Enhanced high-sensitivity multi-layer neutron detector based on LiF:ZnS(Ag) scintillator

被引:0
|
作者
Zada, Ilan Cohen [1 ,2 ]
Osovizky, Alon [2 ,3 ]
Orion, Itzhak [1 ]
机构
[1] Ben Gurion Univ Negev, Nucl Engn Unit, Beer Sheva, Israel
[2] Nucl Res Ctr Negev, Elect & Control Labs, Dimona, Israel
[3] Rotem Ind Ltd, Radiat Detect Dept, Dimona, Israel
来源
SCIENTIFIC REPORTS | 2024年 / 14卷 / 01期
关键词
Neutron detection; LiF:ZnS(Ag); Silicon photomultiplier; WLS FIBERS; READ-OUT; HE-3;
D O I
10.1038/s41598-024-83129-5
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
This study proposes a novel, highly sensitive neutron detector design utilizing a unique multi-layered configuration. Each layer consists of a LiF: ZnS(Ag) scintillator coupled with a transparent neutron moderator that also functions as a light guide for the Silicon Photomultiplier (SiPM) light sensor. This design offers a cost-effective and readily available alternative for existing neutron detectors. The research focused on optimizing a single layer for maximum sensitivity and proper gamma rejection capabilities. This optimized configuration is then replicated to form the multi-layer detector. A primary challenge with the LiF:ZnS(Ag) scintillator is its inherent opacity, limiting its width and consequently, its detection efficiency. Our proposed multi-layer structure addresses this limitation by employing a thin scintillator in each layer. This strategic design minimizes emitted light attenuation while simultaneously enhancing sensitivity through the cumulative effect of multiple layers. Our experiments demonstrate a significant improvement in detection efficiency compared to the single-layer setup. Additionally, our architecture offers an actual improvement in differentiating between gamma and neutron signals. By analyzing count rates across the detector's layers, we gain valuable operational insights, such as the ability to predict the source direction. Our finding demonstrates an improved sensitivity achieved by minimal loss of neutrons during the moderation of 329% compared to a single layer, aligned with the potential range of improvement while maintaining extremely high gamma rejection. Supported by the presented findings, this design represents a noticeable advancement over existing solutions, offering scalable customization to user requirements. Notably, it outperforms traditional He-3 tube-based detection configurations, positioning it as a compelling and advantageous replacement option.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] CHARACTERISTICS OF ZnS(Ag) + 6LiF SCINTILLATOR-BASED DETECTOR USED AS A NEUTRON DOSIMETER
    Kolesnikov, S. V.
    Denisenko, A. P.
    Ryabeva, E. V.
    ATOMIC ENERGY, 2023, 133 (04) : 212 - 216
  • [2] Characteristics of ZnS(Ag) + 6LiF Scintillator-Based Detector Used as a Neutron Dosimeter
    S. V. Kolesnikov
    A. P. Denisenko
    E. V. Ryabeva
    Atomic Energy, 2023, 133 : 212 - 216
  • [3] A new type of thermal-neutron detector based on ZnS(Ag)/LiF scintillator and avalanche photodiodes
    Marin, V. N.
    Sadykov, R. A.
    Trunov, D. N.
    Litvin, V. S.
    Aksenov, S. N.
    Stolyarov, A. A.
    TECHNICAL PHYSICS LETTERS, 2015, 41 (09) : 912 - 914
  • [4] A new type of thermal-neutron detector based on ZnS(Ag)/LiF scintillator and avalanche photodiodes
    V. N. Marin
    R. A. Sadykov
    D. N. Trunov
    V. S. Litvin
    S. N. Aksenov
    A. A. Stolyarov
    Technical Physics Letters, 2015, 41 : 912 - 914
  • [5] Selection of silicon photomultipliers for a (LiF)-Li-6:ZnS(Ag) scintillator based cold neutron detector
    Osovizky, A.
    Pritchard, K.
    Yehuda-Zada, Y.
    Ziegler, J. B.
    Tsai, P.
    Ghelman, M.
    Thompson, A. K.
    Ibberson, R. M.
    Baltic, G. M.
    Majkrzak, C. F.
    Maliszewskyj, N. C.
    JOURNAL OF PHYSICS COMMUNICATIONS, 2018, 2 (04):
  • [6] Development of a Coated-Micro-Particle Neutron Detector Based on LiF/ZnS Scintillator
    Logoglu, Faruk
    Flaska, Marek
    Wolfe, Douglas E.
    ADVANCEMENTS IN NUCLEAR INSTRUMENTATION MEASUREMENT METHODS AND THEIR APPLICATIONS (ANIMMA 2019), 2020, 225
  • [7] Study on Performance of 6LiF/ZnS(Ag) Scintillation Screen for Scintillator Neutron Detector in China Spallation Neutron Source
    Jiang J.
    Huang C.
    Wei Y.
    Tang B.
    Yu Q.
    Zhou S.
    Cai X.
    Deng Q.
    Yuanzineng Kexue Jishu/Atomic Energy Science and Technology, 2023, 57 (08): : 1650 - 1657
  • [8] High-Sensitivity Activation Neutron Detector
    Antropov, S. Y.
    Ermilov, A. P.
    Measurement Techniques (English translation of Izmeritel'naya Tekhnika), 38 (02):
  • [9] HIGH-SENSITIVITY ACTIVATION NEUTRON DETECTOR
    ANTROPOV, SY
    ERMILOV, AP
    MEASUREMENT TECHNIQUES USSR, 1995, 38 (02): : 227 - 228
  • [10] Novel high-efficiency 2D position-sensitive ZnS:Ag/6LiF scintillator detector for neutron diffraction
    Mauri, Giacomo
    Sykora, G. Jeff
    Schooneveld, E. M.
    Capelli, S. C.
    Gutmann, M. J.
    Howarth, S.
    Mann, S. E.
    Zuddas, F.
    Rhodes, N. J.
    JOURNAL OF APPLIED CRYSTALLOGRAPHY, 2024, 57 : 690 - 699