Automatic Image Registration Provides Superior Accuracy Compared with Surface Matching in Cranial Navigation

被引:0
|
作者
Frisk, Henrik [1 ,2 ]
Jensdottir, Margret [1 ,2 ]
Coronado, Luisa [3 ]
Conrad, Markus [3 ]
Hager, Susanne [3 ]
Arvidsson, Lisa [1 ,2 ]
Bartek Jr, Jiri [1 ,2 ]
Burstroem, Gustav [1 ,2 ]
El-Hajj, Victor Gabriel [1 ]
Edstroem, Erik [1 ,4 ]
Elmi-Terander, Adrian [1 ,4 ,5 ]
Persson, Oscar [1 ,2 ]
机构
[1] Karolinska Inst, Dept Clin Neurosci, SE-17177 Stockholm, Sweden
[2] Karolinska Univ Hosp, Dept Neurosurg, SE-17176 Stockholm, Sweden
[3] Brainlab AG, Clin Affairs, D-81829 Munich, Germany
[4] Lowenstromska Hosp, Capio Spine Ctr Stockholm, SE-19489 Upplands Vasby, Sweden
[5] Uppsala Univ, Dept Surg Sci, SE-75236 Uppsala, Sweden
关键词
patient tracking; reference frame; surface matching; CBCT; neurosurgery; surgical navigation; automatic image registration; accuracy; GUIDED NEUROSURGERY; PATIENT;
D O I
10.3390/s24227341
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Objective: The precision of neuronavigation systems relies on the correct registration of the patient's position in space and aligning it with radiological 3D imaging data. Registration is usually performed by the acquisition of anatomical landmarks or surface matching based on facial features. Another possibility is automatic image registration using intraoperative imaging. This could provide better accuracy, especially in rotated or prone positions where the other methods may be difficult to perform. The aim of this study was to validate automatic image registration (AIR) using intraoperative cone-beam computed tomography (CBCT) for cranial neurosurgical procedures and compare the registration accuracy to the traditional surface matching (SM) registration method based on preoperative MRI. The preservation of navigation accuracy throughout the surgery was also investigated. Methods: Adult patients undergoing intracranial tumor surgery were enrolled after consent. A standard SM registration was performed, and reference points were acquired. An AIR was then performed, and the same reference points were acquired again. Accuracy was calculated based on the referenced and acquired coordinates of the points for each registration method. The reference points were acquired before and after draping and at the end of the procedure to assess the persistency of accuracy. Results: In total, 22 patients were included. The mean accuracy was 6.6 +/- 3.1 mm for SM registration and 1.0 +/- 0.3 mm for AIR. The AIR was superior to the SM registration (p < 0.0001), with a mean improvement in accuracy of 5.58 mm (3.71-7.44 mm 99% CI). The mean accuracy for the AIR registration pre-drape was 1.0 +/- 0.3 mm. The corresponding accuracies post-drape and post-resection were 2.9 +/- 4.6 mm and 4.1 +/- 4.9 mm, respectively. Although a loss of accuracy was identified between the preoperative and end-of-procedure measurements, there was no statistically significant decline during surgery. Conclusions: AIR for cranial neuronavigation consistently delivered greater accuracy than SM and should be considered the new gold standard for patient registration in cranial neuronavigation. If intraoperative imaging is a limited resource, AIR should be prioritized in rotated or prone position procedures, where the benefits are the greatest.
引用
收藏
页数:12
相关论文
共 50 条
  • [31] Surface matching of multimodality image volumes by a fuzzy elastic registration technique
    Butt, Amer
    Acharya, Raj
    Sibata, Claudio
    Shin, K.H.
    Computerized Medical Imaging and Graphics, 22 (01): : 13 - 23
  • [32] Surface matching of multimodality image volumes by a fuzzy elastic registration technique
    Butt, A
    Acharya, R
    Sibata, C
    Shin, KH
    COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 1998, 22 (01) : 13 - 23
  • [33] Anatomical surface contour matching for efficient registration in image guided neurosurgery
    Bucholz, R
    Macneil, WR
    Henderson, J
    CAR '97 - COMPUTER ASSISTED RADIOLOGY AND SURGERY, 1997, 1134 : 772 - 777
  • [34] Automatic multi-sensor image registration by edge matching using genetic algorithms
    Inglada, J
    Adragna, F
    IGARSS 2001: SCANNING THE PRESENT AND RESOLVING THE FUTURE, VOLS 1-7, PROCEEDINGS, 2001, : 2313 - 2315
  • [35] An automatic evaluation method for retinal image registration based on similar vessel structure matching
    Shu, Yifan
    Feng, Yunlong
    Wu, Guannan
    Kang, Jieliang
    Li, Huiqi
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2020, 58 (01) : 117 - 129
  • [36] An automatic evaluation method for retinal image registration based on similar vessel structure matching
    Yifan Shu
    Yunlong Feng
    Guannan Wu
    Jieliang Kang
    Huiqi Li
    Medical & Biological Engineering & Computing, 2020, 58 : 117 - 129
  • [37] Automatic image registration method based on SIFT and Harris-Affine features matching
    School of Mathematical Science, Xiamen University, Xiamen 361005, China
    不详
    不详
    不详
    Huazhong Ligong Daxue Xuebao, 2008, 8 (13-16):
  • [38] An optimized method for automatic registration of remotely sensed images based on structural image matching
    Zhang, Jingtao
    Cheng, Bo
    PROCEEDINGS OF THE 2009 2ND INTERNATIONAL CONGRESS ON IMAGE AND SIGNAL PROCESSING, VOLS 1-9, 2009, : 2898 - 2902
  • [39] Accuracy and practicability of laser surface scanning for registration in image guided neurosurgery
    Krishnan, R
    Raabe, A
    Seifert, V
    PERSPECTIVE IN IMAGE-GUIDED SURGERY, 2004, : 31 - 36
  • [40] Image overlay navigation by markerless surface registration in gastrointestinal, hepatobiliary and pancreatic surgery
    Sugimoto, Maki
    Yasuda, Hideki
    Koda, Keiji
    Suzuki, Masato
    Yamazaki, Masato
    Tezuka, Tohru
    Kosugi, Chihiro
    Higuchi, Ryota
    Watayo, Yoshihisa
    Yagawa, Yohsuke
    Uemura, Shuichiro
    Tsuchiya, Hironori
    Azuma, Takeshi
    JOURNAL OF HEPATO-BILIARY-PANCREATIC SCIENCES, 2010, 17 (05) : 629 - 636