ENERGY SPREAD OF PARTICLES IN BEAM OF ELECTROSTATIC ACCELERATOR.

被引:0
|
作者
Dzagurov, O.B. [1 ]
Eremin, N.V. [1 ]
Krivolap, V.V. [1 ]
Sirotinin, E.I. [1 ]
Strizhov, V.F. [1 ]
机构
[1] Moscow State Univ, USSR, Moscow State Univ, USSR
来源
Instruments and experimental techniques New York | 1987年 / 30卷 / 2 pt 1期
关键词
ELECTROSTATIC DEVICES - PROTONS;
D O I
暂无
中图分类号
学科分类号
摘要
The resonance reactions **2**7 Al(p, gamma ) for E//p equals 922 KeV and **2**7 Al(p, p//2 gamma ) and **2**7 Al(p, a//2 gamma ) for E//p equals 1679. 5 keV are used to determine the energy spread in the proton beam of an electrostatic accelerator. The results show that use of an inductive charging system improves energy homogeneity to 1 multiplied by (times) 10**4.