An instrument is described for measuring the energy of a laser pulse in a prescribed time interval. The device operates by integrating the signal from a photosensor, the integrating circuit being switched off at a prescribed moment of time. The time interval over which energy is measured is DELTA t equals 10 mu sec to 1 msec. Measurement accuracy is approximately 2%.