Electrochemical detection of glucose in sweat using a paper-based origami biosensor based on PEDOT:PSS/Ti3C2/MWCNT composite

被引:3
作者
Moradi, Sara [1 ]
Hosseini, Morteza [1 ,2 ]
Firoozbakhtian, Ali [1 ]
Tashnizi, Mohammad Barshan [3 ]
Ganjali, Mohammad Reza [4 ]
机构
[1] Univ Tehran, Fac New Sci & Technol, Dept Life Sci Engn, Nanobiosensors Lab, Tehran 1439817435, Iran
[2] Univ Tehran Med Sci, Med Biomat Res Ctr, Tehran, Iran
[3] Univ Tehran, Fac New Sci & Technol, Dept Life Sci Engn, Tehran, Iran
[4] Univ Tehran, Coll Sci, Ctr Excellence Electrochem, Sch Chem, Tehran 1439817435, Iran
关键词
Origami; Paper-based-Biosensor; Glucose; Electrochemistry; Sweat; CARBON NANOTUBES; IMMOBILIZATION; ELECTRODE; DEVICE; ENZYME; SENSOR;
D O I
10.1016/j.jelechem.2024.118675
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Origami glucose biosensors utilize folding techniques to create compact, cost-effective devices for monitoring glucose levels and offer great potential in healthcare and diagnostics applications. The origami design of the biosensor allows for easy handling and transport, making it ideal for point-of-care testing in resource-limited areas, especially for individuals with diabetes to monitor glucose levels. This study introduced a novel paper-based origami glucose biosensor with graphite pencil electrodes which directly drawn onto the paper surface. A simple method used to draw two circular areas for effective immobilization of the mediator and glucose oxidase enzyme. A novel mediator composed of MWCNT and Ti3C2 dissolved in PEDOT:PSS ink was introduced to the mediator zone to enhance the paper surface's conductivity, thereby boosting the electron transfer rate from the enzyme region to the electrodes. The mediator ink was produced by using a straightforward one-step process, eliminating the necessity for specialized equipment and intricate synthesis techniques. Also, the mediator composite lowered the applied potential for glucose oxidation reaction according to the high surface area, unique electronic properties, and excellent catalytic activity of the composite. The lower potentials for glucose oxidation are required to avoid the electrochemical oxidation of interfering substances, to reach a high accuracy in the detection of the analyte. With a detection limit of 0.007 mM and linear range of 0.01 mM to 0.4 mM, it covers a wide range of glucose concentrations found in sweat.
引用
收藏
页数:8
相关论文
共 46 条
  • [1] Non-enzymatic glucose sensor based on copper oxide and multi-wall carbon nanotubes using PEDOT:PSS matrix
    Amirzadeh, Zhila
    Javadpour, Sirus
    Shariat, Mohammad Hossein
    Knibbe, Ruth
    [J]. SYNTHETIC METALS, 2018, 245 : 160 - 166
  • [2] Sensor and biosensor application of a new redox mediator: Rosmarinic acid modified screen-printed carbon electrode for electrochemical determination of NADH and ethanol
    Bilgi, Melike
    Sahin, Elif Merve
    Ayranci, Erol
    [J]. JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2018, 813 : 67 - 74
  • [3] Recent advances in electrochemical glucose biosensors: a review
    Chen, Chao
    Xie, Qingji
    Yang, Dawei
    Xiao, Hualing
    Fu, Yingchun
    Tan, Yueming
    Yao, Shouzhuo
    [J]. RSC ADVANCES, 2013, 3 (14): : 4473 - 4491
  • [4] An origami paper-based electrochemical biosensing platform for quality control of agri-food waste in the valorization strategy
    Colozza, Noemi
    Di Meo, Erika
    Mucaria, Angelica
    Moscone, Danila
    Arduini, Fabiana
    [J]. MICROCHIMICA ACTA, 2022, 189 (08)
  • [5] Ti3C2/Ni/Sm-based electrochemical glucose sensor for sweat analysis using bipolar electrochemistry
    Damirchi, Zahra
    Firoozbakhtian, Ali
    Hosseini, Morteza
    Ganjali, Mohammad Reza
    [J]. MICROCHIMICA ACTA, 2024, 191 (03)
  • [6] Paper-based enzymatic reactors for batch injection analysis of glucose on 3D printed cell coupled with amperometric detection
    Dias, Anderson A.
    Cardoso, Thiago M. G.
    Cardoso, Rafael M.
    Duarte, Lucas C.
    Munoz, Rodrigo A. A.
    Richter, Eduardo M.
    Coltro, Wendell K. T.
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2016, 226 : 196 - 203
  • [7] Glucose measurement: Confounding issues in setting targets for inpatients management
    Dungan, Kathleen
    Chapman, John
    Braithwaite, Susan S.
    Buse, John
    [J]. DIABETES CARE, 2007, 30 (02) : 403 - 409
  • [8] Thermoelectric Properties of PEDOT:PSS
    Fan, Zeng
    Ouyang, Jianyong
    [J]. ADVANCED ELECTRONIC MATERIALS, 2019, 5 (11):
  • [9] 3D-printed holder for drawing highly reproducible pencil-on-paper electrochemical devices
    Faustino, Lucas C. C.
    Cunha, Joao P. C.
    Cantanhede, Welter
    Kubota, Lauro T. T.
    Geroncio, Everson T. S.
    [J]. MICROCHIMICA ACTA, 2023, 190 (08)
  • [10] Ganjali MR, 2011, INT J ELECTROCHEM SC, V6, P2808