EVERTING (TOPOSCOPIC) CATHETER FOR BROAD CLINICAL APPLICATION.

被引:0
作者
Shook, D.R. [1 ]
Doppman, J.L. [1 ]
Cattau Jr., E.L. [1 ]
Goldstein, S.R. [1 ]
机构
[1] Natl Inst of Health, Bethesda, MD,, USA, Natl Inst of Health, Bethesda, MD, USA
关键词
POLYURETHANES - Elasticity - THERMOPLASTIC ELASTOMERS - Biocompatibility;
D O I
暂无
中图分类号
学科分类号
摘要
The advanced development of the clinical everting (toposcopic) catheter is described. A detailed discussion of the design and outline of the fabrication techniques are followed by a thorough performance evaluation and summary of the first two clinical applications. The everting element is a low-durometer thermoplastic polyurethane elastomer. Surface treatments include the bonding of a hydrophilic polymeric coating, optimized for lubricity, to the sliding internal surfaces of the catheter. Eversion pressures and infusion/aspiration flow rates have been measured under various conditions and the infusate-in-blood mixing potential investigated. A preliminary assessment is given of the clinical performance of the catheter in the vascular delivery of chemotherapy and standard endoscopic retrograde cholangiopancreatography.