Simultaneous co-axial multi-modal inspection using a laser driven x-ray and neutron source

被引:1
|
作者
Armstrong, C. D. [1 ]
Scott, G. G. [1 ]
Richards, S. [2 ]
Fedorov, K. [1 ]
Laker-Illoul, A. [1 ]
Hull, G. [1 ]
Patel, J. K. [1 ,3 ]
Booth, N. [1 ]
Clarke, R. [1 ]
Rajeev, P. P. [1 ]
Simpson, R. [4 ]
Ma, T. [4 ]
Hernandez-Gomez, C. [1 ]
机构
[1] Rutherford Appleton Lab, Cent Laser Facil, Harwell OX11 6FQ, England
[2] Rutherford Appleton Lab, STFC Technol, Harwell OX11 6FQ, England
[3] Univ Strathclyde, SUPA Dept Phys, Glasgow G4 0NG, Scotland
[4] Lawrence Livermore Natl Lab, Livermore, CA USA
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2024年 / 95卷 / 10期
关键词
Compendex;
D O I
10.1063/5.0199999
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Laser-plasma interactions have been demonstrated to produce bright sources of energetic radiation including ions, electrons, photons across the electro-magnetic spectrum, and neutrons. Combinations of species can significantly increase information from non-destructive imaging. Here we demonstrate single-shot co-axial radiography with both x-ray and fast-neutron radiation from a laser-driven source using a pair of gated microchannel plate photomultiplier tube channels and a fast scintillator medium. The outlined system demonstrates recovery full-width-half-maximum of (18 +/- 3) ns, which is sufficient to isolate x-rays from neutrons up to (72 +/- 20) MeV and could be isolated only a short distance (2 m) from the target.
引用
收藏
页数:7
相关论文
共 2 条
  • [1] Demonstration of a robust laser-driven X-ray source for CT
    ALPhANOV, Rue François Mitterand, Talence
    33400, France
    不详
    33600, France
    不详
    F-33400, France
    eJ. Nondestruct. Test., Special Issue
  • [2] CONDITIONS FOR USING A LASER-STIMULATED X-RAY SOURCE IN CONTACT LITHOGRAPHY.
    Bokov, Yu.S.
    Kas'yanov, Yu.S.
    Korobkin, V.V.
    Leonov, Yu.S.
    Mishachev, V.I.
    Soviet physics. Technical physics, 1982, 27 (03): : 346 - 347