [4 ]
Tilman Emrich [1 ]
Akos Varga-Szemes [1 ]
机构
[1] Department of Radiology and Imaging Sciences, Emory University, Atlanta, GA
[2] Department of Biomedical Engineering, Georgia Institute of Technology, Emory University, Atlanta, GA
[3] Winship Cancer Institute, Emory University, Atlanta, GA
[4] Department of Radiology, Medical University of South Carolina, Clinical Science Building, Charleston, SC
[5] Department of Diagnostic and Interventional Radiology, University Medical Center of the Johannes Gutenberg-University, Mainz
关键词
Artifacts; Bismuth; Gadolinium; Iodine; Tomography (x-ray computed);
D O I
10.1186/s41747-024-00530-5
中图分类号
学科分类号
摘要
Background: Beam hardening (BH) artifacts negatively influence computed tomography (CT) measurements, especially when due to dense materials or materials with high effective atomic numbers. Photon-counting detectors (PCD) are more susceptible to BH due to equal weighting of photons regardless of their energies. The problem is further confounded by the use of contrast agents (CAs) with K-edge in the diagnostic CT energy range. We quantified the BH effect of different materials comparing energy-integrating detector (EID)-CT and PCD-CT. Methods: Pairs of test tubes were filled with dense CA (iodine-, gadolinium-, and bismuth-based) and placed inside a water phantom. The phantoms were scanned on EID- and PCD-CT systems, at all available tube voltages for the PCD scanner. Images were reconstructed with standard water BH correction but without any iodine/bone BH corrections. Virtual monoenergetic images (VMI) were calculated from PCD-CT data. Results: PCD-CT had higher CT numbers in all x-ray spectra for all CAs (p < 0.001) and produced larger cupping artifacts in all test cases (p < 0.001). Bismuth-based CA artifacts were 3- to 5-fold smaller than those of iodine- or gadolinium-based CA. PCD-CT-based VMI completely removed iodine BH artifacts. Iodine BH artifacts decreased with increasing tube voltage. However, gadolinium-based BH artifacts had a different trend increasing at 120 kVp. Conclusion: EID had fewer BH artifacts compared to PCD at x-ray tube voltages of 120 kVp and higher. The inherent spectral information of PCDs can be used to eliminate BH artifacts. Special care is needed to correct BH artifacts for gadolinium- and bismuth-based CAs. Relevance statement: With the increasing availability of clinical photon-counting CT systems offering the possibility of dual contrast imaging capabilities, addressing and comprehending the BH artifacts attributed to old and novel CT CAs grows in research and ultimately clinical relevance. Key Points: EID-CT provides fewer BH artifacts compared to PCD-CT at x-ray tube voltages of 120 kVp and higher. K-edge CAs, such as those based on gadolinium, further confound BH artifacts. The inherent spectral information of photon counting detector CT can be used to effectively eliminate BH artifacts. © The Author(s) 2025.
引用
收藏
相关论文
共 14 条
  • [1] Intraindividual reproducibility of myocardial radiomic features between energy-integrating detector and photon-counting detector CT angiography
    Tremamunno, Giuseppe
    Varga-Szemes, Akos
    Schoepf, U. Joseph
    Laghi, Andrea
    Zsarnoczay, Emese
    Fink, Nicola
    Aquino, Gilberto J.
    O'Doherty, Jim
    Emrich, Tilman
    Vecsey-Nagy, Milan
    EUROPEAN RADIOLOGY EXPERIMENTAL, 2024, 8 (01)
  • [2] Assessment of metal artifacts from titanium wrist prostheses: photon-counting versus energy-integrating detector CT
    Nina Kämmerling
    Simon Farnebo
    Mårten Sandstedt
    Ronald Booij
    Anders Persson
    Erik Tesselaar
    European Radiology Experimental, 9 (1)
  • [3] Photon-counting detector CT and energy-integrating detector CT for trabecular bone microstructure analysis of cubic specimens from human radius
    Benjamin Klintström
    Lilian Henriksson
    Rodrigo Moreno
    Alexandr Malusek
    Örjan Smedby
    Mischa Woisetschläger
    Eva Klintström
    European Radiology Experimental, 6
  • [4] Photon-counting detector CT and energy-integrating detector CT for trabecular bone microstructure analysis of cubic specimens from human radius
    Klintstrom, Benjamin
    Henriksson, Lilian
    Moreno, Rodrigo
    Malusek, Alexandr
    Smedby, Orjan
    Woisetschlager, Mischa
    Klintstrom, Eva
    EUROPEAN RADIOLOGY EXPERIMENTAL, 2022, 6 (01)
  • [5] Experimental research of the energy bins for K-edge imaging using a photon counting detector: a phantom and mice study
    Zhang, Zhidu
    Hu, Jinming
    Zhang, Xiaomei
    Xu, Qiong
    Li, Mohan
    Wei, Cunfeng
    Wei, Long
    Wang, Zhe
    RADIATION DETECTION TECHNOLOGY AND METHODS, 2020, 4 (03) : 303 - 311
  • [6] Photon-counting versus energy-integrating CT of abdomen-pelvis: a phantom study on the potential for reducing iodine contrast media
    Djamel Dabli
    Maxime Pastor
    Sebastian Faby
    Julien Erath
    Cédric Croisille
    Fabricio Pereira
    Jean-Paul Beregi
    Joël Greffier
    European Radiology Experimental, 9 (1)
  • [7] Low-dose high-resolution chest CT in adults with cystic fibrosis: intraindividual comparison between photon-counting and energy-integrating detector CT
    Frings, Marko
    Welsner, Matthias
    Mousa, Christin
    Zensen, Sebastian
    Salhoefer, Luca
    Meetschen, Mathias
    Beck, Nikolas
    Bos, Denise
    Westhoelter, Dirk
    Wienker, Johannes
    Taube, Christian
    Umutlu, Lale
    Schaarschmidt, Benedikt M.
    Forsting, Michael
    Haubold, Johannes
    Sutharsan, Sivagurunathan
    Opitz, Marcel
    EUROPEAN RADIOLOGY EXPERIMENTAL, 2024, 8 (01)
  • [8] High-Pitch Photon-Counting Detector Computed Tomography Angiography of the Aorta Intraindividual Comparison to Energy-Integrating Detector Computed Tomography at Equal Radiation Dose
    Euler, Andre
    Higashigaito, Kai
    Mergen, Victor
    Sartoretti, Thomas
    Zanini, Bettina
    Schmidt, Bernhard
    Flohr, Thomas G.
    Ulzheimer, Stefan
    Eberhard, Matthias
    Alkadhi, Hatem
    INVESTIGATIVE RADIOLOGY, 2022, 57 (02) : 115 - 121
  • [9] Simultaneous dual-contrast imaging using energy-integrating detector multi-energy CT: An in vivo feasibility study
    Zhou, Zhongxing
    Ren, Liqiang
    Rajendran, Kishore
    Diehn, Felix E.
    Fletcher, Joel G.
    McCollough, Cynthia H.
    Yu, Lifeng
    MEDICAL PHYSICS, 2022, 49 (03) : 1458 - 1467
  • [10] Stent appearance in a novel silicon-based photon-counting CT prototype: ex vivo phantom study in head-to-head comparison with conventional energy-integrating CT
    Emma Verelst
    Nico Buls
    Johan De Mey
    Koenraad Hans Nieboer
    Frans Vandenbergh
    Dominic Crotty
    Paul Deak
    Albert Sundvall
    Staffan Holmin
    Aron De Smet
    Steven Provyn
    Gert Van Gompel
    European Radiology Experimental, 7