Rapid determination of Se(IV) and tSe in fungal samples by foam electrode-based electrolytic hydride generation coupled atomic fluorescence spectrometry

被引:0
|
作者
Liu, Xiao-na [1 ]
Yang, Xin-an [1 ]
Chu, Xiang-feng [1 ]
Zhang, Wang-bing [1 ]
机构
[1] Anhui Univ Technol, Dept Appl Chem, Maanshan 243002, Anhui, Peoples R China
基金
中国国家自然科学基金;
关键词
Electrolytic hydride generation; Interface reaction; Interference suppression; Nickel foam; Selenium; Ultrasound-assisted digestion; MODIFIED GRAPHITE ELECTRODE; VAPOR GENERATION; WATER SAMPLES; SEQUENTIAL DETERMINATION; SENSITIVE DETERMINATION; EMISSION-SPECTROMETRY; SELENIUM; SPECIATION; REDUCTION; RICE;
D O I
10.1016/j.talanta.2024.127324
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The key to accurately identifying trace heavy metal elements is to achieve efficient sample introduction while shielding the interference of matrix components. Taking the electrolytic hydride generation (EHG) technology as an example, this paper explored the effects of cathode materials and structural factors on the electrosynthesis of hydrogen selenide (H2Se), particularly on suppressing interference from coexisting components. Systematic electrochemical and spectroscopic tests show that the nickel-based electrode can promote the generation of H2Se, while the multi-layer foam structure with large specific surface area, rich pores and weak gas evolution effect improves the yield and stability of electrosynthesis reaction. Even if the surface state of the electrode changes due to the electrodeposition of high concentration interference ion, the electrochemical behavior of selenium (Se) is basically not affected. After coupling with an atomic fluorescence spectrometer detector, this method has a low detection limit (0.13 mu g L- 1), a wide linear range (2-100 mu g L- 1), and stable signal output (RSD, 3.3%, n = 11). With the assistance of high-frequency ultrasound sample extraction and pre-reduction measures, Se(IV) and total Se (tSe) in fungal samples such as mushrooms can be quickly quantified without pre-separation of the matrix. The contribution of this study is to provide an economical and sustainable electrochemical gas separation strategy for spectroscopic quantification of trace and even ultra-trace heavy metal elements in complex matrices.
引用
收藏
页数:10
相关论文
共 50 条
  • [41] Simultaneous determination of As, Bi, Se, Sn and Te in high alloy steels-re-evaluation of hydride generation inductively coupled plasma atomic emission spectrometry
    Wiltsche, Helmar
    Brenner, Isaac B.
    Knapp, Guenter
    Prattes, Karl
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2007, 22 (09) : 1083 - 1088
  • [42] A simple and rapid copper-assisted microprecipitation method for the on-site separation of inorganic arsenic species in water samples followed by hydride generation atomic fluorescence spectrometry determination
    Mullapudi, Venkata Balarama Krishna
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2024, 211
  • [43] Determination of cadmium in water samples by fast pyrolysis - Chemical vapor generation atomic fluorescence spectrometry using titanium hydride powder as a hydrogen source
    Duan, Xuchuan
    Liu, Shuai
    Gao, Wei
    Sun, Jinsheng
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2019, 162
  • [44] Determination of Te, As, Bi, Sb and Se (TABS) in Geological Reference Materials and GeoPT Proficiency Test Materials by Hydride Generation-Atomic Fluorescence Spectrometry (HG-AFS)
    Mansur, Eduardo Teixeira
    Barnes, Sarah-Jane
    Savard, Dany
    Webb, Peter C.
    GEOSTANDARDS AND GEOANALYTICAL RESEARCH, 2020, 44 (01) : 147 - 167
  • [45] Ion Chromatography Coupled with Bi-anode Electrochemical Hydride Generation-Atomic Fluorescence Spectrometry for Determination of Arsenic Species in Dental Pulp Inactivative Agent Type I
    Shentu Chao
    Hou Yi-Zhong
    Fan Yun-Chang
    Zhu Yan
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2009, 37 (02) : 263 - 266
  • [46] Speciation analysis of inorganic arsenic in coal samples by microwave-assisted extraction and high performance liquid chromatography coupled to hydride generation atomic fluorescence spectrometry
    Sun, Mei
    Liu, Guijian
    Wu, Qianghua
    Liu, Wenqi
    TALANTA, 2013, 106 : 8 - 13
  • [47] Multi-walled Carbon Nanotube-Packed Micro-column for On-line Sorption and Determination of Bismuth in Environmental Water Samples by Hydride Generation Atomic Fluorescence Spectrometry
    Wu, Hong
    Wang, Xuecui
    Xu, Changliang
    Lu, Jusheng
    Tian, Jiuying
    Ma, Juan
    ATOMIC SPECTROSCOPY, 2010, 31 (01) : 14 - 20
  • [48] Cold vapor generation of Zn based on dielectric barrier discharge induced plasma chemical process for the determination of water samples by atomic fluorescence spectrometry
    Zhenli Zhu
    Lu Liu
    Yixiao Li
    Huan Peng
    Zhifu Liu
    Wei Guo
    Shenghong Hu
    Analytical and Bioanalytical Chemistry, 2014, 406 : 7523 - 7531
  • [49] Cold vapor generation of Zn based on dielectric barrier discharge induced plasma chemical process for the determination of water samples by atomic fluorescence spectrometry
    Zhu, Zhenli
    Liu, Lu
    Li, Yixiao
    Peng, Huan
    Liu, Zhifu
    Guo, Wei
    Hu, Shenghong
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2014, 406 (29) : 7523 - 7531
  • [50] New method for mercury determination in microwave digested soil samples based on cold vapor capacitively coupled plasma microtorch optical emission spectrometry: Comparison with atomic fluorescence spectrometry
    Frentiu, Tiberiu
    Mihaltan, Alin Ironim
    Senila, Marin
    Darvasi, Eugen
    Ponta, Michaela
    Frentiu, Maria
    Pintican, Bogdan Petru