EFFECTS OF SECONDARY ELECTRON DETECTOR POSITION ON SCANNING ELECTRON MICROSCOPE IMAGE.

被引:0
|
作者
Kawamoto, H.
Yamazaki, S.
Ishikawa, A.
Buchanan, R.
机构
关键词
ELECTRONS; -; Emission;
D O I
暂无
中图分类号
学科分类号
摘要
The image contrast in a SEM employing an open bore geometry objective lens is seen to vary substantially, depending on whether the images are produced using an SE detector above or below the lens, and on working distance. At a short working distance, and with the specimen in the gap between the objective lens polepiece, there is a very high contribution of Type 1 SEs to the top detector signal. This is advantageous for the production of high resolution images. In addition, the images are very sensitive to surface effects such as the presence of contaminants.
引用
收藏
页码:15 / 22
相关论文
共 50 条
  • [31] Scanning detector of secondary electron spin polarization
    Sokolov, OB
    Kuznetsov, VL
    Gribov, IV
    Nosov, AP
    Kutsin, EV
    PHYSICS OF METALS AND METALLOGRAPHY, 2001, 92 : S271 - S276
  • [32] Adaptive Threshold Processing of Secondary Electron Images in Scanning Electron Microscope
    Bian, Weiguo
    Wang, Mingyu
    Yang, Zhan
    INTELLIGENT ROBOTICS AND APPLICATIONS, ICIRA 2019, PT I, 2019, 11740 : 166 - 173
  • [33] CHARGING EFFECTS IN SCANNING ELECTRON MICROSCOPE
    SHAFFNER, TJ
    VANVELD, RD
    JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1971, 4 (SEP): : 633 - &
  • [34] OBSERVATION OF BIOLOGICAL SPECIMENS USING THE SCANNING-TRANSMISSION ELECTRON-MICROSCOPE EQUIPPED WITH SECONDARY-ELECTRON DETECTOR
    TAMURA, E
    KATOH, M
    YAMAMOTO, I
    JOURNAL OF ELECTRON MICROSCOPY, 1982, 31 (03): : 333 - 333
  • [35] Imaging thin and thick sections of biological tissue with the secondary electron detector in a field-emission scanning electron microscope
    Wergin, WP
    Yaklich, RW
    Roy, S
    Joy, DC
    Erbe, EF
    Murphy, CA
    Pooley, CD
    SCANNING, 1997, 19 (06) : 387 - 395
  • [36] Use of backscattered electron detector arrays for forming backscattered electron images in the scanning electron microscope
    Wells, OC
    Gignac, LM
    Murray, CE
    Frye, A
    Bruley, J
    SCANNING, 2006, 28 (01) : 27 - 31
  • [37] SECONDARY ELECTRON DETECTION IN A FIELD EMISSION SCANNING MICROSCOPE
    CREWE, AV
    ISAACSON, M
    JOHNSON, D
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1970, 41 (01): : 20 - &
  • [38] Accurate virus quantitation using a Scanning Transmission Electron Microscopy (STEM) detector in a scanning electron microscope
    Blancett, Candace D.
    Fetterer, David P.
    Koistinen, Keith A.
    Morazzani, Elaine M.
    Monninger, Mitchell K.
    Piper, Ashley E.
    Kuehl, Kathleen A.
    Kearney, Brian J.
    Norris, Sarah L.
    Rossi, Cynthia A.
    Glass, Pamela J.
    Sun, Mei G.
    JOURNAL OF VIROLOGICAL METHODS, 2017, 248 : 136 - 144
  • [39] SECONDARY-ELECTRON DETECTION IN THE SCANNING-TRANSMISSION ELECTRON-MICROSCOPE
    IMESON, D
    MILNE, RH
    BERGER, SD
    MCMULLAN, D
    ULTRAMICROSCOPY, 1985, 17 (03) : 243 - 249
  • [40] Quantitative secondary electron energy filtering in a scanning electron microscope and its applications
    Kazemian, P.
    Mentink, S. A. M.
    Rodenburg, C.
    Humphreys, C. J.
    ULTRAMICROSCOPY, 2007, 107 (2-3) : 140 - 150