Rapid determination of Se(IV) and tSe in fungal samples by foam electrode-based electrolytic hydride generation coupled atomic fluorescence spectrometry

被引:0
|
作者
Liu, Xiao-na [1 ]
Yang, Xin-an [1 ]
Chu, Xiang-feng [1 ]
Zhang, Wang-bing [1 ]
机构
[1] Anhui Univ Technol, Dept Appl Chem, Maanshan 243002, Anhui, Peoples R China
基金
中国国家自然科学基金;
关键词
Electrolytic hydride generation; Interface reaction; Interference suppression; Nickel foam; Selenium; Ultrasound-assisted digestion; MODIFIED GRAPHITE ELECTRODE; VAPOR GENERATION; WATER SAMPLES; SEQUENTIAL DETERMINATION; SENSITIVE DETERMINATION; EMISSION-SPECTROMETRY; SELENIUM; SPECIATION; REDUCTION; RICE;
D O I
10.1016/j.talanta.2024.127324
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The key to accurately identifying trace heavy metal elements is to achieve efficient sample introduction while shielding the interference of matrix components. Taking the electrolytic hydride generation (EHG) technology as an example, this paper explored the effects of cathode materials and structural factors on the electrosynthesis of hydrogen selenide (H2Se), particularly on suppressing interference from coexisting components. Systematic electrochemical and spectroscopic tests show that the nickel-based electrode can promote the generation of H2Se, while the multi-layer foam structure with large specific surface area, rich pores and weak gas evolution effect improves the yield and stability of electrosynthesis reaction. Even if the surface state of the electrode changes due to the electrodeposition of high concentration interference ion, the electrochemical behavior of selenium (Se) is basically not affected. After coupling with an atomic fluorescence spectrometer detector, this method has a low detection limit (0.13 mu g L- 1), a wide linear range (2-100 mu g L- 1), and stable signal output (RSD, 3.3%, n = 11). With the assistance of high-frequency ultrasound sample extraction and pre-reduction measures, Se(IV) and total Se (tSe) in fungal samples such as mushrooms can be quickly quantified without pre-separation of the matrix. The contribution of this study is to provide an economical and sustainable electrochemical gas separation strategy for spectroscopic quantification of trace and even ultra-trace heavy metal elements in complex matrices.
引用
收藏
页数:10
相关论文
共 50 条
  • [31] Gold determination in geological samples by chelate vapor generation at room temperature coupled with atomic fluorescence spectrometry
    Duan, Xuchuan
    Sun, Rui
    Fang, Jinliang
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2017, 133 : 21 - 25
  • [32] Multivariate optimization of a method for antimony determination by hydride generation atomic fluorescence spectrometry in hair samples of patients undergoing chemotherapy against Leishmaniasis
    Cardozo, Manuelle C.
    Cavalcante, Dannuza D.
    Silva, Daniel L. F.
    Dos Santos, Walter N. L.
    Bezerra, Marcos A.
    ANAIS DA ACADEMIA BRASILEIRA DE CIENCIAS, 2016, 88 (03): : 1179 - 1190
  • [33] Application of Graphene as a Sorbent for the Preconcentration and Determination of Trace Amounts of Mercury in Water Samples by Hydride Generation Atomic Fluorescence Spectrometry
    Liu, Hai Yan
    Meng, Xu Feng
    Li, Chao
    Jia, An Qiang
    Li, Jing Ci
    Ma, Jing Jun
    JOURNAL OF THE CHEMICAL SOCIETY OF PAKISTAN, 2015, 37 (02): : 252 - 255
  • [34] Determination of Cadmium in Soil Samples with Ascorbic Acid and Sulfanilamide as Enhanced Reagent Using Hydride Generation Atomic Fluorescence Spectrometry
    Lai Dong-Mei
    Deng Tian-Long
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2010, 38 (04) : 542 - 546
  • [35] Determination of ultra-trace amounts of selenium in environmental samples by sequential injection-hydride generation-nondispersive atomic fluorescence spectrometry
    Wang, Z
    Fan, SH
    Fang, ZL
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2005, 33 (02) : 195 - 197
  • [36] Determination of ultra-trace selenium in mineral samples by hydride generation atomic fluorescence spectrometry with pressurized-PTFE-vessel acid digestion
    Zhang, Ni
    Sun, Genling
    Ma, Hairui
    MINERALS ENGINEERING, 2007, 20 (15) : 1397 - 1400
  • [37] Atomic fluorescence spectrometry for ultrasensitive determination of bismuth based on hydride generation - the role of excitation source, interference filter and flame atomizers
    Stadlerova, Barbora
    Kolrosova, Marta
    Dedina, Jiri
    Musil, Stanislav
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2020, 35 (05) : 993 - 1002
  • [38] Highly sensitive determination of trace antimony in water samples by cobalt ion enhanced photochemical vapor generation coupled with atomic fluorescence spectrometry or ICP-MS
    Zeng, Wen
    Hu, Zenghui
    Luo, Jin
    Hou, Xiandeng
    Jiang, Xiaoming
    ANALYTICA CHIMICA ACTA, 2022, 1191
  • [39] Speciation of Se(IV) and the selenoamino acids by high-performance liquid chromatography-direct hydride generation-atomic fluorescence spectrometry
    Ipolyi, I
    Stefánka, Z
    Fodor, P
    ANALYTICA CHIMICA ACTA, 2001, 435 (02) : 367 - 375
  • [40] Quantitative Determination of Antimony, Bismuth, and Tellurium in Geological Samples by Inductively Coupled Plasma Mass Spectrometry and Flow Injection Hydride Generation Atomic Absorption Technique
    Saha, Manik C.
    Baskey, Rashmoni
    Lahiri, Sudipta
    Dutta, Saikat
    ATOMIC SPECTROSCOPY, 2017, 38 (01) : 7 - 11