Upgrading a Computed Tomography Dose Index (CTDI) Phantom to an Electron Density (ED) Phantom for Commissioning a CT Simulator

被引:0
|
作者
Elaziz, Rehab A. [1 ]
Azzam, Alia A. [2 ]
Awad, Ibrahim A. [2 ]
Osama, Mohamed [3 ]
Oraby, Ahmed H. [1 ]
机构
[1] Physics Department, Faculty of Science, Mansoura University, Mansoura
[2] Clinical Oncology & Nuclear Medicine Department, Faculty of Medicine, Mansoura University, Mansoura
[3] On-Medical Company, Cairo
关键词
Calibration Curve; Computed Tomography; Density Phantom; Radiotherapy; Relative Electron Density; Treatment Planning System;
D O I
10.22038/IJMP.2023.68799.2206
中图分类号
学科分类号
摘要
Introduction: A crucial step in commissioning a CT simulator is measuring the conversion relationship between the CT number and Relative Electron Density (RED) and transferring it to the radiotherapy TPS for accurate dose calculation. Material and Methods: In order to automatically reconstruct a DE-Rho series with relative electron density maps of the materials used, 14 tissue-mimicking material plugs were made, their properties were measured, and the developed density phantom was scanned by CT-simulator using the Dual Energy^DE_Direct Density protocol. Results: CT numbers (HU) for various densities of materials were determined. Utilizing the resultant HU from CT scans, the relative electron density, or RED, was computed. The HU-RED calibration curve was created using CT scans that were obtained with different tissue replacements. Conclusion: Particularly for the CT-number calibration in radiation therapy planning, the new Tissue Equivalent Materials (TEMs) may simplify the calibration process without sacrificing the accuracy of the stoichiometric calibration. Our objective was met and the purchase cost was avoided thanks to the developed phantom. © (2024), (Mashhad University of Medical Sciences). All Rights Reserved.
引用
收藏
页码:154 / 159
页数:5
相关论文
共 50 条
  • [41] Assessment of the image quality of virtual monochromatic spectral computed tomography images: a phantom study considering object contrast, radiation dose, and frequency characteristics
    Masuda, Shota
    Sugisawa, Koichi
    Minamishima, Kazuya
    Yamazaki, Akihisa
    Jinzaki, Masahiro
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2021, 14 (01) : 41 - 49
  • [42] Assessment of the image quality of virtual monochromatic spectral computed tomography images: a phantom study considering object contrast, radiation dose, and frequency characteristics
    Shota Masuda
    Koichi Sugisawa
    Kazuya Minamishima
    Akihisa Yamazaki
    Masahiro Jinzaki
    Radiological Physics and Technology, 2021, 14 : 41 - 49
  • [43] Evaluation of surface dose and image quality using the half-scan mode in chest computed tomography-guided interventional radiology: a phantom study
    Hasegawa H.
    Sato J.
    Kobayashi I.
    Radiological Physics and Technology, 2018, 11 (2) : 138 - 148
  • [44] Influence of the adaptive iterative dose reduction 3D algorithm on the detectability of low-contrast lesions and radiation dose repeatability in abdominal computed tomography: a phantom study
    Jeong Hee Yoon
    Jeong Min Lee
    Bo Yun Hur
    Jeehyun Baek
    Hackjoon Shim
    Joon Koo Han
    Byung Ihn Choi
    Abdominal Imaging, 2015, 40 : 1843 - 1852
  • [45] Influence of the adaptive iterative dose reduction 3D algorithm on the detectability of low-contrast lesions and radiation dose repeatability in abdominal computed tomography: a phantom study
    Yoon, Jeong Hee
    Lee, Jeong Min
    Hur, Bo Yun
    Baek, Jeehyun
    Shim, Hackjoon
    Han, Joon Koo
    Choi, Byung Ihn
    ABDOMINAL IMAGING, 2015, 40 (06): : 1843 - 1852
  • [47] The Discriminative Power and Stability of Radiomics Features With Computed Tomography Variations Task-Based Analysis in an Anthropomorphic 3D-Printed CT Phantom
    Jimenez-Del-Toro, Oscar
    Aberle, Christoph
    Bach, Michael
    Schaer, Roger
    Obmann, Markus M.
    Flouris, Kyriakos
    Konukoglu, Ender
    Stieltjes, Bram
    Mueller, Henning
    Depeursinge, Adrien
    INVESTIGATIVE RADIOLOGY, 2021, 56 (12) : 820 - 825
  • [48] Measurement of Electron Density and Effective Atomic Number by Dual-Energy Scan Using a 320-Detector Computed Tomography Scanner with Raw Data-Based Analysis: A Phantom Study
    Tatsugami, Fuminari
    Higaki, Toru
    Kiguchi, Masao
    Tsushima, So
    Taniguchi, Akira
    Kaichi, Yoko
    Yamagami, Takuji
    Awai, Kazuo
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2014, 38 (06) : 824 - 827
  • [49] Effects of automatic tube potential selection on radiation dose index, image quality, and lesion detectability in pediatric abdominopelvic CT and CTA: a phantom study
    Brinkley, Michael F.
    Ramirez-Giraldo, Juan C.
    Samei, Ehsan
    Frush, Daniel J.
    Choudhury, Kingshuk Roy
    Wilson, Joshua M.
    Christianson, Olav I.
    Frush, Donald P.
    EUROPEAN RADIOLOGY, 2016, 26 (01) : 157 - 166
  • [50] Effects of automatic tube potential selection on radiation dose index, image quality, and lesion detectability in pediatric abdominopelvic CT and CTA: a phantom study
    Michael F. Brinkley
    Juan C. Ramirez-Giraldo
    Ehsan Samei
    Daniel J. Frush
    Kingshuk Roy Choudhury
    Joshua M. Wilson
    Olav I. Christianson
    Donald P. Frush
    European Radiology, 2016, 26 : 157 - 166