INTRINSIC RESOLVING TIME OF DIAMOND NUCLEAR-RADIATION DETECTORS.

被引:0
|
作者
Kozlov, S.F.
Likhosherstov, V.N.
机构
来源
Instruments and Experimental Techniques (English Translation of Pribory I Tekhnika Eksperimenta) | 1973年 / 16卷 / 5 Part 1期
关键词
DIAMONDS - Industrial Applications;
D O I
暂无
中图分类号
学科分类号
摘要
A description is given of an experimental installation for measuring the intrinsic resolving time of diamond detectors as a function of their thickness, energy resolution, bias voltage, and the type of ionizing radiation, and the results of the measurements are presented. It is concluded that the intrinsic resolving time of diamond detectors does not depend on the energy resolution, on the bias voltage in the working range 150 to 250 V, or on the thickness in the 110 to 340 mu range. For recording alpha particles being absorbed in the surface layer of a diamond detector ( approximately equals 20 mu ) and gamma quanta for which the detectors are ″thin″ the resolving time within the limits of the errors has an identical magnitude which is approximately equal to the intrinsic resolving time of the measurement path.
引用
收藏
页码:1363 / 1364
相关论文
共 50 条
  • [31] STABILITY OF SELECTIVE RADIOMETERS WITH PHOTOELECTRIC RADIATION DETECTORS.
    Gracheva, I.V.
    Koltun, M.M.
    Orshanskii, I.S.
    Measurement Techniques, 1985, 28 (01) : 45 - 48
  • [32] DIAMOND RADIATION DETECTORS
    KEDDY, RJ
    NAM, TL
    RADIATION PHYSICS AND CHEMISTRY, 1993, 41 (4-5): : 767 - 773
  • [33] DIAMOND RADIATION DETECTORS
    KANIA, DR
    LANDSTRASS, MI
    PLANO, MA
    PAN, LS
    HAN, S
    DIAMOND AND RELATED MATERIALS, 1993, 2 (5-7) : 1012 - 1019
  • [34] Diamond radiation detectors .2. CVD diamond development for radiation detectors
    Kania, DR
    PHYSICS OF DIAMOND, 1997, 135 : 565 - 566
  • [35] SPATIAL RESOLUTION OF OPTOACOUSTIC LASER-RADIATION DETECTORS.
    Bagratashvili, V.N.
    Zharov, V.P.
    Lobko, V.V.
    1978, 8 (03): : 366 - 369
  • [36] PROPERTIES AND APPLICATIONS OF SOLID STATE NUCLEAR TRACK DETECTORS.
    Almasi, Gyula
    Boeck, Helmuth
    Suleiman, Mohammed
    Vana, Norbert
    Atomkernenergie, Kerntechnik, 1984, 44 (04): : 304 - 311
  • [37] SEARCH FOR ANOMALONS USING PLASTIC NUCLEAR TRACK DETECTORS.
    Drechsel, H.
    Trakowski, W.
    Dreute, J.
    Sonntag, S.
    Brechtmann, C.
    Rudat, R.
    Beer, J.
    Heinrich, W.
    Benton, E.V.
    Cassou, R.M.
    Nuclear tracks, 1983, 8 (1-4): : 563 - 566
  • [38] The optimization of the properties of the beta-radiation scintillation detectors.
    Sysoeva, EP
    Tarasov, VA
    Meshman, AP
    Gen, NS
    1995 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD, VOLS 1-3, 1996, : 790 - 791
  • [39] ADVANCES IN MERCURIC IODIDE SINGLE CRYSTAL NUCLEAR DETECTORS.
    Randtke, P.T.
    Ortale, C.
    Whited, R.C.
    van den Berg, L.
    1976, : 71 - 74
  • [40] SEARCH FOR ANOMALONS USING PLASTIC NUCLEAR TRACK DETECTORS.
    Heinrich, W.
    Brechtmann, C.
    Dreute, J.
    Dreschel, H.
    1600, (12): : 1 - 6