DESIGN OF A BISMUTH GERMANATE ANTI-COMPTON SPECTROMETER AND ITS USE IN NUCLEAR SPECTROSCOPY.

被引:0
|
作者
Lieder, R.M.
Jaeger, H.
Neskakis, A.
Venkova, T.
Michel, C.
机构
关键词
GAMMA RAYS - Spectrum Analysis;
D O I
暂无
中图分类号
学科分类号
摘要
The design of an anti-Compton spectrometer consisting of a Ge detector and an asymmetric anti-Compton shield made of bismuth germanate is described. Its dimensions have been optimized by means of Monte Carlo calculations. The features predicted for this asymmetric anti-Compton shield are compared with those of symmetric ones. The design of a gamma - gamma coincidence spectrometer consisting of twelve anti-Compton spectrometers and a sum-energy and gamma -ray multiplicity filter is discussed. Results of neutron-sensitivity measurements for bismuth germanate and sodium iodide detectors are presented.
引用
收藏
页码:363 / 370
相关论文
共 50 条
  • [11] MEASUREMENT OF LASER-INDUCED COMPTON BACKSCATTERED PHOTONS WITH ANTI-COMPTON SPECTROMETER
    OHGAKI, H
    SUGIYAMA, S
    YAMAZAKI, T
    MIKADO, T
    CHIWAKI, M
    YAMADA, K
    SUZUKI, R
    NOGUCHI, T
    TOMIMASU, T
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1991, 38 (02) : 386 - 392
  • [12] SEGMENTED COMPTON SUPPRESSION SPECTROMETER USING BISMUTH-GERMANATE
    OLIN, A
    POFFENBERGER, PR
    BRITTON, DI
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1984, 222 (03): : 463 - 466
  • [13] PROPOSAL FOR A VERSATILE COINCIDENCE SPECTROMETER EMPLOYING AN ANTI-COMPTON ANNULUS
    KANTELE, J
    MARTTILA, OJ
    NUCLEAR INSTRUMENTS & METHODS, 1964, 27 (02): : 235 - 236
  • [14] A HIGH-RESOLUTION GE(LI) ANTI-COMPTON SPECTROMETER FOR RADIATIVE NEUTRON CAPTURE SPECTROSCOPY
    MICHAELIS, W
    KUPFER, H
    NUCLEAR INSTRUMENTS & METHODS, 1967, 56 (02): : 181 - +
  • [15] VERSATILE ALL-GE(LI) ANTI-COMPTON SPECTROMETER
    ALBERTI, G
    ISABELLA, L
    ORIANI, A
    ZAMBRA, A
    NUCLEAR INSTRUMENTS & METHODS, 1971, 95 (03): : 413 - +
  • [16] A PAIR AND ANTI-COMPTON SPECTROMETER FOR STUDY OF NEUTRON CAPTURE GAMMA RAYS
    MICHAUD, B
    HUBER, O
    KERN, J
    RIBORDY, L
    SCHALLER, LA
    HELVETICA PHYSICA ACTA, 1969, 42 (7-8): : 917 - &
  • [17] A CRYOGENICALLY COOLED SINGLE-CRYSTAL BGO ANTI-COMPTON SPECTROMETER
    SCHOTANUS, P
    MICEK, SL
    DEVOIGT, MJA
    WINTRAECKEN, Y
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (04) : 730 - 734
  • [18] Design of an anti-Compton spectrometer for low-level radioactive wastes using Monte Carlo techniques
    Tsutsumi, M
    Oishi, T
    Kinouchi, N
    Sakamoto, R
    Yoshida, M
    JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY, 2002, 39 (09) : 957 - 963
  • [19] INCREASING OF THE GAMMA-RAY DETECTION SENSITIVITY USING AN ANTI-COMPTON SPECTROMETER
    ADAM, J
    KRACIK, B
    HOFFMANN, J
    JADERNA ENERGIE, 1984, 30 (02): : 41 - 44
  • [20] A GE(LI) GAMMA SPECTROMETER EMPLOYING AN ANTI-COMPTON MANTLE OF NAI(TL)
    KANTELE, J
    SUOMINEN, P
    NUCLEAR INSTRUMENTS & METHODS, 1966, 41 (01): : 41 - &