Modeling and application research of absolute gravimeter driven by a V-shaped linear ultrasonic motor

被引:0
作者
Zhou, Lifeng [1 ,2 ]
Xu, Hao [3 ]
Geng, Ranran [1 ,2 ]
Yao, Zhiyuan [3 ]
机构
[1] Nanjing Inst Technol, Ind Ctr, Nanjing 211167, Peoples R China
[2] Jiangsu Prov Engn Res Ctr Bion Control Technol & E, Nanjing 211167, Peoples R China
[3] Nanjing Univ Aeronaut & Astronaut, State Key Lab Mech & Control Mech Struct, Nanjing 210016, Peoples R China
基金
美国国家科学基金会;
关键词
TEMPERATURE RISE; ACCELERATION; TRANSDUCER; JOINT;
D O I
10.1063/5.0238063
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A V-shaped linear ultrasonic motor (LUSM), used as a driving element for the absolute gravimeter, generates structural vibrations when the motor starts and stops, which interfere with and influence the accurate measurement of the absolute gravimeter. In order to address this problem and provide better stability and measurement accuracy, this paper designs a fuzzy proportional integral differential (PID) controller and realizes the drive control of the motor on a field programmable gate array (FPGA) using the Verilog hardware description language (HDL). The advantages of the proposed controller are verified by comparing the simulation results with those of the traditional PID controller. First, the technical index of the LUSM as the driving element is studied through theoretical analysis based on the operation mechanism of the absolute gravimeter. Then, the relationship between the output performance and the input parameters (excitation voltage, preload, and driving frequency) of the motor in a vacuum/natural environment is experimentally investigated. Finally, an experimental system is constructed, and experiments are conducted to measure the acceleration of gravity and compare with the test results of the habitual FG-5 absolute gravimeter. The experimental results show that the output performance of the V-shaped LUSM meets the requirements of the driving element of the absolute gravimeter. The combination of the two control methods can realize the repeatable and stable operation of the system. Moreover, the motor with a mass of only 92 g can output force directly without transmission components. The complexity and the weight of the system are significantly reduced, which has certain advantages.
引用
收藏
页数:12
相关论文
共 28 条
  • [1] Arnautov G. P., 1979, Soviet Journal of Quantum Electronics, V9, P333, DOI 10.1070/QE1979v009n03ABEH008778
  • [2] AN ELASTIC-PLASTIC MODEL FOR THE CONTACT OF ROUGH SURFACES
    CHANG, WR
    ETSION, I
    BOGY, DB
    [J]. JOURNAL OF TRIBOLOGY-TRANSACTIONS OF THE ASME, 1987, 109 (02): : 257 - 263
  • [3] ABSOLUTE DETERMINATION OF ACCELERATION DUE TO GRAVITY
    COOK, AH
    [J]. CONTEMPORARY PHYSICS, 1967, 8 (03) : 251 - &
  • [5] Faller J. E., 1983, Bur. Grav. Int. Bull. dInf, V53, P87
  • [6] RESULTS OF AN ABSOLUTE DETERMINATION OF ACCELERATION OF GRAVITY
    FALLER, JE
    [J]. JOURNAL OF GEOPHYSICAL RESEARCH, 1965, 70 (16): : 4035 - +
  • [7] Linear ultrasonic motor for absolute gravimeter
    Jian, Yue
    Yao, Zhiyuan
    Silberschmidt, Vadim V.
    [J]. ULTRASONICS, 2017, 77 : 88 - 94
  • [8] Transducer for high speed and large thrust ultrasonic linear motor using two sandwich-type vibrators
    Kurosawa, MK
    Kodaira, O
    Tsuchitoi, Y
    Higuchi, T
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1998, 45 (05) : 1188 - 1195
  • [9] Temperature evaluation of traveling-wave ultrasonic motor considering interaction between temperature rise and motor parameters
    Li, Shiyang
    Ou, Wenchu
    Yang, Ming
    Guo, Chao
    Lu, Cunyue
    Hu, Junhui
    [J]. ULTRASONICS, 2015, 57 : 159 - 166
  • [10] Analytical modeling and experimental validation of a V-shaped piezoelectric transducer with a flexible joint for generating an elliptical motion
    Li, Xiaoniu
    Yao, Zhiyuan
    [J]. PRECISION ENGINEERING-JOURNAL OF THE INTERNATIONAL SOCIETIES FOR PRECISION ENGINEERING AND NANOTECHNOLOGY, 2018, 52 : 55 - 63