Evaluation of 3D-CRT Radiotherapy Treatment Planning with Radiobiological Effects in Patient-Specific 3D Printed Anthropomorphic Phantom Postmastectomy Breast Cancer Axillary Metastases

被引:0
作者
Endarko, Endarko [1 ]
Hariyanto, Aditya Prayugo [1 ]
Aisyah, Siti [1 ]
Kavilani, Nandia [2 ]
Syafi’i, Ahmad [2 ]
机构
[1] Department of Physics, Institut Teknologi Sepuluh Nopember, Kampus ITS, Sukolilo Surabaya, East Java
[2] Medical Physicist of Radiotherapy Installation, Naval Hospital Ramelan, Surabaya, East Java
关键词
Breast Cancer; Conformal Radiation; Dosimetry Radiation; Phantom; Protection; Radiation Therapy; Therapy (3D-CRT); Three-Dimensional;
D O I
10.22038/ijmp.2024.73930.2309
中图分类号
学科分类号
摘要
Introduction: The aim of this study was to evaluate and measure the effectiveness of a particular radiotherapy treatment planning system, namely 3D conformal radiation therapy (3D-CRT), and radiobiological effects in postmastectomy breast cancer axillary lymph metastases using a patient-specific 3D printed anthropomorphic phantom of the right breast after surgery. Material and Methods: The chest wall and axilla were the two treatment targets for 3D-CRT planning techniques. Each target was subjected to variations of two and three irradiation fields, with a dose of 50.00 Gy and fractionation of 2.00 Gy. From the variations, the radiation dose received by planning target volume (PTV) chest wall, PTV-Axillary, and organs at risk (OARs) was evaluated to determine the best planning technique through dosimetry verification. The thermoluminescent dosimeter (TLD) 100 Chip and EBT3 film were employed to measure the radiation dose at the PTV and OARs. Results: The planning system evaluation based on dose volume histogram showed no significant difference (p = 0.993) between two and three irradiation fields in both the chest wall and axillary planning 3D-CRT technique. Normal tissue complication probability for OARs has values below 5%, while the smallest value was obtained for the left lung. Finally, the point dose verification between planning and measurement using TLD100 and EBT3 film indicated an average difference of 9.14% and 4.34%, respectively, with no significant difference. Conclusion: This study successfully demonstrated the dosimetry evaluation and radiobiological effects of postmastectomy right breast cancer treatment using a patient-specific 3D printed anthropomorphic phantom, which met the established standards. © (2024), (Mashhad University of Medical Sciences). All rights reserved.
引用
收藏
页码:314 / 322
页数:8
相关论文
共 50 条
[31]   Comparison of Breast Cancer Radiotherapy Techniques Regarding Secondary Cancer Risk and Normal Tissue Complication Probability - Modelling and Measurements Using a 3D-Printed Phantom [J].
Vogel, Marc ;
Gade, Jonas ;
Timm, Bernd ;
Schuermann, Michaela ;
Auerbach, Hendrik ;
Nuesken, Frank ;
Ruebe, Christian ;
Melchior, Patrick ;
Dzierma, Yvonne .
FRONTIERS IN ONCOLOGY, 2022, 12
[32]   Heart dose reduction in breast cancer treatment with simultaneous integrated boost: Comparison of treatment planning and dosimetry for a novel hybrid technique and 3D-CRT; [Herzdosisreduktion bei der Brustkrebsbehandlung mit simultan integriertem Boost: Vergleich von Behandlungsplanung und Dosimetrie einer neuen Hybridtechnik und der 3D-CRT] [J].
Jöst V. ;
Kretschmer M. ;
Sabatino M. ;
Würschmidt F. ;
Dahle J. ;
Ueberle F. ;
Lorenzen J. .
Strahlentherapie und Onkologie, 2015, 191 (9) :734-741
[33]   Clinical added value of 3D printed patient-specific guides in orthopedic surgery (excluding knee arthroplasty): a systematic review [J].
Kampkuiper, Nick ;
ten Heggeler, Romy ;
Nellensteijn, Jorm ;
Brusse-Keizer, Marjolein ;
Tuijthof, Gabrielle ;
Koenrades, Maaike ;
Schroder, Femke .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2025, 145 (01)
[34]   The dosimetric value and safety evaluation of 3D printed bolus in adjuvant intensity-modulated radiotherapy after radical mastectomy for breast cancer: a prospective cohort study [J].
Wang, Yaxue ;
Li, Hanzong ;
Hu, Wendie ;
Li, Bohan ;
Du, Yujie ;
Li, Fengyan .
RADIATION ONCOLOGY, 2025, 20 (01)
[35]   Factors of influence on acute skin toxicity of breast cancer patients treated with standard three-dimensional conformal radiotherapy (3D-CRT) after breast conserving surgery (BCS) [J].
Uta Kraus-Tiefenbacher ;
Andreas Sfintizky ;
Grit Welzel ;
Anna Simeonova ;
Elena Sperk ;
Kerstin Siebenlist ;
Sabine Mai ;
Frederik Wenz .
Radiation Oncology, 7
[36]   Factors of influence on acute skin toxicity of breast cancer patients treated with standard three-dimensional conformal radiotherapy (3D-CRT) after breast conserving surgery (BCS) [J].
Kraus-Tiefenbacher, Uta ;
Sfintizky, Andreas ;
Welzel, Grit ;
Simeonova, Anna ;
Sperk, Elena ;
Siebenlist, Kerstin ;
Mai, Sabine ;
Wenz, Frederik .
RADIATION ONCOLOGY, 2012, 7
[37]   Patient-specific implants made of 3D printed bioresorbable polymers at the point-of-care: material, technology, and scope of surgical application [J].
Maintz, Michaela ;
Tourbier, Celine ;
de Wild, Michael ;
Cattin, Philippe C. ;
Beyer, Michel ;
Seiler, Daniel ;
Honigmann, Philipp ;
Sharma, Neha ;
Thieringer, Florian M. .
3D PRINTING IN MEDICINE, 2024, 10 (01)
[38]   A Multi-Criteria Assessment Strategy for 3D Printed Porous Polyetheretherketone (PEEK) Patient-Specific Implants for Orbital Wall Reconstruction [J].
Sharma, Neha ;
Welker, Dennis ;
Aghlmandi, Soheila ;
Maintz, Michaela ;
Zeilhofer, Hans-Florian ;
Honigmann, Philipp ;
Seifert, Thomas ;
Thieringer, Florian M. .
JOURNAL OF CLINICAL MEDICINE, 2021, 10 (16)
[39]   In vivo surface dose measurement using GafChromic film dosimetry in breast cancer radiotherapy: comparison of 7-field IMRT, tangential IMRT and tangential 3D-CRT [J].
Rudat, Volker ;
Nour, Alaa ;
Alaradi, Abdul Aziz ;
Mohamed, Adel ;
Altuwaijri, Saleh .
RADIATION ONCOLOGY, 2014, 9
[40]   In vivo surface dose measurement using GafChromic film dosimetry in breast cancer radiotherapy: comparison of 7-field IMRT, tangential IMRT and tangential 3D-CRT [J].
Volker Rudat ;
Alaa Nour ;
Abdul Aziz Alaradi ;
Adel Mohamed ;
Saleh Altuwaijri .
Radiation Oncology, 9