An electrical impedance tomography microscope

被引:0
|
作者
Griffiths, H. [1 ]
Tucker, M.G. [2 ]
Sage, J. [2 ]
Herrenden-Harker, W.G. [2 ]
机构
[1] Department of Medical Physics and Bioengineering, University Hospital of Wales, Heath Park, Cardiff CF4 4XW, United Kingdom
[2] Department of Physics and Astronomy, University of Wales, College of Cardiff, The Parade, Cardiff CF2 3YB, United Kingdom
关键词
Electric impedance - Electric impedance measurement - Printed circuit boards - Electric impedance tomography;
D O I
暂无
中图分类号
学科分类号
摘要
A circular array of 16 electrodes has been constructed for use as an electrical impedance tomography (EIT) microscope. The electrodes were made from 60 μm diameter gold wires anchored to a printed circuit board. The internal diameter of the array was 0.9 mm giving a theoretical spatial resolution of about 100 μm. For EIT imaging, the array was connected to an imaging system operating at 82 kHz. Static images of conducting and insulating filaments (copper wire and human hair) in saline solution were obtained as well as dynamic imaging sequences of glass microspheres migrating through the array. The interelectrode impedance was typically 5 kΩ and the transimpedances ranged from 14 to 210 Ω. © 1996 IOP Publishing Ltd.
引用
收藏
相关论文
共 50 条
  • [1] An electrical impedance tomography microscope
    Griffiths, H
    Tucker, MG
    Sage, J
    HerrendenHarker, WG
    PHYSIOLOGICAL MEASUREMENT, 1996, 17 : A15 - A24
  • [2] Electrical impedance tomography
    Lobo, Beatriz
    Hermosa, Cecilia
    Abella, Ana
    Gordo, Federico
    ANNALS OF TRANSLATIONAL MEDICINE, 2018, 6 (02)
  • [3] Electrical impedance tomography
    Saulnier, GJ
    Blue, RS
    Newell, JC
    Isaacson, D
    Edic, PM
    IEEE SIGNAL PROCESSING MAGAZINE, 2001, 18 (06) : 31 - 43
  • [4] Electrical impedance tomography
    Cheney, M
    Isaacson, D
    Newell, JC
    SIAM REVIEW, 1999, 41 (01) : 85 - 101
  • [5] Electrical impedance tomography
    Kotre, CJ
    BRITISH JOURNAL OF RADIOLOGY, 1997, 70 : S200 - S205
  • [6] Electrical Impedance Tomography
    Hinz, Jose
    CURRENT RESPIRATORY MEDICINE REVIEWS, 2009, 5 (02) : 105 - 109
  • [7] Electrical impedance tomography
    Borcea, L
    INVERSE PROBLEMS, 2002, 18 (06) : R99 - R136
  • [8] Electrical impedance tomography
    Costa, Eduardo L. V.
    Lima, Raul Gonzalez
    Amato, Marcelo B. P.
    CURRENT OPINION IN CRITICAL CARE, 2009, 15 (01) : 18 - 24
  • [9] Bioimpedance tomography (Electrical impedance tomography)
    Bayford, R. H.
    ANNUAL REVIEW OF BIOMEDICAL ENGINEERING, 2006, 8 : 63 - 91
  • [10] Multi-frequency Electrical Impedance Tomography and Magnetic Resonance Electrical Impedance Tomography
    Seo, Jin Keun
    Woo, Eung Je
    MATHEMATICAL MODELING IN BIOMEDICAL IMAGING I: ELECTRICAL AND ULTRASOUND TOMOGRAPHIES, ANOMALY DETECTION, AND BRAIN IMAGING, 2009, 1983 : 1 - 71