Evaluation of 3D-CRT Radiotherapy Treatment Planning with Radiobiological Effects in Patient-Specific 3D Printed Anthropomorphic Phantom Postmastectomy Breast Cancer Axillary Metastases

被引:0
作者
Endarko, Endarko [1 ]
Hariyanto, Aditya Prayugo [1 ]
Aisyah, Siti [1 ]
Kavilani, Nandia [2 ]
Syafi’i, Ahmad [2 ]
机构
[1] Department of Physics, Institut Teknologi Sepuluh Nopember, Kampus ITS, Sukolilo Surabaya, East Java
[2] Medical Physicist of Radiotherapy Installation, Naval Hospital Ramelan, Surabaya, East Java
关键词
Breast Cancer; Conformal Radiation; Dosimetry Radiation; Phantom; Protection; Radiation Therapy; Therapy (3D-CRT); Three-Dimensional;
D O I
10.22038/ijmp.2024.73930.2309
中图分类号
学科分类号
摘要
Introduction: The aim of this study was to evaluate and measure the effectiveness of a particular radiotherapy treatment planning system, namely 3D conformal radiation therapy (3D-CRT), and radiobiological effects in postmastectomy breast cancer axillary lymph metastases using a patient-specific 3D printed anthropomorphic phantom of the right breast after surgery. Material and Methods: The chest wall and axilla were the two treatment targets for 3D-CRT planning techniques. Each target was subjected to variations of two and three irradiation fields, with a dose of 50.00 Gy and fractionation of 2.00 Gy. From the variations, the radiation dose received by planning target volume (PTV) chest wall, PTV-Axillary, and organs at risk (OARs) was evaluated to determine the best planning technique through dosimetry verification. The thermoluminescent dosimeter (TLD) 100 Chip and EBT3 film were employed to measure the radiation dose at the PTV and OARs. Results: The planning system evaluation based on dose volume histogram showed no significant difference (p = 0.993) between two and three irradiation fields in both the chest wall and axillary planning 3D-CRT technique. Normal tissue complication probability for OARs has values below 5%, while the smallest value was obtained for the left lung. Finally, the point dose verification between planning and measurement using TLD100 and EBT3 film indicated an average difference of 9.14% and 4.34%, respectively, with no significant difference. Conclusion: This study successfully demonstrated the dosimetry evaluation and radiobiological effects of postmastectomy right breast cancer treatment using a patient-specific 3D printed anthropomorphic phantom, which met the established standards. © (2024), (Mashhad University of Medical Sciences). All rights reserved.
引用
收藏
页码:314 / 322
页数:8
相关论文
共 50 条
[1]   Dosimetry Evaluation of Treatment Planning Systems in Patient-Specific 3D Printed Anthropomorphic Phantom for Breast Cancer after Mastectomy using a Single-Beam 3D-CRT Technique for Megavoltage Electron Radiation Therapy [J].
Endarko E. ;
Aisyah S. ;
Hariyanto A.P. ;
Haekal M. ;
Kavilani N. ;
Syafi’i A. .
Journal of Biomedical Physics and Engineering, 2023, 13 (03) :217-226
[2]   Patient-Specific Quality Assurance Using a 3D-Printed Chest Phantom for Intraoperative Radiotherapy in Breast Cancer [J].
Choi, Yeonho ;
Lee, Ik Jae ;
Park, Kwangwoo ;
Park, Kyung Ran ;
Cho, Yeona ;
Kim, Jun Won ;
Lee, Ho .
FRONTIERS IN ONCOLOGY, 2021, 11
[3]   Dosimetric Comparison of IMRT with 3D-CRT Regarding Their Contribution to the Treatment Plan Optimization Using Rando Phantom with a Realistic Lung Cancer Radiotherapy Treatment Planning [J].
Benkahila K. ;
Kharfi F. ;
Boulakhssaim F. ;
Khoudri S. .
Iranian Journal of Medical Physics, 2021, 18 (03) :154-163
[4]   Dosimetric verification of cancer patient?s treatment plan using an anthropomorphic, 3D-printed phantom [J].
Waluk, Kacper ;
Pietrzak, Jakub .
APPLIED RADIATION AND ISOTOPES, 2023, 191
[5]   Evaluation of a 3D-printed heterogeneous anthropomorphic head and neck phantom for patient-specific quality assurance in intensity-modulated radiation therapy [J].
Kadoya, Noriyuki ;
Abe, Kota ;
Nemoto, Hikaru ;
Sato, Kiyokazu ;
Ieko, Yoshiro ;
Ito, Kengo ;
Dobashi, Suguru ;
Takeda, Ken ;
Jingu, Keiichi .
RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2019, 12 (03) :351-356
[6]   Evaluation of a 3D-printed heterogeneous anthropomorphic head and neck phantom for patient-specific quality assurance in intensity-modulated radiation therapy [J].
Noriyuki Kadoya ;
Kota Abe ;
Hikaru Nemoto ;
Kiyokazu Sato ;
Yoshiro Ieko ;
Kengo Ito ;
Suguru Dobashi ;
Ken Takeda ;
Keiichi Jingu .
Radiological Physics and Technology, 2019, 12 :351-356
[7]   Evaluation of 3D Printed Anthropomorphic Head Phantom for Calibration of TLD in Eye Lens Dosimeter [J].
Endarko E. ;
Pitaloka A.W. ;
Inayah A.A. ;
Ramadhan F.R. ;
Kardianto K. ;
Rahayuningsih B. .
Journal of Biomedical Physics and Engineering, 2023, 13 (02) :193-202
[8]   Different meaning of the mean heart dose between 3D-CRT and IMRT for breast cancer radiotherapy [J].
Prunaretty, Jessica ;
Bourgier, Celine ;
Gourgou, Sophie ;
Lemanski, Claire ;
Azria, David ;
Fenoglietto, Pascal .
FRONTIERS IN ONCOLOGY, 2023, 12
[9]   Treatment of left sided breast cancer for a patient with funnel chest: Volumetric-modulated arc therapy vs. 3D-CRT and intensity-modulated radiotherapy [J].
Haertl, Petra M. ;
Pohl, Fabian ;
Weidner, Karin ;
Groeger, Christian ;
Koelbl, Oliver ;
Dobler, Barbara .
MEDICAL DOSIMETRY, 2013, 38 (01) :1-4
[10]   Dosimetric evaluation of a patient-specific 3D-printed oral positioning stent for head-and-neck radiotherapy [J].
Susannah Cleland ;
Philip Chan ;
Benjamin Chua ;
Scott B. Crowe ;
Jodi Dawes ;
Lizbeth Kenny ;
Charles Lin ;
Elise Obereigner ;
Samuel C. Peet ;
Jamie V. Trapp ;
Tania Poroa ;
Tanya Kairn .
Physical and Engineering Sciences in Medicine, 2021, 44 :887-899