Estimation of weighted computed tomography dose index (CTDIw) in megavoltage computed tomography (MVCT)

被引:1
作者
Devi, Rama [1 ,2 ]
Singh, Balbir [3 ]
Basandrai, Deepak [1 ]
Passi, Kamlesh [2 ]
Singh, Vikram [2 ]
机构
[1] Lovely Profess Univ, Sch Phys Sci & Chem Engn, Dept Phys, Phagwara 144411, Punjab, India
[2] Mohan Dai Oswal Canc Hosp, Res Fdn, Dept Radiat Oncol, Ludhiana 141009, Punjab, India
[3] Shri Lal Bahadur Shastri Govt Med Coll Nerchowk, Dept Radiat Oncol, Mandi 175021, Himachal Prades, India
关键词
Weighted computed tomography dose index; Flattening filter-free; Megavoltage computed tomography; Kilovoltage computed tomography; Dose length product; HELICAL TOMOTHERAPY; RADIOTHERAPY;
D O I
10.1186/s43055-024-01374-0
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
IntroductionAdvanced technologies in radiation treatment deliveries require high accuracy and precision in setup before treatment to get accurate dose delivery for planned target volumes. During imaging procedures, patients are getting extra doses from kilovoltage computed tomography (KVCT) and megavoltage computed tomography (MVCT). Computed tomography dose index is an important dosimetric parameter to know the degree of radiation-induced side effects.AimThe main aim of this study is to quantify the dose received by patients during MVCT imaging procedures, weighted computed tomography dose index (CTDIw), volume computed tomography dose index (CTDIvol) and dose length product (DLP).Methods and materialsThis study has been performed on Radixact X9 LINAC machine with 6MV flattening filter-free (FFF) photon beam for treatment and imaging with 3.5 MeV average photon energy. CTDIw and dose length product (DLP) have been calculated for different scanning lengths 18 cm, 10 cm and 0.7 cm with fine, normal and coarse modes with standard imaging A1SL1 ion chamber, cheese phantom and tomoelectrometer.ResultsCTDIw, CTDIvol, and DLP have been calculated with cheese phantom using A1SL1 ion chamber. CTDIw for 0.7 cm is 2.16 cGy, 1.14 cGy and 0.77 cGy for fine normal and coarse modes, respectively.ConclusionCTDIw is a dosimetric index to know the dose levels at center and periphery of patient according to scan length. It gives an idea about how much additional dose is received by the patient during the imaging procedures. High pitch gives the lesser absorbed dose. DLP gives the knowledge about stochastic effects from radiation to patients after MVCT.
引用
收藏
页数:8
相关论文
共 23 条
[1]  
[Anonymous], 2014, AAPM Report No. 220
[2]   Intrafractional prostate motion during online image guided intensity-modulated radiotherapy for prostate cancer [J].
Budiharto, Tom ;
Slagmolen, Pieter ;
Haustermans, Karin ;
Maes, Frederik ;
Junius, Sara ;
Verstraete, Jan ;
Oyen, Raymond ;
Hermans, Jeroen ;
Van den Heuvel, Frank .
RADIOTHERAPY AND ONCOLOGY, 2011, 98 (02) :181-186
[3]   Quantitative characterization of tomotherapy MVCT dosimetry [J].
Chen, Mingli ;
Chao, Edward ;
Lu, Weiguo .
MEDICAL DOSIMETRY, 2013, 38 (03) :280-286
[4]   Image quality and dose evaluation of MVCT TomoTherapy acquisitions: A phantom study [J].
De Marco, P. ;
Osman, I. Abdi ;
Castellini, F. ;
Ricotti, R. ;
Leonardi, M. C. ;
Miglietta, E. ;
Cambria, R. ;
Origgi, D. ;
Jereczek-Fossa, B. A. ;
Garibaldi, C. ;
Cattani, F. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 57 :200-206
[5]   Performance characteristics of a novel megavoltage cone-beam-computed tomography device [J].
Fast, M. F. ;
Koenig, T. ;
Oelfke, U. ;
Nill, S. .
PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (03) :N15-N24
[6]   Enhanced efficiency in helical tomotherapy quality assurance using a custom-designed water-equivalent phantom [J].
Goddu, S. Murty ;
Mutic, Sasa ;
Pechenaya, Olga L. ;
Chaudhari, Summer R. ;
Garcia-Ramirez, Jose ;
Rangaraj, Dharanipathy ;
Klein, Eric E. ;
Yang, Deshan ;
Grigsby, James ;
Low, Daniel A. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (19) :5663-5674
[7]   Potential of image-guidance, gating and real-time tracking to improve accuracy in pulmonary stereotactic body radiotherapy [J].
Guckenberger, Matthias ;
Krieger, Thomas ;
Richter, Anne ;
Baier, Kurt ;
Wilbert, Juergen ;
Sweeney, Reinhart A. ;
Flentje, Michael .
RADIOTHERAPY AND ONCOLOGY, 2009, 91 (03) :288-295
[8]   QA for helical tomotherapy: Report of the AAPM Task Group 148 [J].
Langen, Katja M. ;
Papanikolaou, Niko ;
Balog, John ;
Crilly, Richard ;
Followill, David ;
Goddu, S. Murty ;
Grant, Walter, III ;
Olivera, Gustavo ;
Ramsey, Chester R. ;
Shi, Chengyu .
MEDICAL PHYSICS, 2010, 37 (09) :4817-4853
[9]   Image guidance for precise conformal radiotherapy [J].
Mackie, TR ;
Kapatoes, J ;
Ruchala, K ;
Lu, WG ;
Wu, C ;
Olivera, G ;
Forrest, L ;
Tome, W ;
Welsh, J ;
Jeraj, R ;
Harari, P ;
Reckwerdt, P ;
Paliwal, B ;
Ritter, M ;
Keller, H ;
Fowler, J ;
Mehta, M .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 56 (01) :89-105
[10]   TOMOTHERAPY - A NEW CONCEPT FOR THE DELIVERY OF DYNAMIC CONFORMAL RADIOTHERAPY [J].
MACKIE, TR ;
HOLMES, T ;
SWERDLOFF, S ;
RECKWERDT, P ;
DEASY, JO ;
YANG, J ;
PALIWAL, B ;
KINSELLA, T .
MEDICAL PHYSICS, 1993, 20 (06) :1709-1719