Dual-Energy Technique at Low Tube Voltages for Small Animal Imaging

被引:4
作者
Cho S. [1 ]
Sidky E.Y. [1 ]
Bian J. [1 ]
Pan X. [1 ]
机构
[1] Department of Radiology, University of Chicago, Chicago
基金
美国国家卫生研究院;
关键词
dual-energy; specimen radiography; spectrum estimation;
D O I
10.1016/S1007-0214(10)70013-4
中图分类号
学科分类号
摘要
We investigate the feasibility of dual-energy method for image contrast enhancement in small animal studies using a low kV X-ray radiographic system. A robust method for X-ray spectrum estimation from transmission measurements, based on expectation-maximization (EM) method, is applied to an X-ray specimen radiographic system for dual energy imaging of a mouse. From transmission measurements of two known attenuators at two different X-ray tube voltages, the X-ray energy spectra are reconstructed using the EM-based method. From the spectra information thus obtained, the transmission data for bone and soft tissue in terms of various thicknesses are generated. Two polynomial functions of transmission data are then sought for to fit the inverted thicknesses of bone and soft-tissue. Scatters in cone-beam projection data acquired at two X-ray energies were corrected. From the scatter-corrected data, a bone thickness map is separated from a soft-tissue thickness map by use of the polynomial functions. © 2010 Tsinghua University Press.
引用
收藏
页码:79 / 86
页数:7
相关论文
共 30 条
[1]  
Grodzins L., Optimum energies for X-ray transmission tomography, Nucl. Inst. Meth., 206, pp. 541-545, (1983)
[2]  
Chen A.S., Marsh D.J., Trumbauer M.E., Inactivation of the mouse melanocortin-3 receptor results in increased fat mass and reduced lean body mass, Nature Genet., 26, pp. 97-102, (2000)
[3]  
Giakos G.C., Chowdhury S., Shah N., Et al., Signal evaluation of a novel dual-energy multimedia imaging sensor, IEEE Trans. Instrum. Meas., 51, pp. 949-954, (2002)
[4]  
Hassler U., Garnero L., Rizo P., X-ray dual-energy calibration based on estimated spectral properties of the experimental system, IEEE Trans. Nucl. Sci., 45, pp. 1699-1712, (1998)
[5]  
Heismann B.J., Leppert J., Stierstorfer K., Density and atomic number measurements with spectral X-ray attenuation method, J. Appl. Phys., 94, pp. 2073-2079, (2003)
[6]  
Jiang H., Computed Tomography-Principles, Designs, Artifacts, and Recent Advances, (2003)
[7]  
Engler P., Friedman W.D., Review of dual-energy computed tomography techniques, Mater Eval., 48, pp. 623-629, (1990)
[8]  
Baird L.C., X-ray spectra vs attenuation data: A theoretical analysis, Med Phys., 8, pp. 319-323, (1981)
[9]  
Huang P.H., Chen T.S., Kase K.R., Reconstruction of diagnostic X-ray spectra by numerical analysis of transmission data, Med. Phys., 13, pp. 707-710, (1986)
[10]  
Boone J.M., X-ray spectral reconstruction from attenuation data using neural networks, Med. Phys., 17, pp. 647-654, (1990)