Absorbed dose, effective dose and lifetime attributable risk of breast and thyroid cancers in coronary CT angiography: A comparison between 64 and 128 slice CT scanners

被引:0
作者
Mirzaie, Ehsan [1 ]
Jafari, Salman [2 ]
Hadei, Seyed Kamaledin [3 ]
Nikzad, Safoora [4 ]
机构
[1] Hamadan Univ Med Sci, Sch Med, Hamadan, Iran
[2] Hamadan Univ Med Sci, Sch Paramed, Dept Radiol Technol, Hamadan, Iran
[3] Hamadan Univ Med Sci, Sch Med, Dept Radiol, Hamadan, Iran
[4] Hamadan Univ Med Sci, Sch Med, Dept Med Phys, Hamadan, Iran
关键词
Cancer; Coronary CT angiography; Breast; Thyroid; Absorbed dose; Thermoluminescence dosimetry; COMPUTED-TOMOGRAPHY; RADIATION-EXPOSURE; IMAGE QUALITY; REDUCTION; THORAX;
D O I
10.1016/j.radphyschem.2024.111845
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
The attentive consideration of absorbed dose impact on the breast, a radiosensitive organ, remains pivotal in coronary computed tomography angiography (CCTA). Additionally, the thyroid is subjected to scatter radiation during this diagnostic procedure. The primary objective of this study was to comprehensively investigate and quantify the absorbed dose, effective dose, and lifetime attributable risks (LARs) associated with breast and thyroid cancers in the context of CCTA. This investigation enrolled a cohort of 180 patients (90 men and 90 women) who underwent CCTA utilizing three distinct CT scanners: the SOMATOM Definition AS 128 slice (Siemens 128), the Lightspeed VCT 64 Slice (GE 64), and the Brilliance 64 Slice (Philips 64). The precise quantification of absorbed doses to the breast and thyroid was achieved through the utilization of a thermoluminescence dosimeter (TLD). LAR for breast and thyroid cancers was evaluated in accordance with the BEIR VII report. Kruskal-Wallis test was employed to compare the means of absorbed doses and LARs. The average absorbed doses to the breast during CCTA with Siemens 128, GE 64, and Philips 64 scanners were 116.32 +/- 39.02, 82.88 +/- 18.33, and 84.11 +/- 15.18 mGy, respectively (P < 0.05). Correspondingly, the average LAR for breast cancer was found to be 84, 68, and 52 cases per 100,000 persons, respectively. Notably, a significant discrepancy in breast cancer LARs was observed among different age groups (P < 0.05). Regarding thyroid doses with Siemens 128, GE 64, and Philips 64 scanners, average values were 11.82 +/- 7.51, 12.20 +/- 10.87, and 14.49 +/- 5.43 mGy, respectively (P > 0.05). Additionally, the average LAR for thyroid cancer was determined as 0.32, 0.56, and 0.22 cases per 100,000 persons, respectively. Across all age groups, the LAR for thyroid cancer exhibited higher values among women compared to men. However, these differences did not achieve statistical significance (P > 0.05), indicating a need for further investigation. The findings prominently indicated high absorbed doses to the breast in CCTA using protocols investigated in this study (with an average of 93.7 +/- 30.26 mGy). Particularly, the LAR for breast cancer was higher among younger women, specifically those aged between 20 and 40 years (249 cases per 100,000). These results underscore the necessity for further justification and optimization of CCTA protocols.
引用
收藏
页数:6
相关论文
共 42 条
[1]   Direct Quantification of Breast Dose During Coronary CT Angiography and Evaluation of Dose Reduction Strategies [J].
Abadi, Sobhi ;
Mehrez, Hatem ;
Ursani, Ali ;
Parker, Mark ;
Paul, Narinder .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 196 (02) :W152-W158
[2]   Radiation dose to the paediatric undergoing diagnostic coronary angiography and percutaneous intervention procedures [J].
Abuzaid, M. ;
Abdelrazig, A. ;
Sulieman, A. ;
Alkhorayef, M. ;
Babikir, E. ;
Alonazi, B. ;
Bradley, D. A. .
RADIATION PHYSICS AND CHEMISTRY, 2020, 167
[3]   A closer look at the utilized radiation doses during computed tomography pulmonary angiography (CTPA) for COVID-19 patients [J].
Abuzaid, Mohamed ;
Elshami, Wiam ;
Cavli, Baris ;
Ozturk, Ceren ;
ALMisned, Ghada ;
Tekin, H. O. .
RADIATION PHYSICS AND CHEMISTRY, 2023, 211
[4]   COMPARISON OF RADIATION DOSE AND IMAGE QUALITY IN HEAD CT SCANS AMONG MULTIDETECTOR CT SCANNERS [J].
Abuzaid, Mohamed M. ;
Elshami, Wiam ;
Tekin, H. O. ;
Sulieman, Abdelmoneim ;
Bradley, D. A. .
RADIATION PROTECTION DOSIMETRY, 2021, 196 (1-2) :10-16
[5]  
[Anonymous], 2020, Coronavirus 2019 After Action Event Planning
[6]  
[Anonymous], 2006, Health risks from exposure to low levels of ionizing radiation: BEIR VII, Phase 2
[7]  
[Anonymous], 2008, Introduction to Radiological Physics and Radiation Dosimetry
[8]   Radiation Dose to the Thyroid and Gonads in Patients Undergoing Cardiac CT Angiography [J].
Behroozi, Hamid ;
Davoodi, Mohammad ;
Aghasi, Shahriar .
IRANIAN JOURNAL OF RADIOLOGY, 2015, 12 (02)
[9]   Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography [J].
Einstein, Andrew J. ;
Henzlova, Milena J. ;
Rajagopalan, Sanjay .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2007, 298 (03) :317-323
[10]   Radiation Dose at Coronary CT Angiography: Second-Generation Dual-Source CT Versus Single-Source 64-MDCT and First-Generation Dual-Source CT [J].
Fink, Christian ;
Krissak, Radko ;
Henzler, Thomas ;
Lechel, Ursula ;
Brix, Gunnar ;
Takx, Richard A. P. ;
Nance, John W. ;
Abro, Joseph A. ;
Schoenberg, Stefan O. ;
Schoepf, U. Joseph .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 196 (05) :W550-W557