The basic characteristics of a linear pulse motor-driven artificial heart, as obtained in mock tests, are discussed. The artificial heart consists mainly of a linear pulse motor and a blood pump which are mechanically coupled. The linear pulse motor has a flat structure and is constructed from laminated sheets of permendure and rare-earth magnets. Two kinds of blood pumps, the diaphragm and the pusher-cup types, were prepared for use in the mock tests. With the diaphragm-type pump, an aortic pressure of 69 to 103 mmHg was obtained at 113 beats/min. Using the pusher-cup type pump, an aortic pressure of 53 to 97 mmHg was obtained at 130 beats/min. The artificial heart was thus confirmed to have the potential for actual implantation.
引用
收藏
页码:796 / 802
相关论文
共 50 条
  • [1] DEVELOPMENT AND TEST OF A LINEAR MOTOR-DRIVEN TOTAL ARTIFICIAL-HEART
    YAMADA, H
    YAMAGUCHI, M
    KOBAYASHI, K
    MATSUURA, Y
    TAKANO, H
    IEEE ENGINEERING IN MEDICINE AND BIOLOGY MAGAZINE, 1995, 14 (01): : 84 - 90
  • [2] In vitro evaluation of linear motor-driven total artificial heart
    Kobayashi, M
    Wakiwaka, H
    Watanabe, M
    Mizuno, H
    Karita, M
    Maeda, M
    Matsuura, Y
    Fukunaga, S
    Yamada, H
    ARTIFICIAL ORGANS, 1996, 20 (12) : 1320 - 1324
  • [3] LOOKING AHEAD - ELECTRICAL MOTOR-DRIVEN ARTIFICIAL-HEART
    COLE, H
    JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1985, 253 (18): : 2618 - 2619
  • [4] Performance comparison of two linear motor-driven totally artificial hearts
    Yamaguchi, M.
    Kobayashi, K.
    Yamamoto, Y.
    Yamada, H.
    IEEE translation journal on magnetics in Japan, 1994, 9 (02): : 166 - 173
  • [5] A NEW TYPE OF THE MOTOR-DRIVEN BLOOD PUMP FOR ARTIFICIAL-HEART
    CHEON, GJ
    KIM, HC
    MIN, BG
    HAN, DC
    JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1990, 112 (04): : 473 - 475
  • [6] A test of centrifugal motor-driven pumps
    Rumsey, SS
    Schwedes, WF
    TRANSACTIONS OF THE AMERICAN INSTITUTE OF MINING AND METALLURGICAL ENGINEERS, 1914, 50 : 779 - 801
  • [7] DEVELOPMENT OF A LINEAR MOTOR-DRIVEN METRO SYSTEM
    MATSUMOTO, A
    MIZUMA, T
    KIMURA, H
    AUTOMATED PEOPLE MOVERS II: NEW LINKS FOR LAND USE - AUTOMATED PEOPLE MOVER OPPORTUNITIES FOR MAJOR ACTIVITY CENTERS, 1989, : 666 - 675
  • [8] A LINEAR ELECTRIC MOTOR-DRIVEN FORGING HAMMER
    MOLDOVAN, V
    METALLURGIA, 1980, 47 (09): : 449 - 450
  • [9] AN ELECTRIC MOTOR-DRIVEN TOTAL ARTIFICIAL-HEART - 7 MONTHS SURVIVAL IN THE CALF
    ROSENBERG, G
    SNYDER, AJ
    LANDIS, DL
    GESELOWITZ, DB
    DONACHY, JH
    PIERCE, WS
    TRANSACTIONS AMERICAN SOCIETY FOR ARTIFICIAL INTERNAL ORGANS, 1984, 30 : 69 - 74
  • [10] MOTOR-DRIVEN LEFT-RIGHT SIMULTANEOUSLY EJECTING TOTAL ARTIFICIAL-HEART (TAH)
    TAKATANI, S
    TAKANO, H
    TAENAKA, Y
    UMEZU, M
    NODA, H
    KINOSHITA, M
    FUKUDA, S
    AKUTSU, T
    ARTIFICIAL ORGANS, 1987, 11 (04) : 297 - 297