Evaluation of 3D-CRT Radiotherapy Treatment Planning with Radiobiological Effects in Patient-Specific 3D Printed Anthropomorphic Phantom Postmastectomy Breast Cancer Axillary Metastases

被引:0
|
作者
Endarko, Endarko [1 ]
Hariyanto, Aditya Prayugo [1 ]
Aisyah, Siti [1 ]
Kavilani, Nandia [2 ]
Syafi’i, Ahmad [2 ]
机构
[1] Department of Physics, Institut Teknologi Sepuluh Nopember, Kampus ITS, Sukolilo Surabaya, East Java
[2] Medical Physicist of Radiotherapy Installation, Naval Hospital Ramelan, Surabaya, East Java
关键词
Breast Cancer; Conformal Radiation; Dosimetry Radiation; Phantom; Protection; Radiation Therapy; Therapy (3D-CRT); Three-Dimensional;
D O I
10.22038/ijmp.2024.73930.2309
中图分类号
学科分类号
摘要
Introduction: The aim of this study was to evaluate and measure the effectiveness of a particular radiotherapy treatment planning system, namely 3D conformal radiation therapy (3D-CRT), and radiobiological effects in postmastectomy breast cancer axillary lymph metastases using a patient-specific 3D printed anthropomorphic phantom of the right breast after surgery. Material and Methods: The chest wall and axilla were the two treatment targets for 3D-CRT planning techniques. Each target was subjected to variations of two and three irradiation fields, with a dose of 50.00 Gy and fractionation of 2.00 Gy. From the variations, the radiation dose received by planning target volume (PTV) chest wall, PTV-Axillary, and organs at risk (OARs) was evaluated to determine the best planning technique through dosimetry verification. The thermoluminescent dosimeter (TLD) 100 Chip and EBT3 film were employed to measure the radiation dose at the PTV and OARs. Results: The planning system evaluation based on dose volume histogram showed no significant difference (p = 0.993) between two and three irradiation fields in both the chest wall and axillary planning 3D-CRT technique. Normal tissue complication probability for OARs has values below 5%, while the smallest value was obtained for the left lung. Finally, the point dose verification between planning and measurement using TLD100 and EBT3 film indicated an average difference of 9.14% and 4.34%, respectively, with no significant difference. Conclusion: This study successfully demonstrated the dosimetry evaluation and radiobiological effects of postmastectomy right breast cancer treatment using a patient-specific 3D printed anthropomorphic phantom, which met the established standards. © (2024), (Mashhad University of Medical Sciences). All rights reserved.
引用
收藏
页码:314 / 322
页数:8
相关论文
共 50 条
  • [1] Dosimetry Evaluation of Treatment Planning Systems in Patient-Specific 3D Printed Anthropomorphic Phantom for Breast Cancer after Mastectomy using a Single-Beam 3D-CRT Technique for Megavoltage Electron Radiation Therapy
    Endarko E.
    Aisyah S.
    Hariyanto A.P.
    Haekal M.
    Kavilani N.
    Syafi’i A.
    Journal of Biomedical Physics and Engineering, 2023, 13 (03) : 217 - 226
  • [2] Patient-Specific Quality Assurance Using a 3D-Printed Chest Phantom for Intraoperative Radiotherapy in Breast Cancer
    Choi, Yeonho
    Lee, Ik Jae
    Park, Kwangwoo
    Park, Kyung Ran
    Cho, Yeona
    Kim, Jun Won
    Lee, Ho
    FRONTIERS IN ONCOLOGY, 2021, 11
  • [3] Dosimetric Comparison of IMRT with 3D-CRT Regarding Their Contribution to the Treatment Plan Optimization Using Rando Phantom with a Realistic Lung Cancer Radiotherapy Treatment Planning
    Benkahila K.
    Kharfi F.
    Boulakhssaim F.
    Khoudri S.
    Iranian Journal of Medical Physics, 2021, 18 (03) : 154 - 163
  • [4] Dosimetric verification of cancer patient?s treatment plan using an anthropomorphic, 3D-printed phantom
    Waluk, Kacper
    Pietrzak, Jakub
    APPLIED RADIATION AND ISOTOPES, 2023, 191
  • [5] Evaluation of a 3D-printed heterogeneous anthropomorphic head and neck phantom for patient-specific quality assurance in intensity-modulated radiation therapy
    Noriyuki Kadoya
    Kota Abe
    Hikaru Nemoto
    Kiyokazu Sato
    Yoshiro Ieko
    Kengo Ito
    Suguru Dobashi
    Ken Takeda
    Keiichi Jingu
    Radiological Physics and Technology, 2019, 12 : 351 - 356
  • [6] Evaluation of a 3D-printed heterogeneous anthropomorphic head and neck phantom for patient-specific quality assurance in intensity-modulated radiation therapy
    Kadoya, Noriyuki
    Abe, Kota
    Nemoto, Hikaru
    Sato, Kiyokazu
    Ieko, Yoshiro
    Ito, Kengo
    Dobashi, Suguru
    Takeda, Ken
    Jingu, Keiichi
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2019, 12 (03) : 351 - 356
  • [7] Evaluation of 3D Printed Anthropomorphic Head Phantom for Calibration of TLD in Eye Lens Dosimeter
    Endarko E.
    Pitaloka A.W.
    Inayah A.A.
    Ramadhan F.R.
    Kardianto K.
    Rahayuningsih B.
    Journal of Biomedical Physics and Engineering, 2023, 13 (02) : 193 - 202
  • [8] Different meaning of the mean heart dose between 3D-CRT and IMRT for breast cancer radiotherapy
    Prunaretty, Jessica
    Bourgier, Celine
    Gourgou, Sophie
    Lemanski, Claire
    Azria, David
    Fenoglietto, Pascal
    FRONTIERS IN ONCOLOGY, 2023, 12
  • [9] Dosimetric evaluation of a patient-specific 3D-printed oral positioning stent for head-and-neck radiotherapy
    Susannah Cleland
    Philip Chan
    Benjamin Chua
    Scott B. Crowe
    Jodi Dawes
    Lizbeth Kenny
    Charles Lin
    Elise Obereigner
    Samuel C. Peet
    Jamie V. Trapp
    Tania Poroa
    Tanya Kairn
    Physical and Engineering Sciences in Medicine, 2021, 44 : 887 - 899
  • [10] Treatment of left sided breast cancer for a patient with funnel chest: Volumetric-modulated arc therapy vs. 3D-CRT and intensity-modulated radiotherapy
    Haertl, Petra M.
    Pohl, Fabian
    Weidner, Karin
    Groeger, Christian
    Koelbl, Oliver
    Dobler, Barbara
    MEDICAL DOSIMETRY, 2013, 38 (01) : 1 - 4