Force modeling for needle insertion into corneal tissue

被引:0
作者
Xuan, Xinxiang [1 ]
Yang, Yang [1 ]
Wang, Zhe [1 ]
Deng, Shijing [2 ]
Liu, Xiaoyu [1 ]
机构
[1] School of Mechanical Engineering and Automation, Beihang University
[2] Research Institute of Beijing Ophthalmology, Capital Medical University
来源
Gaojishu Tongxin/Chinese High Technology Letters | 2009年 / 19卷 / 09期
关键词
Cornea suturing; Force modeling; Needle axial force;
D O I
10.3772/j.issn.1002-0470.2009.09.013
中图分类号
学科分类号
摘要
The mechanical behavior of needle puncturing corneal tissue was analyzed, and the interactive forces between the needle and corneal tissue were resolved into three components, the surface contact force, the friction force and the cutting force. Then a sine loading testing was done based on the modified Karnopp friction model. With the relationship between friction and time, a friction model was established. Then the contact force model was obtained by the quadratic curve fitting from the testing results of surface contact force during needle puncturing tissue. Finally, the cutting force model was concluded according to the testing curve and the friction force model. The research results can be used in fine manipulation during corneal-grafting surgery.
引用
收藏
页码:951 / 956
页数:5
相关论文
共 7 条
  • [1] DiMaio S.P., Salcudean S.E., Needle insertion modeling and simulation, Proceedings of the 2002 IEEE International Conference on Robotics 8 Automation, pp. 2098-2105, (2002)
  • [2] Kataoka H., Washio T., Chinzei K., Et al., Measurement of tip and friction force acting on a needle during penetration, Proceedings of the 5th International Conference on Medical Image Computing and Computer Assisted Intervention, pp. 216-223, (2002)
  • [3] Karnopp D., Computer simulation of stick-slip friction model in dynamical systems, Journal of Dynamic Systems, Measurement, and Control, 107, 1, pp. 00-103, (1985)
  • [4] Allison M.O., Christina S., Mark D.O., Force modeling for needle insertion into soft tissue, IEEE Transactions on Biomedical Engineering, 51, 5, pp. 1707-1716, (2004)
  • [5] Vaughan D., General Ophthalmology, (1999)
  • [6] Simon P.D., Salcudean S., Needle insertion modeling and simulation, IEEE Transactions on Robotics and Automation, 19, 5, pp. 864-873, (2003)
  • [7] Diego D'.A., Remis B., Christian L., A haptic interface for a virtual exam of the human thigh, Proceedings of the 2000 IEEE International Conference on Robotics and Automation, (2000)