A novel proposition of radiation energy conservation in radiation dose deformation using deformable image registration

被引:0
|
作者
Kim, Jihun [1 ]
Yoon, Kyungho [2 ]
Kim, Jun Won [1 ]
Kim, Jin Sung [3 ,4 ]
机构
[1] Yonsei Univ, Gangnam Severance Hosp, Dept Radiat Oncol, Coll Med, Seoul, South Korea
[2] Yonsei Univ, Sch Math & Comp Computat Sci & Engn, Seoul, South Korea
[3] Yonsei Univ, Heavy Ion Therapy Res Inst, Yonsei Canc Ctr, Dept Radiat Oncol,Coll Med, Seoul, South Korea
[4] Oncosoft Inc, Seoul, South Korea
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2024年 / 69卷 / 24期
基金
新加坡国家研究基金会;
关键词
radiation energy conservation; dose accumulation; deformable image registration; uncertainty quantification; energy difference error; MASS CHANGES; EXPERIMENTAL VALIDATION; ACCUMULATION; TUMORS;
D O I
10.1088/1361-6560/ad9542
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. The purpose of this study is to analytically derive and validate a novel radiation energy conservation principle for dose mapping via deformable image registration (DIR). Approach. A radiation energy conservation principle for the DIR-based dose-deforming process was theoretically derived with a consideration of the volumetric Jacobian and proven using synthetic examples and a patient case. Furthermore, an energy difference error was proposed that can be used to evaluate the DIR-based dose accumulation uncertainty. For the analytical validation of the proposed energy conservation principle, a synthetic isotropic deformation was considered, and artificial deformation uncertainties were introduced. For the validation with a patient case, a ground truth set of computed tomography images and the corresponding deformation was generated. Radiation energy calculation was performed using both the ground truth deformation and another deformation with uncertainty. Main results. The suggested energy conservation principle was preserved with uncertainty-free deformation, but not with error-containing deformations using both the synthetic examples and the patient case. For a synthetic example with a tumor volume reduction of 27.1% (10% reduction in length in all directions), the energy difference error was calculated to be -29.8% and 37.2% for an over-deforming and under-deforming DIR uncertainty of 0.3 cm. The energy difference error for the patient case (tumor volume reduction of 37.6%) was 2.9% for a displacement vector field with a registration error of 2.0 +/- 3.2 mm. Significance. A novel energy conservation principle for DIR-based dose deformation and the corresponding energy difference error were mathematically formulated and successfully validated using simple synthetic examples and a patient example. With a consideration of the volumetric Jacobian, this investigation proposed a radiation energy conservation principle which can be met only with uncertainty-free deformations.
引用
收藏
页数:17
相关论文
共 50 条
  • [31] Deformable image registration for composite planned doses during adaptive radiation therapy
    Torchia, Joshua
    Velec, Michael
    JOURNAL OF MEDICAL IMAGING AND RADIATION SCIENCES, 2024, 55 (01) : 82 - 90
  • [32] Application of Deformable Image Registration to Quantify the Effect of Spinal Radiation During Childhood
    Scher, E. D.
    Vatner, R. E.
    Huang, K. T.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2018, 102 (03): : E472 - E472
  • [33] Deformable image registration of CT images for automatic contour propagation in radiation therapy
    Wu, Qian
    Cao, Ruifen
    Pei, Xi
    Jia, Jing
    Hu, Liqin
    BIO-MEDICAL MATERIALS AND ENGINEERING, 2015, 26 : S1037 - S1044
  • [34] Evaluation of dose recalculation vs dose deformation in a commercial platform for deformable image registration with a computational phantom
    Fusella, Marco
    Giglioli, Francesca R.
    Fiandra, Christian
    Ragona, Riccardo
    MEDICAL DOSIMETRY, 2018, 43 (01) : 82 - 90
  • [35] Using patient-specific phantoms to evaluate deformable image registration algorithms for adaptive radiation therapy
    Stanley, Nick
    Glide-Hurst, Carri
    Kim, Jinkoo
    Adams, Jeffrey
    Li, Shunshan
    Wen, Ning
    Chetty, Indrin J.
    Zhong, Hualiang
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (06): : 177 - 194
  • [36] A New Evaluation Method of Deformable Image Registration Algorithms for Image-Guided Radiation Therapy
    Saito, Y.
    Tateoka, K.
    Nakata, A.
    Nakazawa, T.
    Abe, T.
    Yano, M.
    Sakata, K.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [37] Retreatment Composite Dose Planning using a Deformable Image Registration Method
    Palmer, M. B.
    Zhang, Y.
    McAleer, M.
    Zhang, J.
    Dong, L.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (03): : S184 - S184
  • [38] Evaluation of Deformable Image Registration Between High Dose Rate Brachytherapy and Intensity Modulated Radiation Therapy for Prostate Cancer
    Yokoyama, Noriomi
    Takemura, Akihiro
    Kojima, Hironori
    Tsukamoto, Kousuke
    Ueda, Shinichi
    Noto, Kimiya
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2018, VOL 3, 2019, 68 (03): : 505 - 508
  • [39] Role of deformable image registration for delivered dose accumulation of adaptive external beam radiation therapy and brachytherapy in cervical cancer
    van Heerden, Laura E.
    Visser, Jorrit
    Koedooder, Kees
    Rasch, Coen R. N.
    Pieters, Bradley R.
    Bel, Arjan
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2018, 10 (06) : 542 - 550
  • [40] IMPROVEMENT OF ACCURACY IN DEFORMABLE REGISTRATION IN RADIATION THERAPY
    Ye, Xiaojing
    Chen, Yunmei
    2008 15TH IEEE INTERNATIONAL CONFERENCE ON IMAGE PROCESSING, VOLS 1-5, 2008, : 2420 - 2423