New generation biodegradable plate for fracture fixation: Comparison of bending strengths of mandibular osteotomies fixed with absorbable self-reinforced multi-layer poly-l-lactide plates and metallic plates - An experimental study in sheep

被引:0
|
作者
Suuronen, R. [1 ]
Pohjonen, T. [1 ]
Wessman, L. [1 ]
Tormala, P. [1 ]
Vainionpaa, S. [1 ]
机构
[1] Department of Oral and Maxillofacial Surgery, University of Helsinki, Mannerheimintie 127, 00300 Helsinki, Finland
来源
Clinical Materials | 1992年 / 9卷 / 02期
关键词
Calorimetry; -; Plates; Polymers; Radiography;
D O I
暂无
中图分类号
学科分类号
摘要
Eighteen osteotomies of sheep's mandibular corpus were fixed: nine with self-reinforced multi-layer poly-l-lactide (SR-PLLA) plates and nine with metallic plates. Both plates were fixed with similar metallic screws. The right corpus of each sheep acted as a control. Radiographic analysis was carried out at 3 weeks and at sacrifice (6, 12 or 24 weeks) when the bending force needed to break the osteotomy site was measured. At 6 and 12 weeks of follow-up the osteotomy fixed with the PLA-plate seemed to tolerate the bending force better but no significant differences were detected at 3 or 24 weeks. The results have been satisfactory and have led the authors to continue with the development of a SR-PPLA plate-and-screw device for the fracture fixation.