EBT3 Radiochromic film response in time-dependent thermal environment and water submersion conditions: Its clinical relevance and uncertainty estimation

被引:0
作者
Trivedi, Gaurav [1 ,2 ]
Singh, Pushpendra P. [2 ]
Oinam, Arun S. [1 ]
机构
[1] Post Grad Inst Med Educ & Res, Chandigarh, India
[2] Indian Inst Technol, Ropar, India
关键词
Film calibration for 6 MV photon beam; In-vivo dosimetry; EBT3 response to humidity; Thermal effect of EBT3; Film water immersion; DOSIMETRY; TEMPERATURE;
D O I
10.1016/j.radphyschem.2024.112403
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Aim: This study aimed to find the thermal and humidity-dependent behavior of EBT3 film. The changes in the film sensitivity as well as the dosimetric region of interest in the small-size film due to its storage and setup condition were evaluated. Material and methods: Three sets of films were kept at three different thermal storage conditions. 15 films from each set were kept at 2 degrees C, 20 degrees C, and 40 degrees C storage conditions for 24 h and 48 h pre- and post-irradiation. The humidity effect (water submersion) on the film in terms of the film's edge diffusion, and change in the optical density at the central region of the film was estimated by keeping the films underwater. Results: The mean relative film sensitivity was found 1.107 f 0.123 and 0.966 f 0.028 respectively for the films stored at 40 degrees C and 2 degrees C with respect to the sensitivity of the films that were kept at room temperature (20 degrees C) each for 48 h. The dosimetric result shows that the measured dose in the dose range (2-20 Gy) was overestimated by 7.01 % (percentage mean dose difference (MDD)) in the films stored at 40 degrees C for 48 h. In contrast, it was underestimated by -4.54 % (MDD) in the films stored at 2 degrees C for 48 h compared to those kept at an ambient room temperature of 20 degrees C (MDD = 0.24%). The water-diffusion rate at the border of the EBT3 film used in this study was 0.23 mm/h. An increment of 6.8% in the optical density at the central region of the film was observed for the unexposed control film kept underwater for 24 h. The combined standard uncertainty due to five different sources of uncertainties was 2.91%. Conclusion: The sensitivity of the EBT3 film kept at 40 degrees C is always higher (overestimate) whereas lower (underestimate) for the films kept at 2 degrees C as compared to the films at normal room temperature. The correction for the changes in the net optical density due to its storage condition (thermal effect) and the dosimetry condition of water submersion (humidity effect) must be incorporated into the current protocols of film-dosimetry if the facility cannot have temperature-controlled film storage environment and if the film immersion under water is more than 4 h.
引用
收藏
页数:8
相关论文
共 27 条
  • [1] Evaluation of an efficient calibration protocol for use in radiochromic film dosimetry for photon beams
    Akdeniz, Yucel
    Yegingil, Zehra
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2019, 156 : 31 - 37
  • [2] Evaluation of EBT-2 model GAFCHROMIC™ film performance in water
    Aldelaijan, Saad
    Devic, Slobodan
    Mohammed, Huriyyah
    Tomic, Nada
    Liang, Li-Heng
    DeBlois, Francois
    Seuntjens, Jan
    [J]. MEDICAL PHYSICS, 2010, 37 (07) : 3687 - 3693
  • [3] A comprehensive study of the Gafchromic EBT2 radiochromic film. A comparison with EBT
    Andres, C.
    del Castillo, A.
    Tortosa, R.
    Alonso, D.
    Barquero, R.
    [J]. MEDICAL PHYSICS, 2010, 37 (12) : 6271 - 6278
  • [4] GAFCHROMIC® EBT photospectral dose response dependence on temperature and implications for flat bed scanning
    Buchauer, Konrad
    Hillbrand, Elmar
    de Vries, Alexander
    [J]. MEDICAL PHYSICS, 2009, 36 (11) : 5044 - 5051
  • [5] Radiochromic film dosimetry in water phantoms
    Butson, MJ
    Cheung, T
    Yu, PKN
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (01) : N27 - N31
  • [6] A spectroscopic study of the chromatic properties of GafChromic™ EBT3 films
    Callens, M.
    Crijns, W.
    Simons, V.
    De Wolf, I.
    Depuydt, T.
    Maes, F.
    Haustermans, K.
    D'hooge, J.
    D'Agostino, E.
    Wevers, M.
    Pfeiffer, H.
    Van den Abeele, K.
    [J]. MEDICAL PHYSICS, 2016, 43 (03) : 1156 - 1166
  • [7] On the use of Gafchromic EBT3 films for validating a commercial electron Monte Carlo dose calculation algorithm
    Chan, EuJin
    Lydon, Jenny
    Kron, Tomas
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (05) : 2091 - 2102
  • [8] RADIATION-INDUCED POLYMERIZATION
    CHAPIRO, A
    [J]. RADIATION PHYSICS AND CHEMISTRY, 1979, 14 (1-2): : 101 - 116
  • [9] CHAPIRO A, 1974, MAKROMOL CHEM, V175, P1181
  • [10] Reference radiochromic film dosimetry: Review of technical aspects
    Devic, Slobodan
    Tomic, Nada
    Lewis, David
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (04): : 541 - 556