Micromachined LCP connectors for packaging MEMS devices in biological environments

被引:0
作者
Dean, Robert [1 ]
Weller, Jenny [1 ]
Bozack, Mike [1 ]
Farrell, Brian [1 ]
Jauniskis, Linas [1 ]
Ting, Joseph [1 ]
Edell, David [1 ]
Hetke, Jamille [1 ]
机构
[1] Auburn University, 200 Broun Hall, Auburn
来源
Journal of Microelectronics and Electronic Packaging | 2007年 / 4卷 / 01期
关键词
Biomedical; LCP; MEMS; Micromachined; Packaging;
D O I
10.4071/1551-4897-4.1.17
中图分类号
学科分类号
摘要
Micro- and nano-MEMS technology is being increasingly exploited in biological applications, such as large electrode-count neural prosthesis probe arrays. However, a bottleneck in fully utilizing this technology has been the interconnect between the implanted MEMS device and the external system connected to the implanted device. Since the implanted MEMS device is capable of having a large number of elements, the interconnect must have a sufficient number of electrical connections to communicate with each and every element. Complicating this is the fact that the interconnect requirements may include electrical signals, microfluidics transport and optical signals, all packaged in a miniature biocompatible interconnect cable. Micromachined liquid crystal polymer (LCP) is a promising technology for this application, due to LCP's biocompatibility, chemical inertness, electromechanical properties and its ability to be micromachined. This paper presents the results from the development of MEMS packaging technology suitable for use in biological environments, and is demonstrated in the realization of a prototype micromachined LCP biomedical interconnect device. In particular, the development of the interconnect device demonstrates the realization of biocompatible connectors with high-density ultra-fine pitch electrical traces. © International Microelectronics And Packaging Society - JMEP.
引用
收藏
页码:17 / 22
页数:5
相关论文
共 22 条
  • [1] Hetke J.F., Anderson D.J., Silicon microelectrodes for extracellular recording, Handbook of Neuroprosthetic Methods, (2002)
  • [2] Wise K.D., Anderson D.J., Hetke J.F., Kipke D.R., Najafi K., Wireless implantable microsystems: High-density electronic interfaces to the nervous system, Proceedings of the IEEE, 92, pp. 76-97, (2004)
  • [3] Maynard E.M., Nordhausen C.T., Normann R.A., The Utah intracortical electrode array: A recording structure for potential brain-computer interfaces, Electroencephalogr. Clin. Neurophysiol, 102, pp. 228-239, (1997)
  • [4] Norlin P., Kindlundh M., Mouroux A., Yoshida K., Hofmann U.G., A 32-site neural recording probe fabricated by DRIE of SOI substrates, J. Micromech. Microeng, 12, pp. 414-419, (2002)
  • [5] Yoon T.H., Hwang E.J., Shin D.Y., Park S.I., Oh S.J., Jung S.C., Shin H.C., Kim S.J., A micromachined silicon depth probe for multichannel neural recording, IEEE Trans. Biomed. Eng, 47, pp. 1082-1087, (2000)
  • [6] Cheung K.C., Djupsund K., Dan Y., Lee L.P., Implantable multichannel electrode array based on SOI technology, J. Microelectromech. Systems, 12, pp. 179-184, (2003)
  • [7] Moxon K.A., Kalkhoran N.M., Markert M., Sambito M.A., McKenzie J.L., Webster J.T., Nanostructured surface modification of ceramic-based microelectrodes to enhance biocompatibility for a direct brain-machine interface, IEEE Trans. Biomed. Eng, 51, pp. 881-889, (2004)
  • [8] Buzsaki G., Large scale recording of neuronal ensembles, Nature Neuroscience, 7, pp. 446-451, (2004)
  • [9] Mickey B.J., Middlebrooks J.C., Representation of auditory space by cortical neurons in awake cats, J. Neurosci, 23, pp. 8649-8663, (2003)
  • [10] Martindale J., Mayhew J., Berwick J., Jones M., Martin C., Johnston D., Redgrave P., Zheng Y., The hemodynamic impulse response to a single neural event, J. Cereb. Blood Flow Metab, 23, pp. 546-555, (2003)