Agreement between a new fully automatic ocular biometer based on optical low-coherence reflectometry and an optical biometer based on Scheimpflug imaging combined with partial coherence interferometry
被引:1
|作者:
Ning, Rui
[1
,2
,3
,4
]
Xu, Huilin
[2
,3
,5
]
Li, Zheng
[1
,2
,3
,4
]
Yu, Jinjin
[6
]
Xu, Shuoyu
[1
,2
,3
,4
]
Lei, Chak Seng
[1
,2
,3
,4
]
Wang, Yiran
[1
,2
,3
,4
]
Savini, Giacomo
[7
]
Schiano-Lomoriello, Domenico
[7
]
Zhou, Xingtao
[1
,2
,3
,4
]
Huang, Jinhai
[1
,2
,3
,4
]
机构:
[1] Fudan Univ, Eye & ENT Hosp, Eye Inst, Shanghai, Peoples R China
[2] Fudan Univ, Eye & ENT Hosp, Dept Ophthalmol, NHC Key Lab Myopia & Related Eye Dis, Shanghai, Peoples R China
[3] Chinese Acad Med Sci, Key Lab Myopia & Related Eye Dis, Shanghai, Peoples R China
[4] Shanghai Res Ctr Ophthalmol & Optometry, Shanghai, Peoples R China
[5] Wenzhou Med Univ, Eye Hosp, Sch Ophthalmol & Optometry, Wenzhou, Zhejiang, Peoples R China
[6] Ningbo 2 Hosp, Ningbo, Zhejiang, Peoples R China
[7] GB Bietti Fdn IRCCS, Rome, Italy
基金:
中国国家自然科学基金;
关键词:
Optical low coherence reflectometry;
Scheimpflug Imaging;
Partial coherence interferometry;
Biometry;
ANTERIOR SEGMENT MEASUREMENTS;
ASTIGMATISM ANALYSIS;
REPEATABILITY;
PENTACAM;
LENSTAR;
DEVICES;
POWER;
REPRODUCIBILITY;
COMPARABILITY;
TOPOGRAPHER;
D O I:
10.1186/s12886-024-03716-y
中图分类号:
R77 [眼科学];
学科分类号:
100212 ;
摘要:
BackgroundTo analyze the difference and agreement between measurements obtained by a new fully automatic optical biometer, the SW-9000 mu m Plus, based on optical low-coherence reflectometry (OLCR) and a commonly used optical biometer (Pentacam AXL) based on Scheimpflug imaging with partial coherence interferometry (PCI).MethodsThe central corneal thickness (CCT), anterior chamber depth (ACD, from epithelium to anterior lens surface), lens thickness (LT), mean keratometry (Km), corneal astigmatism, corneal diameter (CD), pupil diameter (PD), and axial length (AL) of 74 eyes (from 74 healthy subjects) were measured using the SW-9000 mu m Plus and the Pentacam AXL to determine the agreement. Double angle plots were used for astigmatism vector analysis. Bland-Altman and 95% limits of agreement (LoA) were calculated.ResultsStatistically significant differences were detected for all parameters but J0 vector. The Bland-Altman analysis of AL, CCT, ACD, Km, CD, J0 and J45 indicated a high level of agreement between the two devices. Among AL, CCT, ACD, Km, J0, J45, CD, and PD, the 95% LoA ranged from -0.07 to 0.05 mm, -9.67 to 7.34 mm, -0.11 to 0.04 mm, -0.25 to 0.50 D, -0.22 to 0.20 D, -0.15 to 0.20 D, -0.23 to 0.35 mm and 1.55 to 3.77 mm, respectively.ConclusionsThe measurements of AL, CCT, ACD, Km, corneal astigmatism, and CD showed a narrow LoA and may be used interchangeably in healthy subjects between the new OLCR optical biometer and the Scheimpflug/PCI biometer; however, a poor agreement was noted for PD values.