Evaluation of cup placement accuracy in computer assisted total hip arthroplasty

被引:0
作者
Kaneta, Hiroki [1 ]
Shoji, Takeshi [2 ]
Ueki, Shinichi [1 ]
Morita, Hiroyuki [1 ]
Kozuma, Yosuke [1 ]
Adachi, Nobuo [1 ]
机构
[1] Hiroshima Univ, Grad Sch Biomed & Hlth Sci, Dept Orthopaed Surg, Hiroshima, Japan
[2] Hiroshima Univ, Grad Sch Biomed & Hlth Sci, Dept Artificial Joints & Biomat, Hiroshima, Japan
关键词
Total hip arthroplasty; Computer-assisted surgery; Robotic system; CT navigation; Cup placement accuracy; Patient-reported outcome measures; NAVIGATION; ORIENTATION; COMPONENT; RANGE; IMPLANTATION; ALIGNMENT; ROBOTICS; MOTION;
D O I
10.1007/s00402-025-05797-w
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
BackgroundTotal hip arthroplasty (THA) accuracy has improved significantly with various advances in computer-assisted equipment (CAE), including robotic systems, computed tomography (CT) navigation, and portable navigation. However, no studies have directly compared the accuracy of acetabular cup placement and its impact on patient-reported outcome measures (PROMs) across these three CAE systems. In this study, we aimed to evaluate cup placement accuracy and PROMs in THA using different CAE systems.MethodsThis retrospective analysis included 196 patients (202 hip joints) who underwent THA with three CAE systems from May 2021 to August 2023. Patients were categorized into the robotic system (73 hips), CT navigation (83 hips), and portable navigation (46 hips). Postoperative CT scans measured cup placement angles-radiographic inclination (RI) and radiographic anteversion (RA) -and compared them with preoperative target angles. Anterior-posterior (AP) cup position differences were evaluated by measuring the distance between the acetabular and cup center in the axial view of the postoperative CT scans. PROMs were evaluated using the Japanese Orthopaedic Association Hip Disease Evaluation Questionnaire (JHEQ) at 3 and 12 months.ResultsDemographic characteristics, including age, sex, primary disease, and Body Mass Index, were similar across groups. The robotic system exhibited significantly smaller deviations in Delta RI and Delta RA compared to CT navigation and portable navigation. AP cup position differences were also smaller in the robotic system versus portable navigation; however, the difference between the robotic and CT navigation systems was not statistically significant. Despite the superior precision of cup placement in the robotic system, no significant differences in JHEQ scores were observed among the groups at 3 and 12 months.ConclusionRobotic systems demonstrated superior accuracy in cup placement. However, short-term PROMs did not significantly differ, suggesting that PROMs may not solely depend on accurate cup placement. Future research should investigate additional factors influencing PROMs.
引用
收藏
页数:10
相关论文
共 52 条
[1]   Comparison of the accuracy of the cup position and orientation in total hip arthroplasty for osteoarthritis secondary to developmental dysplasia of the hip between the Mako robotic arm-assisted system and computed tomography-based navigation [J].
Ando, Wataru ;
Takao, Masaki ;
Hamada, Hidetoshi ;
Uemura, Keisuke ;
Sugano, Nobuhiko .
INTERNATIONAL ORTHOPAEDICS, 2021, 45 (07) :1719-1725
[2]   Abductor muscle strengthening in THA patients operated with minimally-invasive anterolateral approach for developmental hip dysplasia [J].
Benedetti, Maria Grazia ;
Cavazzuti, Lorenzo ;
Amabile, Marilina ;
Tassinari, Enrico ;
Valente, Giordano ;
Zanotti, Gabriele ;
Vaienti, Enrico ;
Orsini, Stefania ;
Mariani, Elisabetta ;
Taddei, Fulvia .
HIP INTERNATIONAL, 2021, 31 (01) :66-74
[3]   Robotic-assisted compared with conventional total hip arthroplasty: systematic review and meta-analysis [J].
Chen, Xi ;
Xiong, Jianping ;
Wang, Peipei ;
Zhu, Shibai ;
Qi, Wenting ;
Peng, Huiming ;
Yu, Lingjia ;
Qian, Wenwei .
POSTGRADUATE MEDICAL JOURNAL, 2018, 94 (1112) :335-341
[4]   The effect of the orientation of the acetabular and femoral components on the range of motion of the hip at different head-neck ratios [J].
D'Lima, DD ;
Urquhart, AG ;
Buehler, KO ;
Walker, RH ;
Colwell, CW .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (03) :315-321
[5]   Comparison of Robotic-assisted and Conventional Acetabular Cup Placement in THA: A Matched-pair Controlled Study [J].
Domb, Benjamin G. ;
El Bitar, Youssef F. ;
Sadik, Adam Y. ;
Stake, Christine E. ;
Botser, Itamar B. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2014, 472 (01) :329-336
[6]  
Elmallah Randa K, 2015, Surg Technol Int, V26, P283
[7]   Precision of acetabular cup placement in robotic integrated total hip arthroplasty [J].
Elson, Leah ;
Dounchis, Jon ;
Illgen, Richard ;
Marchand, Robert C. ;
Padgett, Douglas E. ;
Bragdon, Charles R. ;
Malchau, Henrik .
HIP INTERNATIONAL, 2015, 25 (06) :531-536
[8]   Is there increased value in robotic arm-assisted total hip arthroplasty? A NATIONWIDE OUTCOMES, TRENDS, AND PROJECTIONS ANALYSIS OF 4,699,894 CASES [J].
Emara, A. K. ;
Zhou, G. ;
Klika, A. K. ;
Koroukian, S. M. ;
Schiltz, N. K. ;
Higuera-Rueda, C. A. ;
Molloy, R. M. ;
Piuzzi, N. S. .
BONE & JOINT JOURNAL, 2021, 103B (09) :1488-1496
[9]  
Feng X, 2022, AM J TRANSL RES, V14, P7080
[10]   Effect of screw fixation on acetabular component alignment change in total hip arthroplasty [J].
Fujishiro, Takaaki ;
Hayashi, Shinya ;
Kanzaki, Noriyuki ;
Hashimoto, Shingo ;
Shibanuma, Nao ;
Kurosaka, Masahiro .
INTERNATIONAL ORTHOPAEDICS, 2014, 38 (06) :1155-1158