Development of low-temperature field emission apparatus

被引:0
作者
Kashifuku, Aya [1 ]
Yoshino, Manabu [1 ]
Fukutani, Katsuyuki [1 ]
Okano, Tatsuo [1 ]
机构
[1] Institute of Industrial Science, University of Tokyo, Meguro-ku, Tokyo 153-8505
关键词
D O I
10.3131/jvsj2.52.102
中图分类号
学科分类号
摘要
A field emission apparatus operating at cryogenic temperatures was developed by using an Al-SUS explosion bonded metal composite to get a good thermal contact of the sample and the radiation shield with the liquid helium reservoir. With this apparatus, a tip sample temperature of 4.1 K was observed when the liquid helium reservoir was evacuated. As an electron detector, a micro channel plate (MCP) was set on the bottom of the liquid helium reservoir cooled at ∼10 K. The performance of the MCP at low temperatures was examined by using ultra violet light as an input signal. It was confirmed that the MCP has a sufficient amplification at 11.9 K.
引用
收藏
页码:102 / 104
页数:2
相关论文
共 7 条
  • [1] Gomer R., Field Emission and Field Ionization, (1961)
  • [2] Gadzuk J.W., Surf. Sci, 15, (1969)
  • [3] Klein R., Leder L.B., Phys. Rev, 124, (1961)
  • [4] Nagaoka K., Yamashita T., Uchiyama S., Yamada M., Fujii H., Oshima C., Nature, 396, (1998)
  • [5] Kawauchi T., Matsumoto M., Fukutani K., Okano T., Rev. Sci. Instrum, 78, (2007)
  • [6] Lecomte P., Perez-Mendez V., IEEE Trans. Nucl. Sci, 25, (1978)
  • [7] Sawichi J.A., Tyliszczak T., Gzowski O., Nucl. Instrum. Methods, 190, (1981)