A novel technique of blindly positioning bronchial blockers for one-lung ventilation: a prospective, randomized, crossover study

被引:0
|
作者
Min, Seihee [1 ]
Yoon, Susie [2 ]
Han, Jiwon [1 ]
Seo, Jeong-Hwa [2 ]
Bahk, Jae-Hyon [2 ]
机构
[1] Chung Ang Univ, Coll Med, Gwangmyeong Hosp, Dept Anesthesiol & Pain Med, 110 Deokan Ro, Gwangmyeong Si 14353, Gyeonggi Do, South Korea
[2] Seoul Natl Univ, Seoul Natl Univ Hosp, Coll Med, Dept Anesthesiol & Pain Med, 101 Daehak Ro, Seoul 03080, South Korea
关键词
Airway management; Lung isolation; One-lung ventilation; Bronchial blocker; Thoracic surgery; DOUBLE-LUMEN TUBE; ENDOBRONCHIAL BLOCKER; THORACIC-SURGERY; TRACHEOSTOMY TUBE; EZ-BLOCKER; PATIENT; EZ-BLOCKER(R); PLACEMENT; DEVICE; TRIAL;
D O I
10.1186/s13019-024-03276-8
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Several methods for blindly positioning bronchial blockers (BBs) for one-lung ventilation (OLV) have been proposed. However, these methods do not reliably ensure accurate positioning and proper direction. Here, we developed a clinically applicable two-stage maneuver by modifying a previously reported one-stage maneuver for successful insertion of a BB at the appropriate depth and direction in patients requiring lung isolation where a flexible bronchoscope (FOB) is not applicable. Methods This prospective, randomized, crossover study was conducted at a tertiary university hospital and included 94 patients requiring OLV for elective thoracic surgery under general anesthesia. The patients underwent the one-stage maneuver followed by the two-stage maneuver and vice versa, and the success rates of the two methods were compared. After tracheal intubation, the deflated rigid-angle BB was inserted into the endotracheal tube (ET) until a pre-marked point indicating that the BB was just protruding from the ET tip. To identify the carinal depth without FOB, the BB balloon was inflated and advanced toward the intended side whilst monitoring abrupt changes in peak inspiratory pressure and expiratory tidal volume to indicate placement at the carina. In one-stage maneuver, the BB balloon was deflated and advanced 3 cm further from the estimated carinal depth. During the two-stage maneuver, the same procedure was performed to determine the carinal depth, and the deflated BB was withdrawn and reinserted to a predetermined depth with its tip directed 90 degree toward the target bronchus. The accuracy of BB positioning for both maneuvers was evaluated by a 4-point scale, with grades 1 and 2 considered acceptable for providing OLV. Results BB placement was more accurate in the two-stage maneuver than in the one-stage maneuver [88.0% (81/92) vs. 73.9% (68/92), relative risk (95% confidence interval [CI]), 0.45 (0.23-0.88), P < 0.001]. This improvement was particularly significant when targeting the left main bronchus [84.8% (39/46) vs. 58.7% (27/46), relative risk (95% CI), 0.36 (0.17-0.79), P < 0.001)]. Conclusions The novel two-stage maneuver significantly improved the success rate of blindly positioning the BB. Trial registration This study was registered in ClinicalTrials.gov (NCT02981537) on December 05, 2016.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] A Clinical Comparison of 2 Bronchial Blockers Versus Double-Lumen Tubes for One-Lung Ventilation
    Shum, Serena
    Garijo, Jacobo Moreno
    Tomlinson, George
    Rodrigues, Jonathan
    Greyling, Gerhard
    Shafiepour, Danielle
    McRae, Karen
    Slinger, Peter
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2023, 37 (12) : 2577 - 2583
  • [2] Pro: Bronchial Blockers Should Be Used Routinely for Providing One-Lung Ventilation
    Neustein, Steven M.
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2015, 29 (01) : 234 - 236
  • [3] The Use of Bronchial Blockers for Providing One-Lung Ventilation
    Neustein, Steven M.
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2009, 23 (06) : 860 - 868
  • [4] Effects of hypercarbia on arterial oxygenation during one-lung ventilation: prospective randomized crossover study
    Lee, Jun Ho
    Kim, Yesull
    Mun, Juhan
    Lee, Joseph
    Ko, Seonghoon
    KOREAN JOURNAL OF ANESTHESIOLOGY, 2020, 73 (06) : 534 - 541
  • [5] One-lung ventilation with a bronchial blocker in thoracic patients
    Cano, Paulo Andres
    Mora, Luis Carlos
    Enriquez, Irene
    Reis, Matias Santiago
    Martinez, Eva
    Barturen, Fernando
    BMC ANESTHESIOLOGY, 2023, 23 (01)
  • [6] Systematic Review and Meta-Analysis of Efficiency and Safety of Double-Lumen Tube and Bronchial Blocker for One-Lung Ventilation
    Palaczynski, Piotr
    Misiolek, Hanna
    Szarpak, Lukasz
    Smereka, Jacek
    Pruc, Michal
    Rydel, Mateusz
    Czyzewski, Damian
    Bialka, Szymon
    JOURNAL OF CLINICAL MEDICINE, 2023, 12 (05)
  • [7] Effects of the disconnection technique and preemptive one-lung ventilation on lung collapse during one-lung ventilation in thoracoscopic surgery
    Zhang, Hongru
    Xiang, Silin
    Mei, Longyong
    Feng, Yonggeng
    She, Han
    Hu, Yi
    Wang, Li
    BMC ANESTHESIOLOGY, 2025, 25 (01):
  • [8] One-lung ventilation with a bronchial blocker in thoracic patients
    Paulo Andrés Cano
    Luis Carlos Mora
    Irene Enríquez
    Matías Santiago Reis
    Eva Martínez
    Fernando Barturen
    BMC Anesthesiology, 23
  • [9] A new bronchial blocker tube for one-lung ventilation
    Kraenzler, EJ
    Insler, SR
    Grubb, G
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 1999, 13 (01) : 40 - 41
  • [10] Effects of Positive End-Expiratory Pressure on Pulmonary Oxygenation and Biventricular Function during One-Lung Ventilation: A Randomized Crossover Study
    Kim, Namo
    Lee, Su Hyun
    Choi, Kwan Woong
    Lee, Haeyeon
    Oh, Young Jun
    JOURNAL OF CLINICAL MEDICINE, 2019, 8 (05)